510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Gemini Raptor

K052640510(k) TraditionalCleared Oct 7, 200511-day reviewClass II21 CFR 892.1200Recalls 36

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · KPS System, Tomography, Computed, Emission

Cleared on Oct 7, 2005 as substantially equivalent to K051170 (Philips Medical Systems (Cleveland), Inc., 2005) after an 11-day review. Its family tree reaches back 8 generations to 1993. 12 later devices descend from it. It has been recalled 36 times.

Decided
Oct 7, 2005
Review
11 days
Class
Class II
Product code
KPS
Recalls
36
Predicates
1

Its longest line of descent runs back 8 generations to K930090 (1993). 21 ancestors in all. 12 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 3 other ancestors2001: 3 other ancestors2004: 1 other ancestor2000: 2 other ancestors1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2017: 2 descendants cleared2008: 2 descendants cleared2013: 1 descendant cleared2011: 1 descendant cleared2014: 1 descendant cleared2025: 1 descendant cleared2021: 2 descendants cleared2018: 1 descendant cleared2023: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalled×2: 2 generations cleared close together, K041955 (2004), K042935 (2004) K051170 · Gemini GXL · Philips Medical Systems (Cleveland), Inc. · 2005 · recalled1990200020102020
  1. Root · 8 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 6 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 5 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK032036 Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · Recalled
  6. 3 generations backK041955 Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  7. 2 generations backK042935 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  8. 1 generation backK051170 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  9. This deviceK052640 Gemini RaptorPhilips Medical Systems (Cleveland), Inc. · 2005
  10. Descendant · cites it directly · 4 of its ownK123599 Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013 · Recalled
  11. Descendant · 2 generations on · 2 of its ownK210880 Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021
  12. Descendant · cites it directly · 2 of its ownK140651 Celesteion, Pca-9000A/2Toshibamedical Systems Corporation · 2014 · Recalled
  13. Descendant · 2 generations on · 1 of its ownK170490 Celesteion, PCA-9000A/3Toshibamedical Systems Corporation · 2017
  14. Descendant · 2 generations on · 0 of its ownK251401 PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania · 2025

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2008: 22011: 12013: 12014: 12017: 22018: 12021: 22023: 12025: 1
2008high 2 (2008) · low 0 (2009)2026

Most influential descendants

Showing 5 of 12 · See all
  • K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013 · cites it directly4descendants
  • K210880Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021 · 2 generations on2descendants
  • K140651Celesteion, Pca-9000A/2Toshibamedical Systems Corporation · 2014 · cites it directly2descendants
  • K170490Celesteion, PCA-9000A/3Toshibamedical Systems Corporation · 2017 · 2 generations on1descendant
  • K251401PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K930090HelicatElscint, Inc.1993—
−7K961464Select/SPElscint, Inc.19963
−6K982060Volumax CT Imaging SystemElscint, Inc.1998—
−5K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−4K032036Gemini 16Philips Medical Systems (Cleveland), Inc.20032
−3K041955Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc.20042
−2K042935Gemini GXLPhilips Medical Systems (Cleveland), Inc.20042
−1K051170Gemini GXLPhilips Medical Systems (Cleveland), Inc.200512
0 (this device)K052640Gemini RaptorPhilips Medical Systems (Cleveland), Inc.2005—
+1K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland)20135
+2K210880Vereos PET/CTPhilips Medical Systems Nederland B.V.2021—
+1K140651Celesteion, Pca-9000A/2Toshibamedical Systems Corporation20142
+2K170490Celesteion, PCA-9000A/3Toshibamedical Systems Corporation2017—
+2K251401PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200020
200130
200330
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200820
201110
201310
201410
201720
201810
202120
202310
202510
FIG. 1 — Lineage of K052640, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051170Gemini GXLPhilips Medical Systems (Cleveland), Inc.200512

What descends from it? (12)

See all 12

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland)20134
K140651Celesteion, Pca-9000A/2Toshibamedical Systems Corporation20142
K172406Ingenuity TFPhilips Medical Systems (Cleveland), Inc.20170
K103483Ingenuity TF Pet/MRPhilips Healthcare (Cleveland)20110
K081426Nexstar Liftoff Pet Application Software SuitePhilips Medical Systems (Cleveland), Inc.20080
K081135Gemini CondorPhilips Medical Systems (Cleveland), Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-0279-2022Device DesignSignificant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in…
May 1, 2021Class IIZ-0281-2022Device DesignSignificant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in…
May 1, 2021Class IIZ-1987-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Mar 28, 2018Class IIZ-2276-2018OtherA problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Feb 15, 2018Class IIZ-1392-2018Under Investigation by firmFollowing a period of inactivity, the mass storage device may cause the acquisition console to become unresponsive. This may prevent an acquisition from proceeding.
Nov 1, 2017Class IIZ-0522-2018Vendor change controlOne of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
Jul 1, 2015Class IIZ-2194-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Jul 1, 2015Class IIZ-2196-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Jul 1, 2015Class IIZ-2197-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Jul 1, 2015Class IIZ-2195-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Jul 1, 2015Class IIZ-2198-2015Software designPhilips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Mar 6, 2015Class IIZ-1787-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1785-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1783-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1788-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1782-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1786-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Jun 18, 2014Class IIZ-2037-2014Software designWhen beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not…
Dec 4, 2013Class IIZ-0932-2014Software designThe PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.
Aug 16, 2013Class IIZ-0067-2014Software designIssues with the Ingenuity TF PET/CT Software Version 4.0 could pose a risk for patients in the event that a patient rescan were to require reinjection of PET radiopharmaceutical.
Jan 21, 2013Class IIZ-0897-2013Software designThe system may not default to the same gating trigger with which the data were acquired. This could cause a risk for patients.
Dec 20, 2012Class IIZ-1029-2013Software designIssues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, including inaccurate standard uptake values…
Oct 24, 2012Class IIZ-0882-2013Software designSoftware anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was…
Aug 8, 2012Class IIZ-2239-2012Material/Component ContaminationPhilips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS…
Jan 26, 2012Not classifiedZ-1369-2012Software design (manufacturing process)TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving. In TumorLOC, when saving while a contour is in edit mode, that…
Dec 13, 2011Not classifiedZ-1056-2012Software designA number of non-conformances have been noted in the TruFlight Select PET/CT System. A number of these non-conformances have been corrected to improve the overall quality and functionality of the suspect system units.
Dec 2, 2011Not classifiedZ-0805-2012Software designPhilips received notification by a customer of an issue relating to unwanted auto delete occurrences resulting in raw data being deleted and, in some cases, raw data being deleted immediately after scan acquisition but before images are…
Oct 24, 2011Not classifiedZ-0937-2012Material/Component ContaminationPhilips became aware of an issue relating to a UPS cabinet that contained a melted battery which caused damage to other batteries lower in the cabinet, leading to battery acid leaking on the floor.
May 31, 2011Not classifiedZ-3200-2011Software designThe firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact the performance of reliability of the…
May 31, 2011Not classifiedZ-3199-2011Software designThe firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact the performance of reliability of the…
Feb 19, 2009Not classifiedZ-1580-2009Software designA software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the…
Jan 26, 2009Not classifiedZ-0261-2010Process change controlUpper patient pallet component of the recalling firm's Gemini PET/CT system may be thinner than normally required which may reduce the product's load bearing capacity, resulting in cracking under less than the designed full-load weight.
Nov 24, 2008Not classifiedZ-0287-2010Process designA leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
May 1, 2008Class IIZ-0197-2016Unknown/Undetermined by firmThe reason for this inspection is to determine if the Spindle Block Servo Controllers on the CT subsystem of potentially affected GEMINI PET/CT systems are susceptible to premature failure.
Sep 21, 2007Not classifiedZ-0462-2008Software designArtifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.
Sep 21, 2007Not classifiedZ-0461-2008Software designArtifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • 12 of the 21 ancestors have a recall of any class.
  • Of its descendants, K123599 (Philips Healthcare (Cleveland), 2013) has gone furthest, with 4 descendants of its own.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart…

Read the full statement (278 words)

The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions.

Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in:

o The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders.

o The detection, localization, and staging of tumors and diagnosing cancer patients.

o Treatment planning and interventional radiology procedures.

The device includes software that provides a quantified analysis of regional cerebral activity from PET images.

Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium.

Both subsystems (PET and CT) can also be operated independently as fully functional, diagnostic imaging systems including application of the CT scanner as a radiation therapy simulation scanner.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052640: Gemini Raptor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052640

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026