Philips Ingenuity CT

K160743510(k) TraditionalCleared Aug 8, 2016144-day reviewClass II21 CFR 892.1750Recalls 24

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Aug 8, 2016 as substantially equivalent to K033326 (Philips Medical Systems, Inc., 2003) after a 144-day review. Its family tree reaches back 5 generations to 1993. 9 later devices descend from it. It has been recalled 24 times.

Decided
Aug 8, 2016
Review
144 days
Class
Class II
Product code
JAK
Recalls
24
Predicates
1

Its longest line of descent runs back 5 generations to K930090 (1993). 10 ancestors in all. 9 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2022: 3 descendants cleared2020: 1 descendant cleared2018: 1 descendant cleared2021: 2 descendants cleared2024: 1 descendant cleared2026: 1 descendant clearedReference device: K131773 · Brilliance Ict · Philips Healthcare (Cleveland) · 2013Reference device: K033357 · Acqsim-Multislice-CT, Model 453567042271 · Philips Medical Systems (Cleveland), Inc. · 2004 · recalledReference device: K060937 · Brilliance Volume · Philips Medical Systems (Cleveland), Inc. · 2006 · recalledReference device: K123599 · Ingenuity Digital Pet/CT · Philips Healthcare (Cleveland) · 2013 · recalledReference device: K133674 · Philips Iqon Spectral CT · Philips Healthcare (Cleveland) · 2014 · recalledReference device: K123576 · Imr Software Application · Philips Healthcare (Cleveland) · 2013Reference device: K091195 · Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511 · Philips Medical Systems (Cleveland), Inc. · 2009 · recalledReference device: K033677 · Brain Perfusion Option · Philips Medical Systems (Cleveland), Inc. · 2004Reference device: K042293 · Brilliance CT, Private Practice CV Configuration, CT Scanner · Philips Medical Systems (Cleveland), Inc. · 2004 · recalledK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor4 generations back: 3 other ancestors5 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 5 descendants3 generations forward: 1 descendant K033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50
  1. Root · 5 generations backK930090 HelicatElscint, Inc. · 1993
  2. 4 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 3 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 2 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 1 generation backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. This deviceK160743 Philips Ingenuity CTPhilips Medical Systems (Cleveland), Inc. · 2016
  7. Descendant · cites it directly · 7 of its ownK180015 Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2018 · Recalled
  8. Descendant · cites it directly · 3 of its ownK212441 Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2022 · Recalled
  9. Descendant · 2 generations on · 1 of its ownK232491 CT 5300Philips Healthcare (Suzhou) Co., Ltd. · 2024 · Recalled
  10. Descendant · 2 generations on · 1 of its ownK211168 Philips Incisive CT on TrailerPhilips Healthcare (Suzhou) Co., Ltd. · 2021 · Recalled
  11. Descendant · 3 generations on · 0 of its ownK252992 CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2018: 12020: 12021: 22022: 32024: 12026: 1
2018high 3 (2022) · low 0 (2019)2026

Most influential descendants

Showing 5 of 9 · See all
  • K180015Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2018 · cites it directly7descendants
  • K212441Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · 2022 · cites it directly3descendants
  • K232491CT 5300Philips Healthcare (Suzhou) Co., Ltd. · 2024 · 2 generations on1descendant
  • K211168Philips Incisive CT on TrailerPhilips Healthcare (Suzhou) Co., Ltd. · 2021 · 2 generations on1descendant
  • K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K930090HelicatElscint, Inc.1993—
−4K961464Select/SPElscint, Inc.19963
−3K982060Volumax CT Imaging SystemElscint, Inc.1998—
−2K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−1K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
0 (this device)K160743Philips Ingenuity CTPhilips Medical Systems (Cleveland), Inc.2016—
+1K180015Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.201823
+1K212441Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.202217
+2K232491CT 5300Philips Healthcare (Suzhou) Co., Ltd.20245
+2K211168Philips Incisive CT on TrailerPhilips Healthcare (Suzhou) Co., Ltd.20211
+3K252992CT Rembra RT; CT Areta RT; CT RembraPhilips Healthcare (Suzhou) Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202010
202120
202230
202410
202610
FIG. 1 — Lineage of K160743, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033326Philips Plus Computed Tomography System
Summary page 6
…92.1750 Class: II Product Code: JAK Panel: Radiology Primary Predicate device K033326 – Philips Plus CT Scanner Premarket Notification [510(k)] Submission Philips Medical Systems (Cleveland), Inc.…
Philips Medical Systems, Inc.2003131
†K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
†K033357Acqsim-Multislice-CT, Model 453567042271Philips Medical Systems (Cleveland), Inc.200450
†K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200661
†K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland)20135
†K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)20146
†K123576Imr Software ApplicationPhilips Healthcare (Cleveland)2013—
†K091195Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511Philips Medical Systems (Cleveland), Inc.20098
†K033677Brain Perfusion OptionPhilips Medical Systems (Cleveland), Inc.2004—
†K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.20041

Also cleared against K033326: 8 other devices →

What descends from it? (9)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180015Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.20187
K212441Philips Incisive CTPhilips Healthcare (Suzhou) Co., Ltd.20223
K192590CURA 778Fmi Medical Systems, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 7, 2026Class IIZ-3042-2026Radiation Control for Health and Safety ActPhilips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Jun 11, 2026Class IIZ-2652-2026Radiation Control for Health and Safety ActThe clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis…
Aug 13, 2025Class IIZ-2600-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Jun 8, 2021Class IIZ-2315-2021Software designWhen setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Jun 8, 2021Class IIZ-2316-2021Software designWhen setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
May 1, 2021Class IIZ-1984-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Jun 24, 2019Class IIZ-2188-2019OtherIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Jun 24, 2019Class IIZ-2186-2019OtherIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Jun 24, 2019Class IIZ-2187-2019OtherIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Mar 28, 2018Class IIZ-2277-2018OtherA problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Mar 28, 2018Class IIZ-2274-2018OtherA problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Mar 28, 2018Class IIZ-2275-2018OtherA problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Nov 10, 2017Class IIZ-0390-2018Under Investigation by firmPhilips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the…
Nov 10, 2017Class IIZ-0391-2018Under Investigation by firmPhilips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the…
Oct 6, 2017Class IIZ-0520-2018Software designNumerous issues related to software Brilliance iCT 4.1.6 software version.
Oct 6, 2017Class IIZ-0518-2018Software designNumerous issues related to software Brilliance iCT 4.1.6 software version.
Oct 6, 2017Class IIZ-0519-2018Software designNumerous issues related to software Brilliance iCT 4.1.6 software version.
Oct 6, 2017Class IIZ-0521-2018Software designNumerous issues related to software Brilliance iCT 4.1.6 software version.
Nov 12, 2016Class IIZ-0695-2017Software designSoftware error due to the filament on timer
Nov 12, 2016Class IIZ-0697-2017Software designSoftware error due to the filament on timer
Nov 12, 2016Class IIZ-0696-2017Software designSoftware error due to the filament on timer
Oct 13, 2016Class IIIZ-0595-2017Labeling Change ControlThe product label does not include the correct current rating.
Oct 13, 2016Class IIIZ-0597-2017Labeling Change ControlThe product label does not include the correct current rating.
Oct 13, 2016Class IIIZ-0596-2017Labeling Change ControlThe product label does not include the correct current rating.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • Of its descendants, K180015 (Philips Healthcare (Suzhou) Co., Ltd., 2018) has gone furthest, with 7 descendants of its own.
  • 8 other devices also cite its predicate K033326.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.…

Read the full statement (152 words)

The Ingenuity CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. The Ingenuity CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K160743: Philips Ingenuity CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K160743

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026