510(k) · Cleared
Recalled · Class II · 2026Allura Xper FD series and Allura Xper OR Table series
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems Nederland B.V. · OWB Interventional Fluoroscopic X-Ray System
Cleared on Jul 29, 2016 as substantially equivalent to K141979 (Philips Medical Systems Netherland BV, 2014) after a 53-day review. Its family tree reaches back 12 generations to 1980. 2 of the 56 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 45 times.
- Decided
- Jul 29, 2016
- Review
- 53 days
- Class
- Class II
- Product code
- OWB
- Recalls
- 45
- Predicates
- 1
Its longest line of descent runs back 12 generations to K801623 (1980). 53 ancestors in all. 2 later devices descend from it, 1 citing it directly. 2 of the 56 devices in this family, counting this one, have a Class I recall.
- Root · 12 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 8 generations between
- 3 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
- 2 generations backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
- 1 generation backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
- This deviceK161563 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016
- Descendant · cites it directly · 1 of its ownK162859 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK220311 Cardiovascular Angiography SystemAllengers Medical Systems Limited · 2022
Show all 12 generations
- Root · 12 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
- 11 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
- 10 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
- 9 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
- 8 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
- 7 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
- 6 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
- 5 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
- 4 generations backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
- 3 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
- 2 generations backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
- 1 generation backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
- This deviceK161563 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016
Class I recalls in this family
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −12 | K801623 | Cavitron Ultrasonic Surg. Aspir. Cusa | Cavitron Corp. | 1980 | — |
| −11 | K962952 | Olympus Ultrasonic Surgical System (Usu) | Olympus Optical, Co. | 1996 | — |
| −10 | K972114 | Olympus Sonosurg System | The Olympus Optical Co. | 1998 | — |
| −9 | K981993 | Olympus Integrated Endosurgery System Endoalpha | The Olympus Optical Co. | 1998 | — |
| −8 | K994231 | Siemens Integrated Operating System (Sios) | Siemens Medical Corp. | 2000 | — |
| −7 | K052202 | Axiom Artis - Modular Angiographic System | Siemens Medical Solutions USA, Inc. | 2006 | 29 |
| −6 | K073290 | Artis Zee and Zeego | Siemens Medical Solutions USA, Inc. | 2008 | 39 |
| −5 | K090745 | Artis Zee / Zeegos with SW VC14 | Siemens Medical Solutions USA, Inc. | 2009 | 7 |
| −4 | K102005 | Allura Xper or Table Series | Philips Medical Systems North America, Inc. | 2010 | 39 |
| −3 | K130638 | Alluraclarity Xper FD Series X-Ray System | Philips Medical Systems Nederland B.V. | 2013 | 39 |
| −2 | K133292 | Allura Xper FD; or Table | Philips Medical Systems Nederland B.V. | 2014 | 45 |
| −1 | K141979 | Allura Xper FD Series / Allura Xper or Table Series | Philips Medical Systems Netherland BV | 2014 | 81 |
| 0 (this device) | K161563 | Allura Xper FD series and Allura Xper OR Table series | Philips Medical Systems Nederland B.V. | 2016 | — |
| +1 | K162859 | Allura Xper FD series and Allura Xper OR Table series | Philips Medical Systems Nederland B.V. | 2016 | Class I |
| +2 | K220311 | Cardiovascular Angiography System | Allengers Medical Systems Limited | 2022 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1989 | 1 | 0 |
| 1991 | 3 | 0 |
| 1992 | 1 | 0 |
| 1993 | 2 | 0 |
| 1994 | 1 | 0 |
| 1995 | 10 | 0 |
| 1996 | 3 | 0 |
| 1997 | 2 | 0 |
| 1999 | 2 | 0 |
| 2000 | 4 | 0 |
| 2001 | 1 | 0 |
| 2002 | 5 | 0 |
| 2003 | 2 | 1 |
| 2004 | 3 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 1 | 1 |
| 2022 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWB device
- #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K141979 | Allura Xper FD Series / Allura Xper or Table SeriesSummary page 4…series Manufacturer: Philips Medical Systems Nederland B.V. 510(k) Clearance: K141979 (August 19, 2014) Classification Name: Image-intensified fluoroscopic x-ray system… | 2014 |
What descends from it? (2)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K162859 | Allura Xper FD series and Allura Xper OR Table series Class I | 2016 | 1 |
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 29, 2026 | Class II | Z-2928-2026 | Software design | Firmware issue potentially leading to loss of geometry movements. |
| Jun 17, 2026 | Class II | Z-2893-2026 | Under Investigation by firm | Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. |
| May 18, 2026 | Class II | Z-2412-2026 | Radiation Control for Health and Safety Act | Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection |
| Apr 30, 2026 | Class II | Z-2259-2026 | Under Investigation by firm | It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the… |
| Apr 30, 2026 | Class II | Z-2258-2026 | Under Investigation by firm | It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the… |
| Dec 15, 2025 | Class II | Z-1062-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1063-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1065-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1066-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1067-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1071-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1072-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1074-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1077-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1076-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1064-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1075-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1069-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1070-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1073-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Dec 15, 2025 | Class II | Z-1068-2026 | Other | Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not… |
| Sep 3, 2025 | Class II | Z-2624-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2634-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2626-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2625-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2629-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2635-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2628-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2633-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2623-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2630-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2632-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2631-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Sep 3, 2025 | Class II | Z-2627-2025 | Device Design | Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become… |
| Jun 30, 2025 | Class II | Z-2189-2025 | Software design | Potential for temporary loss of imaging (X-ray) functionality due to software issue. |
| Nov 11, 2024 | Class II | Z-0676-2025 | Equipment maintenance | Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life… |
| Aug 8, 2024 | Class II | Z-2716-2024 | Under Investigation by firm | A half-threaded bolt was included in the LTE kits instead of the correct full-threaded bolt. |
| Jan 2, 2024 | Class II | Z-1156-2024 | Process design | Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure. |
| Jan 2, 2024 | Class II | Z-1151-2024 | Process design | Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure. |
| Nov 3, 2023 | Class II | Z-0558-2024 | Device Design | Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility… |
| Jun 30, 2023 | Class II | Z-2485-2023 | Under Investigation by firm | Loss of availability of the wireless foot switch during procedures. |
| Nov 24, 2021 | Class II | Z-0724-2022 | Process control | Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to… |
| Nov 24, 2021 | Class II | Z-0725-2022 | Process control | Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to… |
| Nov 19, 2021 | Class II | Z-0416-2022 | Other | Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure… |
| Oct 29, 2021 | Class II | Z-0476-2022 | Device Design | Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure |
The 5 latest of 45 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 113 Class I recalls
- K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
What does this lineage say?
- The line of descent spans 36 years and passes through 8 companies; a new generation was cleared every 1.9 years on median.
- 19 of the 53 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K162859 (Philips Medical Systems Nederland B.V., 2016) has gone furthest, with 1 descendant of its own.
- 2 other devices also cite its predicate K141979.
How does its category compare?
All OWB devicesOWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · OWB
22in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWB
148in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWB
42so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWB
14,008in 2025
MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 52 | 0 | 1 |
| 2008 | 4 | 135 | 1 | 0 |
| 2009 | 9 | 99 | 0 | 0 |
| 2010 | 7 | 66 | 0 | 0 |
| 2011 | 6 | 79 | 0 | 1 |
| 2012 | 6 | 37 | 0 | 1 |
| 2013 | 11 | 119 | 10 | 27 |
| 2014 | 15 | 158 | 9 | 103 |
| 2015 | 11 | 66 | 35 | 2,134 |
| 2016 | 16 | 103 | 13 | 4,074 |
| 2017 | 20 | 76 | 13 | 1,604 |
| 2018 | 23 | 54 | 6 | 1,449 |
| 2019 | 9 | 174 | 7 | 1,558 |
| 2020 | 16 | 53 | 8 | 1,581 |
| 2021 | 24 | 60 | 31 | 1,981 |
| 2022 | 21 | 90 | 24 | 2,932 |
| 2023 | 12 | 152 | 22 | 7,604 |
| 2024 | 13 | 126 | 11 | 7,942 |
| 2025 | 22 | 148 | 30 | 14,008 |
| 2026 | 15 | 221 | 42 | — |
What is it indicated for?
The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform: • Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. • Cardiac imaging…
Read the full statement (168 words)
The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform:
• Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.
• Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).
• Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.
Additionally:
• The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room.
• Allura Xper FD10 is compatible with specified magnetic navigation systems.
• Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular, Non Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K161563 · printed Nov 6, 2026
Cite this page
510(k) Report. “K161563: Allura Xper FD series and Allura Xper OR Table series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161563