510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Philips Plus Computed Tomography System

K033326510(k) TraditionalCleared Oct 29, 200313-day reviewClass II21 CFR 892.1750Recalls 131

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Oct 29, 2003 as substantially equivalent to K012009 (Marconi Medical Systems, Inc., 2001) and 1 other predicate after a 13-day review. Its family tree reaches back 4 generations to 1993. 214 later devices descend from it. It has been recalled 131 times.

Decided
Oct 29, 2003
Review
13 days
Class
Class II
Product code
JAK
Recalls
131
Predicates
2

Its longest line of descent runs back 4 generations to K930090 (1993). 9 ancestors in all. 214 later devices descend from it, 9 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2005: 3 descendants cleared2007: 1 descendant cleared2016: 2 descendants cleared2004: 1 descendant cleared2006: 2 descendants cleared2015: 2 descendants cleared2018: 12 descendants cleared2013: 6 descendants cleared2011: 3 descendants cleared2010: 2 descendants cleared2009: 2 descendants cleared2017: 10 descendants cleared2008: 4 descendants cleared2020: 20 descendants cleared2012: 4 descendants cleared2014: 7 descendants cleared2023: 17 descendants cleared2022: 21 descendants cleared2021: 14 descendants cleared2024: 17 descendants cleared2025: 32 descendants cleared2019: 7 descendants cleared2026: 25 descendants clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 9 descendants2 generations forward: 26 descendants3 generations forward: 28 descendants4 generations forward: 23 descendants5 generations forward: 25 descendants6 generations forward: 22 descendants7 generations forward: 27 descendants8 generations forward: 17 descendants9 generations forward: 11 descendants10 generations forward: 7 descendants11 generations forward: 10 descendants12 generations forward: 6 descendants13 generations forward: 2 descendants14 generations forward: 1 descendant K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 19930+5+10
  1. Root · 4 generations backK930090 HelicatElscint, Inc. · 1993
  2. 3 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 2 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 1 generation backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. This deviceK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003
  6. Descendant · cites it directly · 147 of its ownK042293 Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  7. Descendant · 2 generations on · 92 of its ownK130159 Qlab Quantification Software Heart ModelPhilips Health Care · 2013
  8. Descendant · 3 generations on · 91 of its ownK132165 Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · Recalled
  9. Descendant · 2 generations on · 49 of its ownK062650 EP Navigator SoftwarePhilips Medical Systems · 2006
  10. Descendant · 4 generations on · 45 of its ownK171314 QLAB Advanced Quantification SoftwarePhilips Health Care · 2017

Showing 5 of 214 descendants. See all descendants

Descendants cleared per year

2004: 12005: 32006: 22007: 12008: 42009: 22010: 22011: 32012: 42013: 62014: 72015: 22016: 22017: 102018: 122019: 72020: 202021: 142022: 212023: 172024: 172025: 322026: 25
2004high 32 (2025) · low 1 (2004)2026

Most influential descendants

Showing 5 of 214 · See all
  • K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · cites it directly147descendants
  • K130159Qlab Quantification Software Heart ModelPhilips Health Care · 2013 · 2 generations on92descendants
  • K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · 3 generations on91descendants
  • K062650EP Navigator SoftwarePhilips Medical Systems · 2006 · 2 generations on49descendants
  • K171314QLAB Advanced Quantification SoftwarePhilips Health Care · 2017 · 4 generations on45descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K930090HelicatElscint, Inc.1993—
−3K961464Select/SPElscint, Inc.19963
−2K982060Volumax CT Imaging SystemElscint, Inc.1998—
−1K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
0 (this device)K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003—
+1K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.20041
+2K130159Qlab Quantification Software Heart ModelPhilips Health Care2013—
+3K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc.20131
+2K062650EP Navigator SoftwarePhilips Medical Systems2006—
+4K171314QLAB Advanced Quantification SoftwarePhilips Health Care2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200530
200620
200710
200840
200920
201020
201130
201240
201360
201470
201520
201620
2017100
2018120
201970
2020200
2021140
2022210
2023170
2024170
2025320
2026250
FIG. 1 — Lineage of K033326, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
*K010817MX8000 V5.0 CT SystemPhilips Medical Systems (Cleveland), Inc.200112

Also cleared against K012009: 10 other devices →

What descends from it? (214)

See all 214

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.2004147
K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200624
K051170Gemini GXLPhilips Medical Systems (Cleveland), Inc.200513
K172135uCT Computed Tomography X-ray SystemShanghai United Imaging Healthcare Co., Ltd.201812
K160743Philips Ingenuity CTPhilips Medical Systems (Cleveland), Inc.20169
K151383NeuViz 128 Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20152
K153444Philips Multislice CT System with Low Dose CT Lung CancerPhilips Medical Systems (Cleveland), Inc.20161
K071308Modification to Neuviz Dual Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20071
K051351Precedence Spect/CT Imaging SystemAdac Laboratories20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 7, 2026Class IIZ-3043-2026Radiation Control for Health and Safety ActPhilips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
Aug 13, 2025Class IIZ-2595-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Aug 13, 2025Class IIZ-2604-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Aug 13, 2025Class IIZ-2602-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Aug 13, 2025Class IIZ-2603-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
May 1, 2021Class IIZ-1982-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Oct 18, 2018Class IIZ-0543-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Oct 18, 2018Class IIZ-0544-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Jul 3, 2018Class IIZ-1143-2019Device DesignPhilips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator…
Mar 30, 2018Class IIZ-2174-2018Device DesignThe patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent…
Feb 21, 2018Class IIZ-1396-2018Under Investigation by firmSurview scan with tube current lower than 30 mA is unable to be initialized.
Feb 8, 2018Class IIZ-1406-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1413-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1407-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1405-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1404-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Nov 10, 2017Class IIZ-0389-2018Under Investigation by firmPhilips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the…
Nov 12, 2016Class IIZ-0694-2017Software designSoftware error due to the filament on timer
Nov 12, 2016Class IIZ-0693-2017Software designSoftware error due to the filament on timer
Aug 24, 2016Class IIZ-0012-2017Software designThe infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image…
Aug 24, 2016Class IIZ-0009-2017Software designThe infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image…
Aug 24, 2016Class IIZ-0010-2017Software designThe infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image…
Aug 24, 2016Class IIZ-0011-2017Software designThe infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image…
Jul 20, 2016Class IIZ-2652-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jul 20, 2016Class IIZ-2658-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jul 20, 2016Class IIZ-2657-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jun 21, 2016Class IIZ-2113-2016Nonconforming Material/ComponentSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
Jun 21, 2016Class IIZ-2115-2016Nonconforming Material/ComponentSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
Jun 21, 2016Class IIZ-2114-2016Nonconforming Material/ComponentSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
Apr 29, 2016Class IIZ-2380-2016Software designSoftware issues found in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 products that could affect the performance of the equipment.
Apr 29, 2016Class IIZ-2385-2016Software designSoftware issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the performance of the equipment.
Apr 29, 2016Class IIZ-2383-2016Software designSoftware issues in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 and Ingenuity CT products that could affect the performance of the equipment.
Apr 29, 2016Class IIZ-2384-2016Software designSoftware issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the performance of the equipment.
Apr 8, 2016Class IIZ-1875-2016Software designThe firm became aware of a problem where the system may not map Varian drive after CT user logout/login.
Apr 8, 2016Class IIZ-1872-2016Software designThe firm became aware of a problem where the system may not map Varian drive after CT user logout/login.
Apr 8, 2016Class IIZ-1873-2016Software designThe firm became aware of a problem where the system may not map Varian drive after CT user logout/login.
Apr 8, 2016Class IIZ-1874-2016Software designThe firm became aware of a problem where the system may not map Varian drive after CT user logout/login.
Apr 1, 2016Class IIZ-2117-2016Software design (manufacturing process)Philips Healthcare received reports from the field that certain Ingenuity Core systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be…
Apr 1, 2016Class IIZ-2118-2016Software design (manufacturing process)Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be…
Apr 1, 2016Class IIZ-2116-2016Software design (manufacturing process)Philips Healthcare received reports from the field that certain Brilliance 64 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be…
Mar 24, 2016Class IIZ-2366-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 24, 2016Class IIZ-2371-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 24, 2016Class IIZ-2372-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 4, 2016Class IIZ-1506-2016Software designThe AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data…
Feb 5, 2016Class IIZ-1355-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Feb 5, 2016Class IIZ-1351-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Feb 5, 2016Class IIZ-1356-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Feb 5, 2016Class IIZ-1354-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Oct 29, 2015Class IIZ-0548-2016Software designPerfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion…
Oct 29, 2015Class IIZ-0547-2016Software designPerfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
  • 3 of the 9 ancestors have a recall of any class.
  • Of its descendants, K042293 (Philips Medical Systems (Cleveland), Inc., 2004) has gone furthest, with 147 descendants of its own.
  • 10 other devices also cite its predicate K012009.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The “Philips Plus” is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033326: Philips Plus Computed Tomography System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033326

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026