510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Gemini GXL

K051170510(k) TraditionalCleared May 23, 200517-day reviewClass II21 CFR 892.1200Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · KPS System, Tomography, Computed, Emission

Cleared on May 23, 2005 as substantially equivalent to K042935 (Philips Medical Systems (Cleveland), Inc., 2004) and 1 other predicate after a 17-day review. Its family tree reaches back 7 generations to 1993. 13 later devices descend from it. It has been recalled 12 times.

Decided
May 23, 2005
Review
17 days
Class
Class II
Product code
KPS
Recalls
12
Predicates
2

Its longest line of descent runs back 7 generations to K930090 (1993). 20 ancestors in all. 13 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 3 other ancestors2001: 3 other ancestors2004: 1 other ancestor2000: 2 other ancestors1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2005: 1 descendant cleared2017: 2 descendants cleared2008: 2 descendants cleared2013: 1 descendant cleared2011: 1 descendant cleared2014: 1 descendant cleared2025: 1 descendant cleared2021: 2 descendants cleared2018: 1 descendant cleared2023: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalled×2: 2 generations cleared close together, K041955 (2004), K042935 (2004) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 6 other ancestors5 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 6 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants K042935 · Gemini GXL · Philips Medical Systems (Cleveland), Inc. · 2004 · recalledK041955 · Gemini Pet/CT Imaging System · Philips Medical Systems (Cleveland), Inc. · 2004 · recalledK032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50
  1. Root · 7 generations backK930090 HelicatElscint, Inc. · 1993
  2. 6 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 5 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 4 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 3 generations backK032036 Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · Recalled
  6. 2 generations backK041955 Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  7. 1 generation backK042935 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  8. This deviceK051170 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2005
  9. Descendant · cites it directly · 12 of its ownK052640 Gemini RaptorPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  10. Descendant · 2 generations on · 4 of its ownK123599 Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013 · Recalled
  11. Descendant · 3 generations on · 2 of its ownK210880 Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021
  12. Descendant · 2 generations on · 2 of its ownK140651 Celesteion, Pca-9000A/2Toshibamedical Systems Corporation · 2014 · Recalled
  13. Descendant · 3 generations on · 1 of its ownK170490 Celesteion, PCA-9000A/3Toshibamedical Systems Corporation · 2017

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2005: 12008: 22011: 12013: 12014: 12017: 22018: 12021: 22023: 12025: 1
2005high 2 (2008) · low 0 (2006)2026

Most influential descendants

Showing 5 of 13 · See all
  • K052640Gemini RaptorPhilips Medical Systems (Cleveland), Inc. · 2005 · cites it directly12descendants
  • K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013 · 2 generations on4descendants
  • K210880Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021 · 3 generations on2descendants
  • K140651Celesteion, Pca-9000A/2Toshibamedical Systems Corporation · 2014 · 2 generations on2descendants
  • K170490Celesteion, PCA-9000A/3Toshibamedical Systems Corporation · 2017 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K930090HelicatElscint, Inc.1993—
−6K961464Select/SPElscint, Inc.19963
−5K982060Volumax CT Imaging SystemElscint, Inc.1998—
−4K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−3K032036Gemini 16Philips Medical Systems (Cleveland), Inc.20032
−2K041955Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc.20042
−1K042935Gemini GXLPhilips Medical Systems (Cleveland), Inc.20042
0 (this device)K051170Gemini GXLPhilips Medical Systems (Cleveland), Inc.2005—
+1K052640Gemini RaptorPhilips Medical Systems (Cleveland), Inc.200536
+2K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland)20135
+3K210880Vereos PET/CTPhilips Medical Systems Nederland B.V.2021—
+2K140651Celesteion, Pca-9000A/2Toshibamedical Systems Corporation20142
+3K170490Celesteion, PCA-9000A/3Toshibamedical Systems Corporation2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200020
200130
200330
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200820
201110
201310
201410
201720
201810
202120
202310
202510
FIG. 1 — Lineage of K051170, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042935Gemini GXL
Summary page 1
…Code: 90 KPS and 90 JAK Classification Panel: Radiology C. Date prepared: April 8, 2005 D. Predicate Device: Gemini System (K042935) "Philips Plus" CT Scanner (K033326) E_ Performance Standards: * 21 CER 1020.30 - 1020.33…
Philips Medical Systems (Cleveland), Inc.20042
*K033326Philips Plus Computed Tomography System
Summary page 1
…iology C. Date prepared: April 8, 2005 D. Predicate Device: Gemini System (K042935) "Philips Plus" CT Scanner (K033326) E_ Performance Standards: * 21 CER 1020.30 - 1020.33 Performance Standards for Ionizing Radiation Emitting P…
Philips Medical Systems, Inc.2003131

What descends from it? (13)

See all 13

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052640Gemini RaptorPhilips Medical Systems (Cleveland), Inc.200512

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-0280-2022Device DesignSignificant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in…
May 1, 2021Class IIZ-0278-2022Device DesignSignificant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in…
Jul 1, 2015Class IIZ-2193-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Mar 6, 2015Class IIZ-1781-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Aug 8, 2012Class IIZ-2239-2012Material/Component ContaminationPhilips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS…
Jan 26, 2012Not classifiedZ-1368-2012Software design (manufacturing process)TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving. In TumorLOC, when saving while a contour is in edit mode, that…
Feb 19, 2009Not classifiedZ-1579-2009Software designA software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the…
Jan 26, 2009Not classifiedZ-0260-2010Process change controlUpper patient pallet component of the recalling firm's Gemini PET/CT system may be thinner than normally required, which may reduce the product's load bearing capacity, resulting in cracking under less than the designed full-load weight.
Nov 24, 2008Not classifiedZ-0286-2010Process designA leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
May 1, 2008Class IIZ-0197-2016Unknown/Undetermined by firmThe reason for this inspection is to determine if the Spindle Block Servo Controllers on the CT subsystem of potentially affected GEMINI PET/CT systems are susceptible to premature failure.
Sep 21, 2007Not classifiedZ-0462-2008Software designArtifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.
Sep 21, 2007Not classifiedZ-0461-2008Software designArtifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • 11 of the 20 ancestors have a recall of any class.
  • Of its descendants, K052640 (Philips Medical Systems (Cleveland), Inc., 2005) has gone furthest, with 12 descendants of its own.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, head, heart and…

Read the full statement (259 words)

The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, head, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions.

Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images nray be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in:

The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders.

The detection, localization, and staging of tumors and diagnosing cancer patients.

Treatment planning and interventional radiology procedures.

Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium,

Both subsystems (PET and CT) can also be operated independently as fully functional, diagnostic imaging systems including application of the CT scanner as a radiation therapy simulation scanner.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051170: Gemini GXL.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051170

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026