iSR'obot MRI-US Fusion

K161109510(k) TraditionalCleared Oct 6, 2016169-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biobot Surgical Pte., Ltd. · LLZ System, Image Processing, Radiological

Cleared on Oct 6, 2016 as substantially equivalent to K120187 (Igt, LLC Dba Eigen, 2012) after a 169-day review. Its family tree reaches back 8 generations to 1993. 6 of the 89 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it.

Decided
Oct 6, 2016
Review
169 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
1

Its longest line of descent runs back 8 generations to K930090 (1993). 86 ancestors in all, the oldest K882383 (1988). 2 later devices descend from it, 1 citing it directly. 6 of the 89 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2006: 2 other ancestors2008: 1 other ancestor2007: 1 other ancestor2005: 4 other ancestors2004: 2 other ancestors, 1 with a Class I recall12003: 6 other ancestors2002: 8 other ancestors2001: 10 other ancestors2000: 9 other ancestors, 1 with a Class I recall11999: 5 other ancestors1998: 10 other ancestors, 3 with a Class I recall31997: 3 other ancestors1996: 6 other ancestors, 1 with a Class I recall11995: 3 other ancestors1994: 4 other ancestors1993: 1 other ancestor1988: 1 other ancestor1991: 1 other ancestor2021: 1 descendant cleared2023: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalledK041182 · Syntegra · Adac Laboratories · 2004K081987 · Unisyn Image Fusion, Version 1.0 · Convergent Imaging Solutions · 2008K120187 · Multi-Modality Image Fusion · Igt, LLC Dba Eigen · 20121990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 41 other ancestors, 6 with a Class I recall64 generations back: 10 other ancestors5 generations back: 13 other ancestors6 generations back: 7 other ancestors7 generations back: 3 other ancestors8 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K120187 · Multi-Modality Image Fusion · Igt, LLC Dba Eigen · 2012K081987 · Unisyn Image Fusion, Version 1.0 · Convergent Imaging Solutions · 2008K041182 · Syntegra · Adac Laboratories · 2004K032036 · Gemini 16 · Philips Medical Systems (Cleveland), Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50
  1. Root · 8 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 6 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 5 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK032036 Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · Recalled
  6. 3 generations backK041182 SyntegraAdac Laboratories · 2004
  7. 2 generations backK081987 Unisyn Image Fusion, Version 1.0Convergent Imaging Solutions · 2008
  8. 1 generation backK120187 Multi-Modality Image FusionIgt, LLC Dba Eigen · 2012
  9. This deviceK161109 iSR'obot MRI-US FusionBiobot Surgical Pte., Ltd. · 2016
  10. Descendant · cites it directly · 1 of its ownK213411 iSR’obot Mona Lisa 2.0Biobot Surgical Pte., Ltd. · 2021
  11. Descendant · 2 generations on · 0 of its ownK232320 iSR`obot Mona Lisa 2.0Biobot Surgical Pte., Ltd. · 2023

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K213411iSR’obot Mona Lisa 2.0Biobot Surgical Pte., Ltd. · 2021 · cites it directly1descendant
  • K232320iSR`obot Mona Lisa 2.0Biobot Surgical Pte., Ltd. · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K930090HelicatElscint, Inc.1993—
−7K961464Select/SPElscint, Inc.19963
−6K982060Volumax CT Imaging SystemElscint, Inc.1998—
−5K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−4K032036Gemini 16Philips Medical Systems (Cleveland), Inc.20032
−3K041182SyntegraAdac Laboratories2004—
−2K081987Unisyn Image Fusion, Version 1.0Convergent Imaging Solutions2008—
−1K120187Multi-Modality Image FusionIgt, LLC Dba Eigen2012—
0 (this device)K161109iSR'obot MRI-US FusionBiobot Surgical Pte., Ltd.2016—
+1K213411iSR’obot Mona Lisa 2.0Biobot Surgical Pte., Ltd.2021—
+2K232320iSR`obot Mona Lisa 2.0Biobot Surgical Pte., Ltd.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199110
199310
199440
199530
199661
199730
1998103
199950
200091
2001100
200280
200360
200421
200540
200620
200710
200810
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202310
FIG. 1 — Lineage of K161109, 1993–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120187Multi-Modality Image Fusion
Summary page 7
…, measurements and data reporting. Technology UroFusion employs the same fundamental scientific technology as that of its predicate device, Multi-modality Image Fusion (K120187). The majority of the software features and functions are common between the two products. The table below outlines the major sub…
Igt, LLC Dba Eigen2012—

Also cleared against K120187: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213411iSR’obot Mona Lisa 2.0Biobot Surgical Pte., Ltd.20211

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 3 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 7 companies; a new generation was cleared every 3.3 years on median.
  • 29 of the 86 ancestors have a recall of any class; 6 have a Class I recall.
  • Of its descendants, K213411 (Biobot Surgical Pte., Ltd., 2021) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K120187.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

The iSR'obot MRI-US Fusion is a software application to be used by Clinicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of Magnetic Resonance and Ultrasound images for mapping planning information of the prostate gland and region of interest. The software features also include multi-modality data communication, surface and volume rendering, segmentation, multi-planar reconstruction, organ delineation, region of interest delineation, landmark selection, measurements and data reporting.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161109: iSR'obot MRI-US Fusion.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161109

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026