Baxter Sigma Spectrum Infusion Pump with Master Drug Library

K133801510(k) TraditionalCleared May 7, 2014145-day reviewClass II21 CFR 880.5725Recalls 15 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corporation · FRN Pump, Infusion

Cleared on May 7, 2014 as substantially equivalent to K042121 (Sigma Intl., 2004) after a 145-day review. Its family tree reaches back 4 generations to 1990. 9 of the 27 devices in its family, counting itself, have a Class I recall. 18 later devices descend from it. It has been recalled 15 times, including 5 Class I recalls.

Decided
May 7, 2014
Review
145 days
Class
Class II
Product code
FRN
Recalls
15 (5 Class I)
Predicates
1

Its longest line of descent runs back 4 generations to K904518 (1990). 8 ancestors in all. 18 later devices descend from it, 3 citing it directly. 9 of the 27 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2000: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2023: 2 descendants cleared, 2 with a Class I recall22018: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2016: 2 descendants cleared2022: 4 descendants cleared2024: 3 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2021: 1 descendant cleared2020: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK042121 · Spectrum and Spectrum with Master Drug Library · Sigma Intl. · 2004 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 3 descendants, 2 with a Class I recall22 generations forward: 7 descendants, 2 with a Class I recall23 generations forward: 3 descendants4 generations forward: 3 descendants, 1 with a Class I recall15 generations forward: 2 descendants K042121 · Spectrum and Spectrum with Master Drug Library · Sigma Intl. · 2004 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0+5
  1. Root · 4 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 3 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 2 generations backK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Recalled (Class I)
  4. 1 generation backK042121 Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Recalled (Class I)
  5. This deviceK133801 Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014
  6. Descendant · cites it directly · 9 of its ownK173084 Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · Recalled (Class I)
  7. Descendant · cites it directly · 5 of its ownK142374 ivWatch Model 400Ivwatch, LLC · 2015
  8. Descendant · 2 generations on · 4 of its ownK153605 ivWatch Model 400Ivwatch, LLC · 2016
  9. Descendant · 2 generations on · 3 of its ownK220417 Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022
  10. Descendant · 3 generations on · 3 of its ownK162478 ivWatchIvwatch, LLC · 2016

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2015: 12016: 22018: 12019: 12020: 12021: 12022: 42023: 22024: 32025: 2
2015high 4 (2022) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 18 · See all
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · cites it directly9descendants
  • K142374ivWatch Model 400Ivwatch, LLC · 2015 · cites it directly5descendants
  • K153605ivWatch Model 400Ivwatch, LLC · 2016 · 2 generations on4descendants
  • K220417Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022 · 2 generations on3descendants
  • K162478ivWatchIvwatch, LLC · 2016 · 3 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−3K994375Horizon LightB.Braun Medical, Inc.2000Class I
−2K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001Class I
−1K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl.2004Class I
0 (this device)K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation2014—
+1K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation2018Class I
+1K142374ivWatch Model 400Ivwatch, LLC2015—
+2K153605ivWatch Model 400Ivwatch, LLC2016—
+2K220417Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation2022—
+3K162478ivWatchIvwatch, LLC2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200010
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201620
201811
201911
202010
202110
202240
202322
202431
202520
FIG. 1 — Lineage of K133801, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042121Spectrum and Spectrum with Master Drug Library
Summary page 1
…device: -SIGMA Spectrum and Spectrum with Master Drug Library. cleared Aug. 26, 2004 (K042121). Device Description: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is a la…
Sigma Intl.2004Class I

Also cleared against K042121: 2 other devices →

What descends from it? (18)

See all 18

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173084Spectrum IQ Infusion System, Dose IQ Safety Software Class IBaxter Healthcare Corporation20189
K142374ivWatch Model 400Ivwatch, LLC20155
K230022Baxter SIGMA Spectrum Infusion Pump with Master Drug Library Class IBaxter Healthcare Corporation20231

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 20, 2025Class IZ-2095-2025Employee errorCertain Spectrum infusion pumps may have an incorrect version of software.
Feb 5, 2025Class IZ-1253-2025Under Investigation by firmThere is the potential for missing motor mounting screws, which may have occurred during the servicing process.
Nov 19, 2024Class IIZ-0775-2025Process controlDevices were identified as released after repair without full testing being performed, which includes flow testing.
Dec 29, 2021Class IZ-0530-2022Device DesignThere is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion…
Jan 6, 2021Class IIZ-1150-2021Under Investigation by firmBattery performance issues related to prolonged storage.
Nov 23, 2020Class IIZ-0714-2021Under Investigation by firmThere is a potential software error during programming.
Jun 30, 2020Class IZ-0121-2021Labeling designBaxter is updating the device IFU for safety communication regarding cleaning practices.
Jun 30, 2020Class IZ-0122-2021Labeling designBaxter is updating the device IFU for safety communication regarding cleaning practices.
Jun 21, 2018Class IIZ-1248-2019Process controlThe flow rate test verification was not properly executed for certain Sigma Spectrum infusion pumps prior to release. The testing was performed for less than the required minimum run time. Improperly performed flow rate testing may not…
May 8, 2018Class IIZ-2113-2018OtherThe firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies…
Oct 6, 2017Class IIZ-0971-2018Nonconforming Material/ComponentCertain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
Jan 26, 2017Class IIZ-1693-2017Process designSIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in…
Sep 28, 2016Class IIZ-0671-2017Process controlA correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Sep 28, 2016Class IIZ-0672-2017Process controlA correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential…
Mar 2, 2015Class IIZ-1451-2015Under Investigation by firmLoading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K230022Baxter SIGMA Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2023 · Cites it directly · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · Cites it directly · ◆ in the 2018 descendants bar of FIG. 14 Class I recalls
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls
  • K211122Novum IQ Large Volume PumpBaxter Healthcare Corporation · 2024 · 2 generations on · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls
  • K230041Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 9.4 years on median.
  • 5 of the 8 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K173084 (Baxter Healthcare Corporation, 2018) has gone furthest, with 9 descendants of its own.
  • 2 other devices also cite its predicate K042121.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master…

Read the full statement (160 words)

The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.

The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas.

The Daxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133801: Baxter Sigma Spectrum Infusion Pump with Master Drug Library.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133801

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026