Spectrum and Spectrum with Master Drug Library

K042121510(k) TraditionalCleared Aug 26, 200420-day reviewClass II21 CFR 880.5725Recalls 18 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sigma Intl. · FRN Pump, Infusion

Cleared on Aug 26, 2004 as substantially equivalent to K030459 (Alaris Medical Systems, Inc., 2003) and 3 other predicates after a 20-day review. Its family tree reaches back 3 generations to 1990. 10 of the 36 devices in its family, counting itself, have a Class I recall. 28 later devices descend from it. It has been recalled 18 times, including 3 Class I recalls.

Decided
Aug 26, 2004
Review
20 days
Class
Class II
Product code
FRN
Recalls
18 (3 Class I)
Predicates
4

Its longest line of descent runs back 3 generations to K904518 (1990). 7 ancestors in all. 28 later devices descend from it, 3 citing it directly. 10 of the 36 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2000: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2013: 1 descendant cleared2009: 1 descendant cleared2014: 3 descendants cleared, 1 with a Class I recall12020: 2 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared2016: 3 descendants cleared2023: 3 descendants cleared, 2 with a Class I recall22018: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2022: 4 descendants cleared2024: 3 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2021: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 8 descendants, 3 with a Class I recall33 generations forward: 9 descendants, 2 with a Class I recall24 generations forward: 3 descendants5 generations forward: 3 descendants, 1 with a Class I recall16 generations forward: 2 descendants K011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0+5
  1. Root · 3 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 2 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 1 generation backK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Recalled (Class I)
  4. This deviceK042121 Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004
  5. Descendant · cites it directly · 18 of its ownK133801 Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Recalled (Class I)
  6. Descendant · 2 generations on · 9 of its ownK173084 Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · Recalled (Class I)
  7. Descendant · cites it directly · 7 of its ownK123049 Sapphire Infusion PumpQ Core Medical, Ltd. · 2013
  8. Descendant · 2 generations on · 5 of its ownK142374 ivWatch Model 400Ivwatch, LLC · 2015
  9. Descendant · 3 generations on · 4 of its ownK153605 ivWatch Model 400Ivwatch, LLC · 2016

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2009: 12013: 12014: 32015: 12016: 32017: 22018: 12019: 12020: 22021: 12022: 42023: 32024: 32025: 2
2009high 4 (2022) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 28 · See all
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · cites it directly18descendants
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · 2 generations on9descendants
  • K123049Sapphire Infusion PumpQ Core Medical, Ltd. · 2013 · cites it directly7descendants
  • K142374ivWatch Model 400Ivwatch, LLC · 2015 · 2 generations on5descendants
  • K153605ivWatch Model 400Ivwatch, LLC · 2016 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−2K994375Horizon LightB.Braun Medical, Inc.2000Class I
−1K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001Class I
0 (this device)K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl.2004—
+1K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation2014Class I
+2K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation2018Class I
+1K123049Sapphire Infusion PumpQ Core Medical, Ltd.2013—
+2K142374ivWatch Model 400Ivwatch, LLC2015—
+3K153605ivWatch Model 400Ivwatch, LLC2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200010
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201310
201431
201510
201630
201720
201811
201911
202021
202110
202240
202332
202431
202520
FIG. 1 — Lineage of K042121, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030459Medley System with Medication Management System (MMS)
Summary page 2
…Master Drug Library are identified as follows: Premarket Device Name Applicant Notification, 510(k) Number K030459 Medley' TM System with Medication ALARIS Medical Systems, Managem…
Alaris Medical Systems, Inc.2003—
*K011975Horizon Outlook with Dosecom
Summary page 2
…59 Medley' TM System with Medication ALARIS Medical Systems, Management System Inc. K011975 Horizon OutlookT M with B. Braun Medical Inc. DoseComTM Th…
B.Braun Medical, Inc.2001Class I
*K950766Sigma 8000 and Sigma 8002
Summary page 3
…510(k) Summary (continuation) Premarket Device Name Applicant Notification, 510(k) Number K950766 SIGMA Model 8000 and 8002 SIGMA International General Infusion Pumps…
Sigma Intl. General Medical Apparatus, Inc.1995—
*K002211Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump 2M8151
Summary page 3
…Number K950766 SIGMA Model 8000 and 8002 SIGMA International General Infusion Pumps Medical Apparatus, LLC. K002211 Colleague® CX Volumetric Baxter Healthcare Infusion Pump…
Baxter Healthcare Corp20004

Also cleared against K030459: 8 other devices →

What descends from it? (28)

See all 28

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133801Baxter Sigma Spectrum Infusion Pump with Master Drug Library Class IBaxter Healthcare Corporation201418
K123049Sapphire Infusion PumpQ Core Medical, Ltd.20137
K093332Delphi Infusion Pump/ Delphi Administration Set Model IV-01000/ Model IV-02XXXX (Where X…Delphi Medical Systems20090

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 6, 2021Class IIZ-1150-2021Under Investigation by firmBattery performance issues related to prolonged storage.
Nov 23, 2020Class IIZ-0713-2021Under Investigation by firmThere is a potential software error during programming.
Nov 23, 2020Class IIZ-0712-2021Under Investigation by firmThere is a potential software error during programming.
Jun 30, 2020Class IZ-0120-2021Labeling designBaxter is updating the device IFU for safety communication regarding cleaning practices.
May 8, 2018Class IIZ-2112-2018OtherThe firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies…
Apr 3, 2018Class IIZ-1667-2018Employee errorService technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at…
Apr 3, 2018Class IIZ-1664-2018Employee errorService technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at…
Apr 3, 2018Class IIZ-1666-2018Employee errorService technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at…
Apr 3, 2018Class IIZ-1665-2018Employee errorService technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at…
Apr 3, 2018Class IIZ-1663-2018Employee errorService technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at…
Sep 28, 2016Class IIZ-0670-2017Process controlA correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Sep 3, 2014Class IIZ-2738-2014Equipment maintenanceOne Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.
Feb 7, 2014Class IZ-1484-2014Software designBaxter Healthcare Corporation has issued an Urgent Device Correction for the SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library due to repeated System Error 322 occurrences.
Sep 24, 2013Class IIZ-0189-2014Under Investigation by firmBaxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.
Jun 10, 2013Class IIZ-0197-2014Process designBaxter Healthcare Corporation issued an Urgent Device Correction for the SIGMA SPECTRUM Infusion Pump with Master Drug Drug Library due to reports of upstream occlusion alarms and air-in-line alarms occurring during infusions.
Apr 23, 2012Class IIIZ-2030-2012Software designSigma is a safety alert for the Master Drug Library (MDL) Editor software associated with the SIGMA Spectrum Infusion Pump System because it may result in a facility drug library that does not represent common clinical practices.
Jul 6, 2011Class IZ-1490-2012Nonconforming Material/ComponentRecall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent…
Sep 15, 2010Not classifiedZ-0146-2011Device DesignThe infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 15 Class I recalls
  • K192860Sapphire Infusion Pump, Sapphire administration setsQ Core Medical, Ltd. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K230022Baxter SIGMA Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls
  • K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 14 Class I recalls
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls
  • K211122Novum IQ Large Volume PumpBaxter Healthcare Corporation · 2024 · 3 generations on · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls
  • K230041Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2023 · 5 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • 4 of the 7 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K133801 (Baxter Healthcare Corporation, 2014) has gone furthest, with 18 descendants of its own.
  • 8 other devices also cite its predicate K030459.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Spectrum and Spectrum with Master Drug Library is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically acceptable routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. The spectrum is intended to be…

Read the full statement (150 words)

The Spectrum and Spectrum with Master Drug Library is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically acceptable routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. The spectrum is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by the user.

The Spectrum and Spectrum with Master Drug Library is suitable for many user facility applications such as but not limited to hospitals, outpatient care areas, homecare and ambulatory care services.

The Spectrum and Spectrum with Master Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming. Parameter programming requires trained healthcare professional confirmation of limits and drug therapy to physician's directive.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042121: Spectrum and Spectrum with Master Drug Library.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K042121

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026