510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Discovery MR750 3.0T ,Discovery MR450 1.5T, Discovery MR750 3.0T,Optima MR450W 1.5T

K142085510(k) TraditionalCleared Sep 26, 201457-day reviewClass II21 CFR 892.1000Recalls 27

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Sep 26, 2014 as substantially equivalent to K132376 (Ge Healthcare (Ge Medical Systems, LLC), 2013) after a 57-day review. Its family tree reaches back 12 generations to 1997. 5 of the 70 devices in its family, counting itself, have a Class I recall. 40 later devices descend from it. It has been recalled 27 times.

Decided
Sep 26, 2014
Review
57 days
Class
Class II
Product code
LNH
Recalls
27
Predicates
1

Its longest line of descent runs back 12 generations to K971684 (1997). 29 ancestors in all, the oldest K962061 (1996). 40 later devices descend from it, 6 citing it directly. 5 of the 70 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2011: 1 other ancestor2005: 3 other ancestors2004: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall12002: 2 other ancestors, 1 with a Class I recall12001: 4 other ancestors, 1 with a Class I recall11999: 2 other ancestors1998: 1 other ancestor1996: 1 other ancestor2016: 4 descendants cleared2015: 3 descendants cleared2017: 5 descendants cleared2022: 4 descendants cleared2023: 4 descendants cleared2019: 4 descendants cleared2018: 1 descendant cleared2020: 4 descendants cleared2024: 2 descendants cleared2025: 4 descendants cleared2026: 4 descendants cleared2021: 1 descendant clearedK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK081028 · Signa 3.0T MR750 System · Ge Medical Systems, LLC · 2008 · recalledK083147 · Discovery MR450 · Ge Medical Systems, LLC · 2008 · recalledK091536 · Optima MR450W · Ge Medical Systems, LLC · 2009 · recalledK113490 · Optima MR450W · Ge Medical Systems, LLC · 2012 · recalledK123522 · Optima MR450W · Ge Healthcare (Ge Medical Systems, LLC) · 2013 · recalledK132376 · Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima MR450W 1.5T · Ge Healthcare (Ge Medical Systems, LLC) · 2013 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor5 generations back: 1 other ancestor6 generations back: 2 other ancestors, 1 with a Class I recall17 generations back: 4 other ancestors, 2 with a Class I recall28 generations back: 4 other ancestors, 1 with a Class I recall19 generations back: 4 other ancestors 1 generation forward: 6 descendants2 generations forward: 9 descendants3 generations forward: 9 descendants4 generations forward: 5 descendants5 generations forward: 2 descendants6 generations forward: 5 descendants7 generations forward: 4 descendants K132376 · Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima MR450W 1.5T · Ge Healthcare (Ge Medical Systems, LLC) · 2013 · recalledK123522 · Optima MR450W · Ge Healthcare (Ge Medical Systems, LLC) · 2013 · recalledK113490 · Optima MR450W · Ge Medical Systems, LLC · 2012 · recalledK091536 · Optima MR450W · Ge Medical Systems, LLC · 2009 · recalledK083147 · Discovery MR450 · Ge Medical Systems, LLC · 2008 · recalledK081028 · Signa 3.0T MR750 System · Ge Medical Systems, LLC · 2008 · recalledK072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalled−10−50+5
  1. Root · 12 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 8 generations between
  3. 3 generations backK113490 Optima MR450WGe Medical Systems, LLC · 2012 · Recalled
  4. 2 generations backK123522 Optima MR450WGe Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled
  5. 1 generation backK132376 Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T…Ge Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled
  6. This deviceK142085 Discovery MR750 3.0T ,Discovery MR450 1.5T, Discovery MR750…Ge Medical Systems, LLC · 2014
  7. Descendant · cites it directly · 16 of its ownK160618 Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T…Ge Medical Systems, LLC · 2016 · Recalled
  8. Descendant · 2 generations on · 15 of its ownK163331 Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T…Ge Medical Systems, LLC · 2017 · Recalled
  9. Descendant · cites it directly · 15 of its ownK143345 SIGNA PioneerGe Healthcare Japan Corporation · 2015 · Recalled
  10. Descendant · 3 generations on · 11 of its ownK171128 SIGNA PremierGe Medical Systems, LLC · 2017 · Recalled
  11. Descendant · 4 generations on · 10 of its ownK183231 SIGNA PremierGe Medical Systems, LLC · 2019 · Recalled
Show all 12 generations
  1. Root · 12 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 11 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 10 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 9 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  5. 8 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  6. 7 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  7. 6 generations backK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  8. 5 generations backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  9. 4 generations backK091536 Optima MR450WGe Medical Systems, LLC · 2009 · Recalled
  10. 3 generations backK113490 Optima MR450WGe Medical Systems, LLC · 2012 · Recalled
  11. 2 generations backK123522 Optima MR450WGe Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled
  12. 1 generation backK132376 Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T…Ge Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled
  13. This deviceK142085 Discovery MR750 3.0T ,Discovery MR450 1.5T, Discovery MR750…Ge Medical Systems, LLC · 2014

Showing 5 of 40 descendants. See all descendants

Descendants cleared per year

2015: 32016: 42017: 52018: 12019: 42020: 42021: 12022: 42023: 42024: 22025: 42026: 4
2015high 5 (2017) · low 1 (2018)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 40 · See all
  • K160618Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5TGe Medical Systems, LLC · 2016 · cites it directly16descendants
  • K163331Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T…Ge Medical Systems, LLC · 2017 · 2 generations on15descendants
  • K143345SIGNA PioneerGe Healthcare Japan Corporation · 2015 · cites it directly15descendants
  • K171128SIGNA PremierGe Medical Systems, LLC · 2017 · 3 generations on11descendants
  • K183231SIGNA PremierGe Medical Systems, LLC · 2019 · 4 generations on10descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−11K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−10K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−9K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
−8K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
−7K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
−6K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
−5K083147Discovery MR450Ge Medical Systems, LLC200810
−4K091536Optima MR450WGe Medical Systems, LLC20098
−3K113490Optima MR450WGe Medical Systems, LLC20123
−2K123522Optima MR450WGe Healthcare (Ge Medical Systems, LLC)20131
−1K132376Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima…Ge Healthcare (Ge Medical Systems, LLC)201315
0 (this device)K142085Discovery MR750 3.0T ,Discovery MR450 1.5T, Discovery MR750 3.0T,Optima MR450W…Ge Medical Systems, LLC2014—
+1K160618Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima…Ge Medical Systems, LLC201640
+2K163331Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery…Ge Medical Systems, LLC201736
+1K143345SIGNA PioneerGe Healthcare Japan Corporation20156
+3K171128SIGNA PremierGe Medical Systems, LLC20178
+4K183231SIGNA PremierGe Medical Systems, LLC20197

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199810
199920
200141
200221
200311
200411
200530
201110
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201530
201640
201750
201810
201940
202040
202110
202240
202340
202420
202540
202640
FIG. 1 — Lineage of K142085, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132376Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima MR450W 1.5T
Summary page 5
…Product Code: LNH Predicate Device(s): Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, and Optima MR450w 1.5T [K132376] Device Description: The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0…
Ge Healthcare (Ge Medical Systems, LLC)201315

Also cleared against K132376: 1 other device →

What descends from it? (40)

See all 40

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K160618Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5TGe Medical Systems, LLC201616
K143345SIGNA PioneerGe Healthcare Japan Corporation201515
K161567SIGNA VoyagerGE Healthcare20164
K143251SIGNA Creator, SIGNA ExplorerGe Healthcare (Ge Medical Systems, LLC)20153
K161397MAGiCGe Medical Systems, LLC20161
K162722HyperSenseGe Medical Systems, LLC20170

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 20, 2025Class IIZ-2306-2025Equipment maintenanceGE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA…
Jun 20, 2025Class IIZ-2302-2025Equipment maintenanceGE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA…
Jun 20, 2025Class IIZ-2301-2025Equipment maintenanceGE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA…
Oct 28, 2024Class IIZ-0517-2025Device DesignGradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
Jun 14, 2024Class IIZ-2464-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Jun 14, 2024Class IIZ-2467-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Jun 14, 2024Class IIZ-2465-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Jun 14, 2024Class IIZ-2468-2024Device DesignGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2…
Dec 23, 2021Class IIZ-0601-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Dec 23, 2021Class IIZ-0593-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Dec 23, 2021Class IIZ-0594-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Dec 23, 2021Class IIZ-0602-2022Process controlThe magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
Aug 26, 2021Class IIZ-0135-2022Software designUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Aug 26, 2021Class IIZ-0136-2022Software designUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Aug 26, 2021Class IIZ-0140-2022Software designUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Jun 4, 2021Class IIZ-2056-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Jun 4, 2021Class IIZ-2055-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Jun 4, 2021Class IIZ-2063-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Jun 4, 2021Class IIZ-2064-2021Software designGE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed…
Jun 6, 2018Class IIZ-3279-2018Nonconforming Material/ComponentGE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient…
May 11, 2017Class IIZ-3024-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
May 11, 2017Class IIZ-3025-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
May 11, 2017Class IIZ-3022-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
May 11, 2017Class IIZ-3023-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Jan 20, 2017Class IIZ-1188-2017Device DesignPotential safety issue with the patient bore heating on the Discovery MR450 narrow bore 1.5T MRI. In the event of a RF component failure, typically happening after extended system use, a small area on the bore surface may become warmer…
Jun 8, 2016Class IIZ-2131-2016Nonconforming Material/ComponentGE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a…
Dec 7, 2015Class IIZ-0201-2019Nonconforming Material/ComponentGE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 6 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 7 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 7 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 8 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 17 years and passes through 4 companies; a new generation was cleared every 11 months on median.
  • 22 of the 29 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K160618 (Ge Medical Systems, LLC, 2016) has gone furthest, with 16 descendants of its own.
  • 1 other device also cites its predicate K132376.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the…

Read the full statement (154 words)

The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T systems reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142085: Discovery MR750 3.0T ,Discovery MR450 1.5T, Discovery MR750 3.0T,Optima MR450W 1.5T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142085

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026