510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR 5300 and MR 7700 MR System

K251397510(k) SpecialCleared Jun 4, 202529-day reviewClass II21 CFR 892.1000Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jun 4, 2025 as substantially equivalent to K223458 (Philips Medical Systems Nederland B.V., 2023) after a 29-day review. Its family tree reaches back 12 generations to 2003. 4 of the 21 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 15 times.

Decided
Jun 4, 2025
Review
29 days
Class
Class II
Product code
LNH
Recalls
15
Predicates
1

Its longest line of descent runs back 12 generations to K031815 (2003). 19 ancestors in all, the oldest K001796 (2000). 1 later device descends from it, 1 citing it directly. 4 of the 21 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2022: 2 other ancestors2016: 1 other ancestor2015: 1 other ancestor2005: 1 other ancestor, 1 with a Class I recall12001: 1 other ancestor2000: 1 other ancestor2026: 1 descendant clearedK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalledK041602 · Panorama 1.0T · Philips Medical Systems North America, Inc. · 2004 · Class I recallK052078 · Achieva, Intera and Panorama 1.0 Release 2-Series · Philips Medical Systems Nederland B.V. · 2005 · Class I recallK063559 · Achieva, Intera and Panorama 1.0T, Release 2.5 · Philips Medical Systems · 2007 · Class I recallK110151 · Achieva R4 1.5T and Achieva R4 3.0T · Philips Health Care · 2011 · recalledK163116 · Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 · Philips Medical Systems Nederland B.V. · 2017 · recalledK173079 · Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 · Philips Medical Systems Nederland B.V. · 2018 · recalledK183063 · Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems · Philips Medical Systems Nederland B.V. · 2019 · recalledK193215 · Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems · Philips Medical Systems Nederland B.V. · 2020 · recalledK212673 · MR 5300 · Philips Medical Systems Nederland B.V. · 2021 · recalledK223442 · MR 5300 and MR 7700 R11 MR Systems · Philips Medical Systems Nederlands B.V. · 2022 · recalledK223458 · Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems · Philips Medical Systems Nederland B.V. · 2023 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 1 other ancestor5 generations back: 1 other ancestor6 generations back: 1 other ancestor, 1 with a Class I recall17 generations back: 1 other ancestor8 generations back: 1 other ancestor 1 generation forward: 1 descendant K223458 · Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems · Philips Medical Systems Nederland B.V. · 2023 · recalledK223442 · MR 5300 and MR 7700 R11 MR Systems · Philips Medical Systems Nederlands B.V. · 2022 · recalledK212673 · MR 5300 · Philips Medical Systems Nederland B.V. · 2021 · recalledK193215 · Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems · Philips Medical Systems Nederland B.V. · 2020 · recalledK183063 · Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems · Philips Medical Systems Nederland B.V. · 2019 · recalledK173079 · Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 · Philips Medical Systems Nederland B.V. · 2018 · recalledK163116 · Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 · Philips Medical Systems Nederland B.V. · 2017 · recalledK110151 · Achieva R4 1.5T and Achieva R4 3.0T · Philips Health Care · 2011 · recalledK063559 · Achieva, Intera and Panorama 1.0T, Release 2.5 · Philips Medical Systems · 2007 · Class I recallK052078 · Achieva, Intera and Panorama 1.0 Release 2-Series · Philips Medical Systems Nederland B.V. · 2005 · Class I recallK041602 · Panorama 1.0T · Philips Medical Systems North America, Inc. · 2004 · Class I recallK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalled−10−50
  1. Root · 12 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
  2. 8 generations between
  3. 3 generations backK212673 MR 5300Philips Medical Systems Nederland B.V. · 2021 · Recalled
  4. 2 generations backK223442 MR 5300 and MR 7700 R11 MR SystemsPhilips Medical Systems Nederlands B.V. · 2022 · Recalled
  5. 1 generation backK223458 Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and…Philips Medical Systems Nederland B.V. · 2023 · Recalled
  6. This deviceK251397 Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR…Philips Medical Systems Nederland B.V. · 2025
  7. Descendant · cites it directly · 0 of its ownK253648 Ingenia, Ingenia CX, Ingenia Elition S/X, Ingenia Ambition S/X…Philips Medical Systems Nederland B.V. · 2026
Show all 12 generations
  1. Root · 12 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
  2. 11 generations backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
  3. 10 generations backK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
  4. 9 generations backK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
  5. 8 generations backK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
  6. 7 generations backK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
  7. 6 generations backK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
  8. 5 generations backK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
  9. 4 generations backK193215 Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia…Philips Medical Systems Nederland B.V. · 2020 · Recalled
  10. 3 generations backK212673 MR 5300Philips Medical Systems Nederland B.V. · 2021 · Recalled
  11. 2 generations backK223442 MR 5300 and MR 7700 R11 MR SystemsPhilips Medical Systems Nederlands B.V. · 2022 · Recalled
  12. 1 generation backK223458 Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and…Philips Medical Systems Nederland B.V. · 2023 · Recalled
  13. This deviceK251397 Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR…Philips Medical Systems Nederland B.V. · 2025

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K253648Ingenia, Ingenia CX, Ingenia Elition S/X, Ingenia Ambition S/X, BlueSeal SE/XE/QE, MR…Philips Medical Systems Nederland B.V. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K031815Intera-Sensation FamilyPhilips Medical Systems20033
−11K041602Panorama 1.0TPhilips Medical Systems North America, Inc.2004Class I
−10K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V.2005Class I
−9K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems2007Class I
−8K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care201120
−7K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V.20177
−6K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V.201811
−5K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.201925
−4K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR…Philips Medical Systems Nederland B.V.2020152
−3K212673MR 5300Philips Medical Systems Nederland B.V.202117
−2K223442MR 5300 and MR 7700 R11 MR SystemsPhilips Medical Systems Nederlands B.V.20225
−1K223458Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR…Philips Medical Systems Nederland B.V.202317
0 (this device)K251397Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR 5300 and…Philips Medical Systems Nederland B.V.2025—
+1K253648Ingenia, Ingenia CX, Ingenia Elition S/X, Ingenia Ambition S/X, BlueSeal…Philips Medical Systems Nederland B.V.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200511
201510
201610
202220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K251397, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K223458Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR Systems
Summary page 5
…Ingenia CX, Ingenia Elition, Ingenia Ambition, MR 5300 and MR 7700 MR System 510(k) Clearance: K223458 Manufacturer: Philips Medical Systems Nederland B.V. Classification Name:…
Philips Medical Systems Nederland B.V.202317

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K253648Ingenia, Ingenia CX, Ingenia Elition S/X, Ingenia Ambition S/X, BlueSeal SE/XE/QE, MR…Philips Medical Systems Nederland B.V.20260

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 28, 2026Class IIZ-3313-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3307-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3312-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3311-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3310-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3315-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Aug 28, 2026Class IIZ-3318-2026Software designWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party…
Apr 14, 2026Class IIZ-1968-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1958-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1969-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1954-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1961-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1965-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1964-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1962-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · 6 generations back · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
  • K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 7 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · 7 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 22 years and passes through 5 companies; a new generation was cleared every 1.2 years on median.
  • 17 of the 19 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K253648 (Philips Medical Systems Nederland B.V., 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the…

Read the full statement (282 words)

Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device.

This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin.

Image appearance is determined by many different physical properties of the tissue and the anatomy, the MR scan technique applied, and presence of contrast agents. The use of contrast agents for diagnostic imaging applications should be performed consistent with the approved labeling for the contrast agent.

The trained clinical user can adjust the MR scan parameters to customize image appearance, accelerate image acquisition, and synchronize with the patient's breathing or cardiac cycle.

The systems can use combinations of images to produce physical parameters, and related derived images. Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user.

In addition, the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities. MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated bv the clinical user.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251397: Ingenia, Ingenia CX, Ingenia Elition, Ingenia Ambition, BlueSeal, MR 5300 and MR 7700 MR System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K251397

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026