510(k) Report

510(k) · Cleared

Recalled · Class I · 2021

Spectrum IQ Infusion System, Dose IQ Safety Software

K173084510(k) TraditionalCleared May 11, 2018224-day reviewClass II21 CFR 880.5725Recalls 8 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corporation · FRN Pump, Infusion

Cleared on May 11, 2018 as substantially equivalent to K133801 (Baxter Healthcare Corporation, 2014) after a 224-day review. Its family tree reaches back 5 generations to 1990. 8 of the 19 devices in its family, counting itself, have a Class I recall. 9 later devices descend from it. It has been recalled 8 times, including 4 Class I recalls.

Decided
May 11, 2018
Review
224 days
Class
Class II
Product code
FRN
Recalls
8 (4 Class I)
Predicates
1

Its longest line of descent runs back 5 generations to K904518 (1990). 9 ancestors in all. 9 later devices descend from it, 5 citing it directly. 8 of the 19 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2000: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2022: 3 descendants cleared2024: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK042121 · Spectrum and Spectrum with Master Drug Library · Sigma Intl. · 2004 · Class I recallK133801 · Baxter Sigma Spectrum Infusion Pump with Master Drug Library · Baxter Healthcare Corporation · 2014 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 3 other ancestors5 generations back: 1 other ancestor 1 generation forward: 5 descendants, 2 with a Class I recall22 generations forward: 2 descendants3 generations forward: 1 descendant, 1 with a Class I recall14 generations forward: 1 descendant K133801 · Baxter Sigma Spectrum Infusion Pump with Master Drug Library · Baxter Healthcare Corporation · 2014 · Class I recallK042121 · Spectrum and Spectrum with Master Drug Library · Sigma Intl. · 2004 · Class I recallK011975 · Horizon Outlook with Dosecom · B.Braun Medical, Inc. · 2001 · Class I recallK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled−50
  1. Root · 5 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 4 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 3 generations backK011975 Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Recalled (Class I)
  4. 2 generations backK042121 Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · Recalled (Class I)
  5. 1 generation backK133801 Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Recalled (Class I)
  6. This deviceK173084 Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2018
  7. Descendant · cites it directly · 3 of its ownK220417 Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022
  8. Descendant · 2 generations on · 2 of its ownK222048 Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022
  9. Descendant · 3 generations on · 1 of its ownK230041 Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2023 · Recalled (Class I)
  10. Descendant · cites it directly · 1 of its ownK183311 Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · Recalled (Class I)
  11. Descendant · 4 generations on · 0 of its ownK251636 Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)Baxter Healthcare Corporation · 2025 · Recalled

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2019: 12021: 12022: 32023: 12024: 22025: 1
2019high 3 (2022) · low 0 (2020)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 9 · See all
  • K220417Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022 · cites it directly3descendants
  • K222048Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2022 · 2 generations on2descendants
  • K230041Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2023 · 3 generations on1descendant
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · cites it directly1descendant
  • K251636Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)Baxter Healthcare Corporation · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−4K994375Horizon LightB.Braun Medical, Inc.2000Class I
−3K011975Horizon Outlook with DosecomB.Braun Medical, Inc.2001Class I
−2K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl.2004Class I
−1K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation2014Class I
0 (this device)K173084Spectrum IQ Infusion System, Dose IQ Safety SoftwareBaxter Healthcare Corporation2018—
+1K220417Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation2022—
+2K222048Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation2022—
+3K230041Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation2023Class I
+1K183311Ivenix Infusion System (IIS)Ivenix, Inc.2019Class I
+4K251636Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)Baxter Healthcare Corporation20251

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200010
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201911
202110
202230
202311
202421
202510
FIG. 1 — Lineage of K173084, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133801Baxter Sigma Spectrum Infusion Pump with Master Drug Library
Summary page 5
…ble 2. Predicate Device Device Company Predicate 510(k) Clearance Date Baxter SIGMA Baxter Healthcare K133801 May 7, 2014 Spectrum Infusion Corporation Pump with Master Drug Library DESCRIPTION OF THE DEV…
Baxter Healthcare Corporation2014Class I

Also cleared against K133801: 2 other devices →

What descends from it? (9)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220417Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation20223
K183311Ivenix Infusion System (IIS) Class IIvenix, Inc.20191
K230665Dose IQ Safety SoftwareBaxter Healthcare Corporation20240
K211122Novum IQ Large Volume Pump Class IBaxter Healthcare Corporation20240
K211124Dose IQ Safety SoftwareBaxter Healthcare Corporation20220

Has anything in its family been recalled? (8 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 29, 2021Class IZ-0529-2022Device DesignThere is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion…
Jul 7, 2021Class IZ-2165-2021Software designSoftware issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
Jun 4, 2021Class IZ-2110-2021Software designSystem errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Mar 19, 2021Class IIZ-1487-2021Device DesignPotentially defective component in the AC power adapter which may fail over time.
Jan 6, 2021Class IIZ-1150-2021Under Investigation by firmBattery performance issues related to prolonged storage.
Nov 23, 2020Class IIZ-0715-2021Under Investigation by firmThere is a potential software error during programming.
Jun 30, 2020Class IZ-0123-2021Labeling designBaxter is updating the device IFU for safety communication regarding cleaning practices.
Sep 18, 2018Class IIZ-0107-2019Software designIf an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133801Baxter Sigma Spectrum Infusion Pump with Master Drug LibraryBaxter Healthcare Corporation · 2014 · Direct predicate · ◆ on the line of descent in FIG. 15 Class I recalls
  • K042121Spectrum and Spectrum with Master Drug LibrarySigma Intl. · 2004 · 2 generations back · ◆ on the line of descent in FIG. 13 Class I recalls
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · Cites it directly · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls
  • K211122Novum IQ Large Volume PumpBaxter Healthcare Corporation · 2024 · Cites it directly · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls
  • K230041Spectrum IQ Infusion System with Dose IQ Safety SoftwareBaxter Healthcare Corporation · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 28 years and passes through 4 companies; a new generation was cleared every 4.0 years on median.
  • 6 of the 9 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K220417 (Baxter Healthcare Corporation, 2022) has gone furthest, with 3 descendants of its own.
  • 2 other devices also cite its predicate K133801.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, and blood products. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, or epidural. The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used in conjunction with legally marketed…

Read the full statement (170 words)

The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, and blood products. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, or epidural. The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.

The Spectrum IQ Infusion System with Dose IQ Safety Software is suitable for a variety of patient care environments such as, but not limited to, hospitals and outpatient care areas.

The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to reduce operator interaction through guided programming, including a way to automate the programming of infusion parameters and documentation of infusion therapies. This automation is intended to reduce pump programming errors.

The Spectrum IQ Infusion System with Dose IQ Safety Software is intended to be used by trained healthcare professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173084: Spectrum IQ Infusion System, Dose IQ Safety Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173084

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026