Recalled · Class II · 2026Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ingenia Ambition S; Ingenia Ambition X; and MR 5300 MR Systems
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Jul 11, 2025 as substantially equivalent to K212673 (Philips Medical Systems Nederland B.V., 2021) and 1 other predicate after a 29-day review. Its family tree reaches back 10 generations to 2003. 4 of the 16 devices in its family, counting itself, have a Class I recall. It has been recalled 19 times.
- Decided
- Jul 11, 2025
- Review
- 29 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 19
- Predicates
- 2
Its longest line of descent runs back 10 generations to K031815 (2003). 15 ancestors in all, the oldest K001796 (2000). No later device cites it as a predicate yet. 4 of the 16 devices in this family, counting this one, have a Class I recall.
- Root · 10 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 6 generations between
- 3 generations backK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
- 2 generations backK193215 Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia…Philips Medical Systems Nederland B.V. · 2020 · Recalled
- 1 generation backK212673 MR 5300Philips Medical Systems Nederland B.V. · 2021 · Recalled
- This deviceK251808 Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX…Philips Medical Systems Nederland B.V. · 2025
Show all 10 generations
- Root · 10 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 9 generations backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
- 8 generations backK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
- 7 generations backK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
- 6 generations backK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
- 5 generations backK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
- 4 generations backK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
- 3 generations backK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
- 2 generations backK193215 Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia…Philips Medical Systems Nederland B.V. · 2020 · Recalled
- 1 generation backK212673 MR 5300Philips Medical Systems Nederland B.V. · 2021 · Recalled
- This deviceK251808 Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX…Philips Medical Systems Nederland B.V. · 2025
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −10 | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | 3 |
| −9 | K041602 | Panorama 1.0T | Philips Medical Systems North America, Inc. | 2004 | Class I |
| −8 | K052078 | Achieva, Intera and Panorama 1.0 Release 2-Series | Philips Medical Systems Nederland B.V. | 2005 | Class I |
| −7 | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5 | Philips Medical Systems | 2007 | Class I |
| −6 | K110151 | Achieva R4 1.5T and Achieva R4 3.0T | Philips Health Care | 2011 | 20 |
| −5 | K163116 | Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 | Philips Medical Systems Nederland B.V. | 2017 | 7 |
| −4 | K173079 | Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 | Philips Medical Systems Nederland B.V. | 2018 | 11 |
| −3 | K183063 | Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems | Philips Medical Systems Nederland B.V. | 2019 | 25 |
| −2 | K193215 | Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR… | Philips Medical Systems Nederland B.V. | 2020 | 152 |
| −1 | K212673 | MR 5300 | Philips Medical Systems Nederland B.V. | 2021 | 17 |
| 0 (this device) | K251808 | Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX… | Philips Medical Systems Nederland B.V. | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2005 | 1 | 1 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K212673 | MR 5300Summary page 5…LNI Code: Primary Predicate Trade Name: MR 5300 MR Systems Device 510(k) Clearance: K212673 Manufacturer: Philips Medical Systems Nederland B.V. Classific… | 2021 |
| * | K193215 | Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR SystemsSummary page 6…a, Ingenia CX, Ingenia Predicate Device Elition, and Ingenia Ambition MR Systems 510(k) Clearance: K193215 Manufacturer: Philips Medical Systems Nederland B.V. Classifica… | 2020 |
| † | K230972 | Ingenia Elition and MR 7700 MR Systems | 2023 |
| † | K232030 | Ingenia Elition R5.7.1 SP4 MR Systems | 2023 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 28, 2026 | Class II | Z-3309-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 28, 2026 | Class II | Z-3312-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 28, 2026 | Class II | Z-3311-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 28, 2026 | Class II | Z-3314-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 28, 2026 | Class II | Z-3310-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 28, 2026 | Class II | Z-3318-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 21, 2026 | Class II | Z-3243-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3238-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3235-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3237-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3239-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3241-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Apr 14, 2026 | Class II | Z-1958-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Apr 14, 2026 | Class II | Z-1959-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Apr 14, 2026 | Class II | Z-1954-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Apr 14, 2026 | Class II | Z-1961-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Apr 14, 2026 | Class II | Z-1967-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Apr 14, 2026 | Class II | Z-1963-2026 | Under Investigation by firm | The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol… |
| Dec 3, 2025 | Class II | Z-1204-2026 | Software design | The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS). |
The 5 latest of 19 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · 4 generations back · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
- K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
- 13 of the 15 ancestors have a recall of any class; 4 have a Class I recall.
- 1 other device also cites its predicate K212673.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the…
Read the full statement (282 words)
Philips Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device.
This MR system enables trained physicians to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin.
Image appearance is determined by many different physical properties of the tissue and the anatomy, the MR scan technique applied, and presence of contrast agents. The use of contrast agents for diagnostic imaging applications should be performed consistent with the approved labeling for the contrast agent.
The trained clinical user can adjust the MR scan parameters to customize image appearance, accelerate image acquisition, and synchronize with the patient’s breathing or cardiac cycle.
The systems can use combinations of images to produce physical parameters, and related derived images. Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters and must be controlled and validated by the clinical user.
In addition, the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities.
MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K251808 · printed Nov 4, 2026
Cite this page
510(k) Report. “K251808: Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ingenia Ambition S; Ingenia Ambition X; and MR 5300 MR Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K251808