510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Alaris System PC Unit, Model 8001

K051641510(k) TraditionalCleared Oct 20, 2005122-day reviewClass II21 CFR 880.5725Recalls 26 (10 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardinal Health, Alaris Products · FRN Pump, Infusion

Cleared on Oct 20, 2005 as substantially equivalent to K950419 (Imed Corp., 1995) after a 122-day review. Its family tree reaches back 1 generation to 1995. 4 of the 8 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 26 times, including 10 Class I recalls.

Decided
Oct 20, 2005
Review
122 days
Class
Class II
Product code
FRN
Recalls
26 (10 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K950419 (1995). 6 later devices descend from it, 2 citing it directly. 4 of the 8 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared2014: 1 descendant cleared, 1 with a Class I recall12012: 1 descendant cleared2022: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK950419 · The Imed Orion Infusion Pump and Administration Sets · Imed Corp. · 1995 · Class I recall1995200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 1 descendant K950419 · The Imed Orion Infusion Pump and Administration Sets · Imed Corp. · 1995 · Class I recall0
  1. Root · 1 generation backK950419 The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · Recalled (Class I)
  2. This deviceK051641 Alaris System PC Unit, Model 8001Cardinal Health, Alaris Products · 2005
  3. Descendant · cites it directly · 5 of its ownK091308 Alaris PC Unit, Models 8000 and 8015Cardinal Health 303,Inc. · 2009
  4. Descendant · cites it directly · 3 of its ownK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Recalled (Class I)
  5. Descendant · 2 generations on · 1 of its ownK211218 BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · Recalled (Class I)
  6. Descendant · 3 generations on · 0 of its ownK243855 BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025
  7. Descendant · 2 generations on · 0 of its ownK210074 Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia…Fresenius Kabi AG · 2022

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2009: 12012: 12014: 12022: 12023: 12025: 1
2009high 1 (2009) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K091308Alaris PC Unit, Models 8000 and 8015Cardinal Health 303,Inc. · 2009 · cites it directly5descendants
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · cites it directly3descendants
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on1descendant
  • K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc. · 2025 · 3 generations on0descendants
  • K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo…Fresenius Kabi AG · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K950419The Imed Orion Infusion Pump and Administration SetsImed Corp.1995Class I
0 (this device)K051641Alaris System PC Unit, Model 8001Cardinal Health, Alaris Products2005—
+1K091308Alaris PC Unit, Models 8000 and 8015Cardinal Health 303,Inc.2009—
+1K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014Class I
+2K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc.2023Class I
+3K243855BD Alaris Infusion System with Guardrails Suite MXCarefusion 303, Inc.2025—
+2K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link…Fresenius Kabi AG2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201210
201411
202210
202311
202510
FIG. 1 — Lineage of K051641, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K950419The Imed Orion Infusion Pump and Administration Sets
Summary page 2
…Predicate Device Manufacturer 510(k No. Date Cleared Alaris System Cardinal Health, Alaris K950419 June 21, 1995 (originally submitted as Products IMED Orion Infusion System)_ INTEN…
Imed Corp.1995Class I

Also cleared against K950419: 4 other devices →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091308Alaris PC Unit, Models 8000 and 8015Cardinal Health 303,Inc.20095
K133532Alaris System with Guardrails Suite MX Class ICarefusion 303, Inc.20143

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 11, 2026Class IIZ-1645-2026Under Investigation by firmDue to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units
Sep 15, 2023Class IZ-0185-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0184-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Sep 15, 2023Class IZ-0186-2024Under Investigation by firmAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Feb 4, 2020Class IZ-1360-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Feb 4, 2020Class IZ-1362-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Feb 4, 2020Class IZ-1361-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Jun 12, 2017Class IIZ-2671-2017Software designBD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions.
Nov 1, 2016Class IIZ-1520-2017Use errorReports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped.
May 12, 2016Class IIZ-2064-2016Software designA patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.
Mar 24, 2016Class IIZ-1606-2016Nonconforming Material/ComponentThe Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.
Feb 12, 2016Class IIZ-1239-2016Under Investigation by firmCareFusion is recalling the Alaris PC unit because a component on the PC unit power supply may cause a "System Error" or "Missing Battery" error code (120.4630).
May 13, 2015Class IIZ-1811-2015Software designAn issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.
Mar 12, 2015Class IIZ-1311-2015Process controlCareFusion is recalling the Alaris PC unit because of an error code. The error code may occur upon power on during the "Power-On Self Test" due to a keypad issue.
Dec 5, 2013Class IZ-0534-2014Software designCareFusion is recalling the Alaris PC unit (model 8000 and 8015) because the Alaris PC unit will produce an audible alarm and the Alaris PC unit will display a System Error message.
Jun 17, 2013Class IZ-1835-2013Process designCareFusion is recalling the Alaris PC units model 8015 (PC unit), version 9.12, because it is operating at an incorrect voltage and it could potentially experience a loss of communication between the PC Unit main Processor and Keyboard…
Mar 6, 2013Class IIZ-1112-2013Software designThe recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic…
Mar 6, 2013Class IZ-1098-2013Software designThe recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.
May 25, 2012Class IZ-2076-2012Component design/selectionThe recall was initiated because the Alaris PC unit model 8015 has a component on the PC unit power supply board that is causing an error code(120.4630) "System Error" or "Missing Battery Error" at start up. The error code is accompanied…
Sep 13, 2011Not classifiedZ-0876-2012Nonconforming Material/ComponentThe recall was initiated because Carefusion has identified a potential risk with a PLD component on the Alaris PC unit model 8015 power supply board. The PLD component used in production between June 21, 2011 and August 15, 2011 has a…
Aug 2, 2010Not classifiedZ-0492-2011Device DesignThan recall was initiated because CareFusion has identified identified a potential risk associated with the Alaris PC unit formerly known as Medley PC unit (PC Unit) model 8000 only when used with Alaris Pump module (Pump module) or…
Jul 13, 2010Not classifiedZ-0006-2011Component design/selectionUnder certain wireless network conditions a communication error can occur, which freezes the PC Unit screen, which may result in a delay of therapy. A delay of therapy may result in serious injury and/or death. CareFusion identified this…
Jun 12, 2009Not classifiedZ-1790-2009Device DesignThere is a potential risk for medication not being delivered when IV tubing sets are occluded in the Alaris Pump module. This can occur if IV sets are set up in the infusion pumps but not used for an extended period of time. If users fail…
Jun 12, 2009Not classifiedZ-1791-2009Device DesignAffected Unit: Alaris PC Unit (Models 8000 and 8015) with software a versions 8 and above when used with the Alaris PCA module (Model 8120) There is a potential risk of over infusion related to a warning message on the user interface of…
Jun 12, 2009Not classifiedZ-1794-2009Device DesignThere is a potential risk of IUI connector failure in the Alaris System when these connectors become corroded or contaminated, which may result in an interruption of infusion with an audible and visual alarm. IUI connectors can become…
Jun 12, 2009Not classifiedZ-1792-2009Device DesignThere is a potential in the specified ESD (electrostatic discharge) protection circuitry in the keypad of Alaris PC units manufactured or serviced between April 2008 through January 2009. In these units there is a potential for the keypad…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · Direct predicate · ◆ on the line of descent in FIG. 14 Class I recalls
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • Its one ancestor has a recall, Class I.
  • Of its descendants, K091308 (Cardinal Health 303,Inc., 2009) has gone furthest, with 5 descendants of its own.
  • 4 other devices also cite its predicate K950419.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Alaris PC Unit is the main user interface unit and power supply of the Alaris System, a modular system to be used with Alaris System modules (aka Medley System modules) intended for use in today's growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices. The specific intended use for each Alaris System module is specified in its respective submission.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051641: Alaris System PC Unit, Model 8001.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K051641

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026