510(k) Report

510(k) · Cleared

Recalled · 2011

Flo-Gard (R) 6201

K915522510(k) TraditionalCleared Feb 28, 199280-day reviewClass II21 CFR 880.5725Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · FRN Pump, Infusion

Cleared on Feb 28, 1992 after an 80-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 6 of the 22 devices in its family, counting itself, have a Class I recall. 21 later devices descend from it. It has been recalled 6 times.

Decided
Feb 28, 1992
Review
80 days
Class
Class II
Product code
FRN
Recalls
6
Predicates
0

21 later devices descend from it, 5 citing it directly. 6 of the 22 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1997: 1 descendant cleared2000: 1 descendant cleared2003: 1 descendant cleared1996: 1 descendant cleared2007: 1 descendant cleared2006: 1 descendant cleared, 1 with a Class I recall12014: 2 descendants cleared, 1 with a Class I recall12020: 3 descendants cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12009: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12022: 3 descendants cleared2016: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant cleared2021: 1 descendant cleared1990200020102020
  1. This deviceK915522 Flo-Gard (R) 6201Baxter Healthcare Corp · 1992
  2. Descendant · cites it directly · 16 of its ownK972678 Millennium CRT Infusion Pump and Millennium CRT Infusion Pump…Mcgaw, Inc. · 1997
  3. Descendant · 2 generations on · 15 of its ownK062700 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Recalled (Class I)
  4. Descendant · 3 generations on · 6 of its ownK140838 IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014
  5. Descendant · 3 generations on · 4 of its ownK083689 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · Recalled (Class I)
  6. Descendant · 4 generations on · 3 of its ownK170595 IV Administration SetsB.Braun Medical, Inc. · 2017 · Recalled

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

1996: 11997: 12000: 12003: 12006: 12007: 12009: 12013: 12014: 22015: 12016: 12017: 12020: 32021: 12022: 32025: 1
1996high 3 (2020) · low 0 (1998)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 21 · See all
  • K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration SetsMcgaw, Inc. · 1997 · cites it directly16descendants
  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on15descendants
  • K140838IV Administration Set with Hand PumpB.Braun Medical, Inc. · 2014 · 3 generations on6descendants
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · 3 generations on4descendants
  • K170595IV Administration SetsB.Braun Medical, Inc. · 2017 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K915522Flo-Gard (R) 6201Baxter Healthcare Corp1992—
+1K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration…Mcgaw, Inc.1997—
+2K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2006Class I
+3K140838IV Administration Set with Hand PumpB.Braun Medical, Inc.2014—
+3K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2009Class I
+4K170595IV Administration SetsB.Braun Medical, Inc.201719

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199610
199710
200010
200310
200611
200710
200911
201311
201421
201511
201610
201710
202031
202110
202230
202510
FIG. 1 — Lineage of K915522, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (21)

See all 21

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K972678Millennium CRT Infusion Pump and Millennium CRT Infusion Pump Administration SetsMcgaw, Inc.199716
K070529Aff (Mark I) Volumetric Infusion PumpDefinitive Medical Technologies, Ltd.20070
K032789Replacement Pump Head Door Assembly, P1, Model BX10468, Replacement Pump Head Door…American I.V. Products, Inc.20030
K993524Advantis Asc Infusion Pump and Administration SetsAlaris Medical Systems, Inc.20000
K961703Colleague 2 Volumetric Infusion Pump (Dual Channel)/Colleague 3 Volumetric Infusion Pump…Baxter Healthcare Corp19960

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 24, 2011Not classifiedZ-2122-2011Employee errorPumps submitted for service may not have been properly serviced before being returned to the customer.
Nov 17, 2010Not classifiedZ-0933-2011Process controlThe FloGard Infusion Pump was released from service without having the Current Leakage test performed. This could lead to the patient/user to receive an electrical shock.
Jun 21, 2007Not classifiedZ-1111-2007OtherFalsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
Nov 27, 2006Not classifiedZ-0632-2007OtherThere is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
May 3, 2005Not classifiedZ-1552-05OtherThe mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.
May 3, 2005Not classifiedZ-1553-05OtherThe mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K142036Infusomat Space Volumetric Infusion Pump Administration SetsB.Braun Medical, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · 4 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K972678 (Mcgaw, Inc., 1997) has gone furthest, with 16 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—
Cite this page

510(k) Report. “K915522: Flo-Gard (R) 6201.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K915522

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026