Synthes Radial Head Prosthesis System

K112030510(k) TraditionalCleared Jun 19, 2012340-day reviewClass II21 CFR 888.3170Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Synthes USA, LLC · KWI Prosthesis, Elbow, Hemi-, Radial, Polymer

Cleared on Jun 19, 2012 as substantially equivalent to K051385 (Biomet, Inc., 2005) and 1 other predicate after a 340-day review. Its family tree reaches back 4 generations to 1999. It has been recalled 2 times.

Decided
Jun 19, 2012
Review
340 days
Class
Class II
Product code
KWI
Recalls
2
Predicates
2

Its longest line of descent runs back 4 generations to K984290 (1999). 9 ancestors in all, the oldest K944507 (1995). No later device cites it as a predicate yet.

  1. Root · 4 generations backK984290 Implex Radial Head Replacement SystemImplex Corp. · 1999
  2. 3 generations backK012551 Liverpool Radial Head Replacement DeviceBiomet, Inc. · 2002
  3. 2 generations backK040611 Modular Radial Head Replacement DeviceBiomet, Inc. · 2004 · Recalled
  4. 1 generation backK051385 Explor Modular Radial Head Replacement DeviceBiomet, Inc. · 2005 · Recalled
  5. This deviceK112030 Synthes Radial Head Prosthesis SystemSynthes USA, LLC · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K984290Implex Radial Head Replacement SystemImplex Corp.1999—
−3K012551Liverpool Radial Head Replacement DeviceBiomet, Inc.2002—
−2K040611Modular Radial Head Replacement DeviceBiomet, Inc.20041
−1K051385Explor Modular Radial Head Replacement DeviceBiomet, Inc.20052
0 (this device)K112030Synthes Radial Head Prosthesis SystemSynthes USA, LLC2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199810
199910
200010
200310
FIG. 1 — Lineage of K112030, 1999–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWI device

  1. #1 K260624 Arthrex Radial Head Replacement Systemscore 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K051385Explor Modular Radial Head Replacement Device
Summary page 2
…s the same indications for use, Equivalence: the same fundamental technological characteristics, and similar materials as the predicate Biomet ExploRTM (K051385) and Ascension® (K032686) Modular Radial Head Devices. In vitro performance testing demonstrates the ability of…
Biomet, Inc.20052
*K032686Ascension Modular Radial Head
Summary page 1
…sis Product Code: KWI Predicate Biomet ExploRTm Radial Head Prosthesis System (KOS51385) Device: Ascension® Modular Radial Head (K032686) Titan Endoskeleton Interbody Fusion Devices (K 111626, K 102067, K 100889, K(083714, and K…
Ascension Orthopedics, Inc.2003—
†K111626Endoskeleton (R) TasTitan Spine, LLC2011—
†K102067Endoskeleton toTitan Spine, LLC2010—
†K100889Endoskeleton TCTitan Spine, LLC2010—
†K083714Endoskeleton TT Interbody Fusion Device (Ibd)Titan Spine, LLC2009—
†K080615Endoskeleton TA Interbody Fusion Device (Ibd)Titan Spine, LLC2008—
†K111272Rhausler Plage Anterior Cervical Fusion SystemRhausler, Inc.2011—
†K053088Slactive ImplantsInstitut Straumann AG2006—
†K033984Straumann Dental Implant SystemInstitut Straumann AG2004—
†K072578Synthes (Usa) Epoca Shoulder Prosthesis SystemSynthes (Usa)2007—
†K083439Synthes Epoca Shoulder Prosthesis SystemSynthes (Usa)2009—

Also cleared against K051385: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 29, 2016Class IIZ-1124-2017Under Investigation by firmThere is the possibility that the radial stem may loosen post-operatively at the stem bone interface.
Feb 22, 2013Class IIZ-1238-2013Device DesignThe Trial Head may come loose from the implant stem during manipulation of the arm during surgery. To date, no adverse events have been reported related to this issue and this recall is not being initiated as a result of adverse events.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
  • 5 of the 9 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K051385.

How does its category compare?

All KWI devices

KWI Prosthesis, Elbow, Hemi-, Radial, Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · KWI

1in 2024

2010: 12012: 12013: 12014: 12017: 12019: 12020: 12021: 12022: 12023: 12024: 12026 (partial year): 1
2007high 1 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWI

Too few KWI 510(k)s in any year since 2007 to give a median.

Recalls per year · KWI

2in 2022

2013: 12014: 12017: 12022: 2
2007high 2 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWI

118in 2025

2007: 152008: 162009: 142010: 132011: 102012: 162013: 312014: 292015: 1032016: 1382017: 3602018: 2022019: 2142020: 1182021: 442022: 442023: 772024: 1242025: 118
2007high 360 (2017) · low 10 (2011)2025

MAUDE reports naming product code KWI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—015
20080—016
20090—014
20101—013
20110—010
20121—016
20131—131
20141—129
20150—0103
20160—0138
20171—1360
20180—0202
20191—0214
20201—0118
20211—044
20221—244
20231—077
20241—0124
20250—0118
20261—0—

What is it indicated for?

The Synthes Radial Head Prosthesis System is intended for:

1. The replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radlo-humeral and/or proximal radlo-ulnar joint with

a. Joint destruction and/or subluxation visible on x-ray

b. Resistance to conservative treatment

2. Primary replacement after fracture of the radial head.

3. Symptomatic sequelae after radial head resection

4. Revision following failed radial head arthroplasty.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112030: Synthes Radial Head Prosthesis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112030

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026