510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

RayStation 10.1

K210645510(k) TraditionalCleared Jun 29, 2021118-day reviewClass II21 CFR 892.5050Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

RaySearch Laboratories AB (PUBL) · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Jun 29, 2021 as substantially equivalent to K200569 (RaySearch Laboratories AB (PUBL), 2020) and 2 other predicates after a 118-day review. Its family tree reaches back 16 generations to 2001. 5 later devices descend from it. It has been recalled 6 times.

Decided
Jun 29, 2021
Review
118 days
Class
Class II
Product code
MUJ
Recalls
6
Predicates
3

Its longest line of descent runs back 16 generations to K010975 (2001). 125 ancestors in all, the oldest K802809 (1981). 5 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2020: 1 other ancestor2018: 2 other ancestors2017: 2 other ancestors2014: 3 other ancestors2015: 1 other ancestor2012: 2 other ancestors2008: 4 other ancestors2011: 3 other ancestors2003: 4 other ancestors2002: 6 other ancestors2000: 8 other ancestors2013: 2 other ancestors2005: 3 other ancestors2001: 10 other ancestors2010: 2 other ancestors2006: 4 other ancestors1995: 2 other ancestors2004: 7 other ancestors2009: 2 other ancestors1999: 6 other ancestors1997: 4 other ancestors2007: 1 other ancestor1998: 2 other ancestors1996: 8 other ancestors1991: 4 other ancestors1993: 1 other ancestor1990: 3 other ancestors1989: 3 other ancestors1992: 4 other ancestors1994: 2 other ancestors1987: 1 other ancestor1986: 1 other ancestor1981: 1 other ancestor2021: 1 descendant cleared2022: 1 descendant cleared2023: 1 descendant cleared2025: 2 descendants clearedK010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalledK030981 · Eclipse · Varian Medical Systems, Inc. · 2003K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005×2: 2 generations cleared close together, K071873 (2007), K073020 (2007) K091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalled×2: 2 generations cleared close together, K140187 (2014), K141860 (2014) K160093 · RayStation · RaySearch Laboratories AB (PUBL) · 2016 · recalled×2: 2 generations cleared close together, K170355 (2017), K171536 (2017) K180379 · RayStation 7 · RaySearch Laboratories AB (PUBL) · 2018 · recalledK190387 · RayStation · RaySearch Laboratories AB (PUBL) · 2019 · recalledK200569 · RayStation · RaySearch Laboratories AB (PUBL) · 2020 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors3 generations back: 2 other ancestors4 generations back: 3 other ancestors5 generations back: 5 other ancestors6 generations back: 5 other ancestors7 generations back: 4 other ancestors8 generations back: 1 other ancestor9 generations back: 5 other ancestors10 generations back: 11 other ancestors11 generations back: 17 other ancestors12 generations back: 20 other ancestors13 generations back: 13 other ancestors14 generations back: 15 other ancestors15 generations back: 4 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant K200569 · RayStation · RaySearch Laboratories AB (PUBL) · 2020 · recalledK190387 · RayStation · RaySearch Laboratories AB (PUBL) · 2019 · recalledK180379 · RayStation 7 · RaySearch Laboratories AB (PUBL) · 2018 · recalledK171536 · RayStation · RaySearch Laboratories AB (PUBL) · 2017 · recalledK170355 · RayStation 6 · RaySearch Laboratories AB (PUBL) · 2017 · recalledK160093 · RayStation · RaySearch Laboratories AB (PUBL) · 2016 · recalledK141860 · Raystation · Raysearch Laboratories AB · 2014 · recalledK140187 · Ray Station · Raysearch Laboratories AB · 2014 · recalledK102216 · Xio RTP System · Computerized Medical Systems, Inc. · 2010 · recalledK091492 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2009 · recalledK073020 · Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK071873 · Modification to Eclipse Treatment Planning System · Varian Medical Systems, Inc. · 2007 · recalledK050296 · Eclipse Treatment Planning System · Varian Medical Systems · 2005K041403 · Eclipse with Aaa · Varian Medical Systems · 2004K030981 · Eclipse · Varian Medical Systems, Inc. · 2003K010975 · Eclipse 6.5 · Varian Medical Systems, Inc. · 2001 · recalled−15−10−50+5
  1. Root · 16 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 12 generations between
  3. 3 generations backK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
  4. 2 generations backK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
  5. 1 generation backK200569 RayStationRaySearch Laboratories AB (PUBL) · 2020 · Recalled
  6. This deviceK210645 RayStation 10.1RaySearch Laboratories AB (PUBL) · 2021
  7. Descendant · cites it directly · 4 of its ownK211867 RayStation 11.0RaySearch Laboratories AB (PUBL) · 2021 · Recalled
  8. Descendant · 2 generations on · 3 of its ownK220141 RayStation 11BRaySearch Laboratories AB (PUBL) · 2022 · Recalled
  9. Descendant · 3 generations on · 2 of its ownK222312 RayStation 12ARaySearch Laboratories AB (PUBL) · 2023 · Recalled
  10. Descendant · 4 generations on · 1 of its ownK240398 RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A…RaySearch Laboratories AB (PUBL) · 2025 · Recalled
  11. Descendant · 5 generations on · 0 of its ownK252109 RayStation (2024A SP3)RaySearch Laboratories AB (PUBL) · 2025
Show all 16 generations
  1. Root · 16 generations backK010975 Eclipse 6.5Varian Medical Systems, Inc. · 2001 · Recalled
  2. 15 generations backK030981 EclipseVarian Medical Systems, Inc. · 2003
  3. 14 generations backK041403 Eclipse with AaaVarian Medical Systems · 2004
  4. 13 generations backK050296 Eclipse Treatment Planning SystemVarian Medical Systems · 2005
  5. 12 generations backK071873 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  6. 11 generations backK073020 Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2007 · Recalled
  7. 10 generations backK091492 Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc. · 2009 · Recalled
  8. 9 generations backK102216 Xio RTP SystemComputerized Medical Systems, Inc. · 2010 · Recalled
  9. 8 generations backK140187 Ray StationRaysearch Laboratories AB · 2014 · Recalled
  10. 7 generations backK141860 RaystationRaysearch Laboratories AB · 2014 · Recalled
  11. 6 generations backK160093 RayStationRaySearch Laboratories AB (PUBL) · 2016 · Recalled
  12. 5 generations backK170355 RayStation 6RaySearch Laboratories AB (PUBL) · 2017 · Recalled
  13. 4 generations backK171536 RayStationRaySearch Laboratories AB (PUBL) · 2017 · Recalled
  14. 3 generations backK180379 RayStation 7RaySearch Laboratories AB (PUBL) · 2018 · Recalled
  15. 2 generations backK190387 RayStationRaySearch Laboratories AB (PUBL) · 2019 · Recalled
  16. 1 generation backK200569 RayStationRaySearch Laboratories AB (PUBL) · 2020 · Recalled
  17. This deviceK210645 RayStation 10.1RaySearch Laboratories AB (PUBL) · 2021

Most influential descendants

Showing 5 of 5 · See all
  • K211867RayStation 11.0RaySearch Laboratories AB (PUBL) · 2021 · cites it directly4descendants
  • K220141RayStation 11BRaySearch Laboratories AB (PUBL) · 2022 · 2 generations on3descendants
  • K222312RayStation 12ARaySearch Laboratories AB (PUBL) · 2023 · 3 generations on2descendants
  • K240398RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation 2024A SP3…RaySearch Laboratories AB (PUBL) · 2025 · 4 generations on1descendant
  • K252109RayStation (2024A SP3)RaySearch Laboratories AB (PUBL) · 2025 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−16K010975Eclipse 6.5Varian Medical Systems, Inc.20013
−15K030981EclipseVarian Medical Systems, Inc.2003—
−14K041403Eclipse with AaaVarian Medical Systems2004—
−13K050296Eclipse Treatment Planning SystemVarian Medical Systems2005—
−12K071873Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20075
−11K073020Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20076
−10K091492Modification to Eclipse Treatment Planning SystemVarian Medical Systems, Inc.20092
−9K102216Xio RTP SystemComputerized Medical Systems, Inc.20107
−8K140187Ray StationRaysearch Laboratories AB20141
−7K141860RaystationRaysearch Laboratories AB20145
−6K160093RayStationRaySearch Laboratories AB (PUBL)20167
−5K170355RayStation 6RaySearch Laboratories AB (PUBL)20171
−4K171536RayStationRaySearch Laboratories AB (PUBL)20176
−3K180379RayStation 7RaySearch Laboratories AB (PUBL)20187
−2K190387RayStationRaySearch Laboratories AB (PUBL)201932
−1K200569RayStationRaySearch Laboratories AB (PUBL)202014
0 (this device)K210645RayStation 10.1RaySearch Laboratories AB (PUBL)2021—
+1K211867RayStation 11.0RaySearch Laboratories AB (PUBL)20218
+2K220141RayStation 11BRaySearch Laboratories AB (PUBL)20226
+3K222312RayStation 12ARaySearch Laboratories AB (PUBL)20236
+4K240398RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation…RaySearch Laboratories AB (PUBL)20252
+5K252109RayStation (2024A SP3)RaySearch Laboratories AB (PUBL)2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198710
198930
199030
199140
199240
199310
199420
199520
199680
199740
199820
199960
200080
2001100
200260
200340
200470
200530
200640
200710
200840
200920
201020
201130
201220
201320
201430
201510
201720
201820
202010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202210
202310
202520
FIG. 1 — Lineage of K210645, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K200569RayStation
Summary page 4
…on name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) 1.7 Predicate devices 1.7.1 Primary predicate device K200569 RayStation 9.1 1.7.2 Secondary predicate devices K203172 MOSAIQ Oncology Information System…
RaySearch Laboratories AB (PUBL)202014
*K203172MOSAIQ Oncology Information System
Summary page 4
…050, Product Code MUJ) 1.7 Predicate devices 1.7.1 Primary predicate device K200569 RayStation 9.1 1.7.2 Secondary predicate devices K203172 MOSAIQ Oncology Information System K181145 Eclipse Treatment Planning System 1.8 Device descrip…
Elekta Solutions AB20201
*K181145Eclipse Treatment Planning System
Summary page 4
…ary predicate device K200569 RayStation 9.1 1.7.2 Secondary predicate devices K203172 MOSAIQ Oncology Information System K181145 Eclipse Treatment Planning System 1.8 Device description RayStation is a treatment planning system for pla…
Varian Medical Systems, Inc.20181

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211867RayStation 11.0RaySearch Laboratories AB (PUBL)20214

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 28, 2025Class IIZ-1102-2026Under Investigation by firmPotential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of…
Aug 8, 2025Class IIZ-2490-2025Under Investigation by firmDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Jan 16, 2025Class IIZ-1188-2025Under Investigation by firmInconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in…
Mar 28, 2024Class IIZ-1747-2024Software designPotential for reported SSD to be too high.
Feb 23, 2024Class IIZ-1482-2024Under Investigation by firmA use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
Sep 9, 2021Class IIZ-0848-2022Under Investigation by firmIf the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 5 companies; a new generation was cleared every 1.3 years on median.
  • 43 of the 125 ancestors have a recall of any class.
  • Of its descendants, K211867 (RaySearch Laboratories AB (PUBL), 2021) has gone furthest, with 4 descendants of its own.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2021, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

The system functionality can be configured based on user needs.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K210645: RayStation 10.1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K210645

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026