510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Plum Solo™ Precision IV Pump

K242114510(k) TraditionalCleared Apr 2, 2025257-day reviewClass II21 CFR 880.5725Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · FRN Pump, Infusion

Cleared on Apr 2, 2025 as substantially equivalent to K223607 (Icu Medical, Inc., 2023) after a 257-day review. Its family tree reaches back 8 generations to 2001. 10 of the 28 devices in its family, counting itself, have a Class I recall. It has been recalled 3 times.

Decided
Apr 2, 2025
Review
257 days
Class
Class II
Product code
FRN
Recalls
3
Predicates
1

Its longest line of descent runs back 8 generations to K011442 (2001). 27 ancestors in all, the oldest K780880 (1978). No later device cites it as a predicate yet. 10 of the 28 devices in this family, counting this one, have a Class I recall.

  1. Root · 8 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 7 generations backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. 6 generations backK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004 · Recalled (Class I)
  4. 5 generations backK052052 Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3…Hospira, Inc. · 2005 · Recalled (Class I)
  5. 4 generations backK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  6. 3 generations backK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  7. 2 generations backK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  8. 1 generation backK223607 Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · Recalled (Class I)
  9. This deviceK242114 Plum Solo™ Precision IV PumpIcu Medical, Inc. · 2025

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−7K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
−6K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004Class I
−5K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump…Hospira, Inc.2005Class I
−4K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
−3K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
−2K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
−1K223607Plum Duo™ Infusion SystemIcu Medical, Inc.2023Class I
0 (this device)K242114Plum Solo™ Precision IV PumpIcu Medical, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198110
199110
199410
199910
200230
200320
200411
200520
200631
201021
201210
FIG. 1 — Lineage of K242114, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K223607Plum Duo™ Infusion System
Summary page 5
…25 Panel 80 Product Code(s) FRN Legally marketed device(s) to Plum Duo™ Infusion System (K223607) which equivalence is claimed The submission is the traditional pre-market notification for n…
Icu Medical, Inc.2023Class I

Also cleared against K223607: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2026Class IIZ-2724-2026Under Investigation by firmIn a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Apr 29, 2026Class IIZ-2311-2026Software Design ChangeUnder certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Dec 15, 2025Class IIZ-1120-2026Software designPlum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · 3 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K101677Hospira Infusion Blood SetsHospira, Inc. · 2010 · 4 generations back · ◆ in the 2010 ancestors bar of FIG. 18 Class I recalls
  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · 4 generations back · ◆ on the line of descent in FIG. 14 Class I recalls
  • K060806Hospira Gemstar Infusion Pump System, 7 Therapy, Model 13087; 6 Therapy, Model 13086; Pain Management, Model…Hospira, Inc. · 2006 · 5 generations back · ◆ in the 2006 ancestors bar of FIG. 17 Class I recalls
  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041550Hospira Phoenix Infusion System with Hospira Mednet SoftwareAbbott Laboratories · 2004 · 5 generations back · ◆ in the 2004 ancestors bar of FIG. 12 Class I recalls
  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 24 years and passes through 3 companies; a new generation was cleared every 2.2 years on median.
  • 21 of the 27 ancestors have a recall of any class; 10 have a Class I recall.
  • 1 other device also cites its predicate K223607.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.

The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.

The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242114: Plum Solo™ Precision IV Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K242114

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026