510(k) Report

510(k) · Cleared

Recalled · 2008

LTV 1000 Ventilator

K981371510(k) TraditionalCleared Oct 30, 1998197-day reviewClass II21 CFR 868.5895Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pulmonetic Systems, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Oct 30, 1998 as substantially equivalent to K950484 (Bird Products Corp., 1996) after a 197-day review. Its family tree reaches back 1 generation to 1996. 21 of the 67 devices in its family, counting itself, have a Class I recall. 65 later devices descend from it. It has been recalled 12 times.

Decided
Oct 30, 1998
Review
197 days
Class
Class II
Product code
CBK
Recalls
12
Predicates
1

Its longest line of descent runs back 1 generation to K950484 (1996). 65 later devices descend from it, 6 citing it directly. 21 of the 67 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared, 1 with a Class I recall12011: 7 descendants cleared, 2 with a Class I recall22009: 2 descendants cleared, 1 with a Class I recall12008: 3 descendants cleared, 2 with a Class I recall22004: 1 descendant cleared2006: 2 descendants cleared, 1 with a Class I recall12017: 4 descendants cleared, 2 with a Class I recall22010: 3 descendants cleared2012: 12 descendants cleared, 4 with a Class I recall42021: 3 descendants cleared, 2 with a Class I recall22015: 2 descendants cleared, 1 with a Class I recall12024: 4 descendants cleared2019: 3 descendants cleared, 2 with a Class I recall22014: 3 descendants cleared2016: 5 descendants cleared, 1 with a Class I recall12023: 1 descendant cleared2025: 2 descendants cleared2026: 3 descendants cleared2013: 2 descendants cleared, 1 with a Class I recall12022: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall1Reference device: K962517 · Model 7700 Quantum Pressure Support Ventilator PSV · Healthdyne Technologies, Inc. · 1997Reference device: K811102 · Siemens-Elema AB Servo Ventilator 900 C · Siemens Elema AB · 1981Reference device: K905244 · Model LP6 Ventilator, Modification · Aequitron Medical, Inc. · 1990K950484 · Tbird VS Ventilator and Tbird Avs Ventilator · Bird Products Corp. · 19961995200020052010201520202025
Descendants, by generation 1 generation forward: 6 descendants, 2 with a Class I recall22 generations forward: 9 descendants, 4 with a Class I recall43 generations forward: 11 descendants, 5 with a Class I recall54 generations forward: 16 descendants, 5 with a Class I recall55 generations forward: 11 descendants, 3 with a Class I recall36 generations forward: 6 descendants7 generations forward: 3 descendants8 generations forward: 2 descendants, 2 with a Class I recall29 generations forward: 1 descendant K950484 · Tbird VS Ventilator and Tbird Avs Ventilator · Bird Products Corp. · 19960+5
  1. Root · 1 generation backK950484 Tbird VS Ventilator and Tbird Avs VentilatorBird Products Corp. · 1996
  2. This deviceK981371 LTV 1000 VentilatorPulmonetic Systems, Inc. · 1998
  3. Descendant · cites it directly · 44 of its ownK040790 Modification to LTV 1000 Ventilator/Breathing CircuitsPulmonetic Systems, Inc. · 2004
  4. Descendant · 2 generations on · 22 of its ownK060705 Carina HomeDraeger Medical B.V. · 2006 · Recalled
  5. Descendant · 3 generations on · 20 of its ownK103167 Stellar 150Resmed Germany, Inc. · 2011
  6. Descendant · 2 generations on · 20 of its ownK062267 Oxylog 3000Drager Medical B.V. · 2006 · Recalled (Class I)
  7. Descendant · 3 generations on · 13 of its ownK112006 Hamilton-TC1Hamilton Medical AG · 2012

Showing 5 of 65 descendants. See all descendants

Descendants cleared per year

2004: 12006: 22007: 12008: 32009: 22010: 32011: 72012: 122013: 22014: 32015: 22016: 52017: 42019: 32020: 12021: 32022: 12023: 12024: 42025: 22026: 3
2004high 12 (2012) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 65 · See all
  • K040790Modification to LTV 1000 Ventilator/Breathing CircuitsPulmonetic Systems, Inc. · 2004 · cites it directly44descendants
  • K060705Carina HomeDraeger Medical B.V. · 2006 · 2 generations on22descendants
  • K103167Stellar 150Resmed Germany, Inc. · 2011 · 3 generations on20descendants
  • K062267Oxylog 3000Drager Medical B.V. · 2006 · 2 generations on20descendants
  • K112006Hamilton-TC1Hamilton Medical AG · 2012 · 3 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K950484Tbird VS Ventilator and Tbird Avs VentilatorBird Products Corp.1996—
0 (this device)K981371LTV 1000 VentilatorPulmonetic Systems, Inc.1998—
+1K040790Modification to LTV 1000 Ventilator/Breathing CircuitsPulmonetic Systems, Inc.2004—
+2K060705Carina HomeDraeger Medical B.V.20061
+3K103167Stellar 150Resmed Germany, Inc.2011—
+2K062267Oxylog 3000Drager Medical B.V.2006Class I
+3K112006Hamilton-TC1Hamilton Medical AG2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200621
200711
200832
200921
201030
201172
2012124
201321
201430
201521
201651
201742
201932
202011
202132
202210
202310
202440
202520
202630
FIG. 1 — Lineage of K981371, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

21 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K950484Tbird VS Ventilator and Tbird Avs Ventilator
Summary page 1
…de Name: LTV 1000 Common Name Ventilator Classification Name: Ventilator, Continuous (Respirator) PREDICATE DEVICE The primary predicate device is the TBird Ventilator (K950484), manufactured by Bird Products Corp. of Palm Springs, CA. DESCRIPTION The LTV 1000 ventilator is intended to provide continuou…
Bird Products Corp.1996—
†K962517Model 7700 Quantum Pressure Support Ventilator PSVHealthdyne Technologies, Inc.1997—
†K811102Siemens-Elema AB Servo Ventilator 900 CSiemens Elema AB1981—
†K905244Model LP6 Ventilator, ModificationAequitron Medical, Inc.1990—

Also cleared against K950484: 4 other devices →

What descends from it? (65)

See all 65

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040790Modification to LTV 1000 Ventilator/Breathing CircuitsPulmonetic Systems, Inc.200444
K070594Palmtop Ventilator, Models PTV-8, PTV-10 Class IPulmonetic Systems, Inc.200712
K082724Newport HTS50 Ventilator with Dual Pac Battery SystemNewport Medical Instruments, Inc.20089
K082982Breathe Technologies Ventilator and AccessoriesBreathe Technologies, Inc.20095
K090888Newport HT70 Family of Ventilators Class INewport Medical Instruments, Inc.20092
K101643Model LTV 1100 VentilatorCare Fusion20111

Has anything in its family been recalled? (21 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 22, 2008Not classifiedZ-1463-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1469-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1468-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1467-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1464-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1470-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1466-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Feb 22, 2008Not classifiedZ-1465-2008Device DesignThe failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver.
Nov 9, 2005Not classifiedZ-0184-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0186-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Nov 9, 2005Not classifiedZ-0185-06OtherPulmonetic Systems has identified the potential of a printed circuit board malfunction, which can result in a ventilator malfunctio (e.g. Vent INOP, HW Fault, Xducer Fault) and possibly resulting in failure of the ventilator to breathe…
Sep 8, 2004Not classifiedZ-1485-04OtherVentilators have not consistently switched to internal battery operation when the external DC power source has become inadequate to supply stable power. This situation results in loss of ventilation to the patient.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K062267Oxylog 3000Drager Medical B.V. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 13 Class I recalls
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K111146Newport HT70 VentilatorNewport Medical Instruments, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K072885Carina System, Model 8419350; Carina Basic Device, Model 5704110Draeger Medical AG & Co. KG · 2008 · 3 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K082600Oxylog 2000 Plus System (Model 5705080), Oxylog 2000 Plus Basic Device (Model 5705082)Draeger Medical B.V. · 2008 · 3 generations on · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
  • K103625Oxylog 3000 Plus Workstation, Oxylog 3000 Plus Basic DeviceDraeger Medical Systems, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 14 Class I recalls
  • K141595O-Two E700,O-Two E600,O-Two E500O-Two Medical Technologies, Inc. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 13 Class I recalls
  • K181166Trilogy EvoRespironics, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 131 Class I recalls
  • K113053Bipap a 30 Ventilatory Support SystemRespironics, Inc. · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 15 Class I recalls

Nearest 12 of 21 shown.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K040790 (Pulmonetic Systems, Inc., 2004) has gone furthest, with 44 descendants of its own.
  • 4 other devices also cite its predicate K950484.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 1bs), who require the following types of ventilatory support:

Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).

Assist/Control, SIMV, or CPAP modes of ventilation.

The ventilator is suitable for use in institutional and transport settings.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K981371: LTV 1000 Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K981371

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.