510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Ingenuity Digital Pet/CT

K123599510(k) TraditionalCleared Jul 12, 2013233-day reviewClass II21 CFR 892.1200Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Healthcare (Cleveland) · KPS System, Tomography, Computed, Emission

Cleared on Jul 12, 2013 as substantially equivalent to K052640 (Philips Medical Systems (Cleveland), Inc., 2005) after a 233-day review. Its family tree reaches back 9 generations to 1993. 4 later devices descend from it. It has been recalled 5 times.

Decided
Jul 12, 2013
Review
233 days
Class
Class II
Product code
KPS
Recalls
5
Predicates
1

Its longest line of descent runs back 9 generations to K930090 (1993). 22 ancestors in all. 4 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 3 other ancestors2001: 3 other ancestors2004: 1 other ancestor2000: 2 other ancestors1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2025: 1 descendant cleared2021: 2 descendants cleared2023: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalled×2: 2 generations cleared close together, K041955 (2004), K042935 (2004) ×2: 2 generations cleared close together, K051170 (2005), K052640 (2005) 1990200020102020
  1. Root · 9 generations backK930090 HelicatElscint, Inc. · 1993
  2. 5 generations between
  3. 3 generations backK042935 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  4. 2 generations backK051170 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  5. 1 generation backK052640 Gemini RaptorPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  6. This deviceK123599 Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013
  7. Descendant · cites it directly · 2 of its ownK210880 Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021
  8. Descendant · cites it directly · 0 of its ownK251401 PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania · 2025
  9. Descendant · 2 generations on · 0 of its ownK231420 Aura 10 PET/CTXeos Medical · 2023
  10. Descendant · 2 generations on · 0 of its ownK211764 Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021
Show all 9 generations
  1. Root · 9 generations backK930090 HelicatElscint, Inc. · 1993
  2. 8 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 7 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 6 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 5 generations backK032036 Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · Recalled
  6. 4 generations backK041955 Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  7. 3 generations backK042935 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  8. 2 generations backK051170 Gemini GXLPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  9. 1 generation backK052640 Gemini RaptorPhilips Medical Systems (Cleveland), Inc. · 2005 · Recalled
  10. This deviceK123599 Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland) · 2013

Most influential descendants

Showing 4 of 4 · See all
  • K210880Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021 · cites it directly2descendants
  • K251401PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania · 2025 · cites it directly0descendants
  • K231420Aura 10 PET/CTXeos Medical · 2023 · 2 generations on0descendants
  • K211764Vereos PET/CTPhilips Medical Systems Nederland B.V. · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K930090HelicatElscint, Inc.1993—
−8K961464Select/SPElscint, Inc.19963
−7K982060Volumax CT Imaging SystemElscint, Inc.1998—
−6K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−5K032036Gemini 16Philips Medical Systems (Cleveland), Inc.20032
−4K041955Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc.20042
−3K042935Gemini GXLPhilips Medical Systems (Cleveland), Inc.20042
−2K051170Gemini GXLPhilips Medical Systems (Cleveland), Inc.200512
−1K052640Gemini RaptorPhilips Medical Systems (Cleveland), Inc.200536
0 (this device)K123599Ingenuity Digital Pet/CTPhilips Healthcare (Cleveland)2013—
+1K210880Vereos PET/CTPhilips Medical Systems Nederland B.V.2021—
+1K251401PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania2025—
+2K231420Aura 10 PET/CTXeos Medical2023—
+2K211764Vereos PET/CTPhilips Medical Systems Nederland B.V.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200020
200130
200330
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202120
202310
202510
FIG. 1 — Lineage of K123599, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052640Gemini Raptor
Summary page 2
…puted Tomography, 21 CIFR 892.1750 Product Code: 9OKPS ___________________ 90JAK Predicate Device(s): K052640, Gemini Raptor Device Description: The device is a hybrid diagnostic imaging system that combines…
Philips Medical Systems (Cleveland), Inc.200536

Also cleared against K052640: 5 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210880Vereos PET/CTPhilips Medical Systems Nederland B.V.20212
K251401PennPET Explorer Positron Emission TomographTrustees of the University of Pennsylvania20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-1988-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Jun 24, 2019Class IIZ-2194-2019OtherIn certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Jun 3, 2019Class IIZ-1929-2019Device DesignThe Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.
Sep 4, 2018Class IIZ-0302-2019Software designSoftware Correction: Software errors that may result in incorrect 4D CT images, PET images with SUV quantification error, or disruption of interventional CT procedures; and Software errors that may result in partial set of images…
May 21, 2018Class IIZ-2496-2018Software designThe block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 4 companies; a new generation was cleared every 1.8 years on median.
  • 13 of the 22 ancestors have a recall of any class.
  • Of its descendants, K210880 (Philips Medical Systems Nederland B.V., 2021) has gone furthest, with 2 descendants of its own.
  • 5 other devices also cite its predicate K052640.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

The Philips Ingenuity Digital PET/CT System is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of x-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purpose of detecting, localizing, diagnosing…

Read the full statement (135 words)

The Philips Ingenuity Digital PET/CT System is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of x-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purpose of detecting, localizing, diagnosing, staging, re-staging and follow-up for monitoring therapy response of various diseases in oncology, cardiology and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools including application of the CT scanner for diagnosis and for use in radiation therapy planning.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123599: Ingenuity Digital Pet/CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123599

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026