Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo, Agilia USB Cable

K210074510(k) TraditionalCleared Mar 1, 2022413-day reviewClass II21 CFR 880.5725

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Kabi AG · FRN Pump, Infusion

Cleared on Mar 1, 2022 as substantially equivalent to K133532 (Carefusion 303, Inc., 2014) after a 413-day review. Its family tree reaches back 4 generations to 1994. 6 of the 18 devices in its family, counting itself, have a Class I recall.

Decided
Mar 1, 2022
Review
413 days
Class
Class II
Product code
FRN
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K940147 (1994). 17 ancestors in all, the oldest K890120 (1989). No later device cites it as a predicate yet. 6 of the 18 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2009: 1 other ancestor2007: 1 other ancestor, 1 with a Class I recall12005: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor2001: 2 other ancestors, 1 with a Class I recall11995: 2 other ancestors, 1 with a Class I recall12000: 1 other ancestor1999: 1 other ancestor1996: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestorK940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalledK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recallK133532 · Alaris System with Guardrails Suite MX · Carefusion 303, Inc. · 2014 · Class I recall1990200020102020
  1. Root · 4 generations backK940147 Auto Syringe AS40A Infusion PumpBaxter Corp. · 1994 · Recalled
  2. 3 generations backK974332 Variable Pressure Syringe PumpAlaris Medical Systems, Inc. · 1998
  3. 2 generations backK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. 1 generation backK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Recalled (Class I)
  5. This deviceK210074 Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia…Fresenius Kabi AG · 2022

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K940147Auto Syringe AS40A Infusion PumpBaxter Corp.19941
−3K974332Variable Pressure Syringe PumpAlaris Medical Systems, Inc.1998—
−2K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
−1K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014Class I
0 (this device)K210074Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link…Fresenius Kabi AG2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199521
199610
199910
200010
200121
200310
200511
200711
200910
FIG. 1 — Lineage of K210074, 1994–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133532Alaris System with Guardrails Suite MX
Summary page 5
…II Infusion pump accessories 80 MRZ Legally Marketed Device Under Which Substantial Equivalence is Being Claimed Alaris System with Guardrails Suite MX (K133532) Device Description The Agilia SP Infusion System includes the Agilia SP MC WiFi Syringe Infusion Pump which is a programmable…
Carefusion 303, Inc.2014Class I

Also cleared against K133532: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Direct predicate · ◆ on the line of descent in FIG. 128 Class I recalls
  • K072105Alaris System with MMS, Model 8XXXCardinal Health 303,Inc. · 2007 · 2 generations back · ◆ in the 2007 ancestors bar of FIG. 13 Class I recalls
  • K051641Alaris System PC Unit, Model 8001Cardinal Health, Alaris Products · 2005 · 2 generations back · ◆ in the 2005 ancestors bar of FIG. 110 Class I recalls
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii)…Alaris Medical Systems, Inc. · 2001 · 2 generations back · ◆ in the 2001 ancestors bar of FIG. 13 Class I recalls
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 28 years and passes through 4 companies; a new generation was cleared every 7.5 years on median.
  • 11 of the 17 ancestors have a recall of any class; 6 have a Class I recall.
  • 1 other device also cites its predicate K133532.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Agilia SP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medication, blood, and blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous, and intraosseous using dedicated administration sets. The Agilia SP Infusion System is also intended for neonatal care for the…

Read the full statement (114 words)

The Agilia SP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medication, blood, and blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous, and intraosseous using dedicated administration sets.

The Agilia SP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, medication, blood, and blood derivatives through clinically accepted parenteral routes of administration and critical drugs under specific conditions. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.

It is intended for use by trained healthcare professionals in healthcare facilities

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K210074: Agilia SP Infusion System, Agilia SP MC WIFI Infusion Pump, Agilia Link, Agilia Duo, Agilia USB Cable.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K210074

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026