510(k) Report

510(k) · Cleared

Cleared

PercuNav Image Fusion and Interventional Navigation System

K223804510(k) TraditionalCleared Jan 18, 202330-day reviewClass II21 CFR 892.1750

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Ultrasound, LLC · JAK System, X-Ray, Tomography, Computed

Cleared on Jan 18, 2023 as substantially equivalent to K201053 (Philips Ultrasound, Inc., 2020) after a 30-day review. Its family tree reaches back 9 generations to 1994. 6 of the 54 devices in its family, counting itself, have a Class I recall.

Decided
Jan 18, 2023
Review
30 days
Class
Class II
Product code
JAK
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K911783 (1994). 53 ancestors in all, the oldest K882383 (1988). No later device cites it as a predicate yet. 6 of the 54 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2005: 2 other ancestors2004: 1 other ancestor, 1 with a Class I recall12003: 3 other ancestors2002: 7 other ancestors2001: 7 other ancestors2000: 5 other ancestors1999: 4 other ancestors1998: 7 other ancestors, 1 with a Class I recall11997: 2 other ancestors1996: 3 other ancestors, 1 with a Class I recall11995: 1 other ancestor1993: 1 other ancestor1988: 1 other ancestorK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994×2: 2 generations cleared close together, K981998 (1998), K983529 (1998) K994270 · Instatrak 3000 System with Fluorotrak Module, Model IT3000 · Visualization Technology, Inc. · 2000 · Class I recallK053610 · Abaris · Traxtal Technologies, Inc. · 2006 · recalledK121498 · Percunav · Philips Health Care · 2012K132087 · Percunav System · Philips Health Care · 2013K170716 · PercuNav Image Fusion and Interventional Navigation · Philips Ultrasound, Inc. · 2017K201053 · PercuNav Image Fusion and Interventional Navigation · Philips Ultrasound, Inc. · 20201990200020102020
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 5 generations between
  3. 3 generations backK132087 Percunav SystemPhilips Health Care · 2013
  4. 2 generations backK170716 PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc. · 2017
  5. 1 generation backK201053 PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc. · 2020
  6. This deviceK223804 PercuNav Image Fusion and Interventional Navigation SystemPhilips Ultrasound, LLC · 2023
Show all 9 generations
  1. Root · 9 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 8 generations backK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  3. 7 generations backK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
  4. 6 generations backK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
  5. 5 generations backK053610 AbarisTraxtal Technologies, Inc. · 2006 · Recalled
  6. 4 generations backK121498 PercunavPhilips Health Care · 2012
  7. 3 generations backK132087 Percunav SystemPhilips Health Care · 2013
  8. 2 generations backK170716 PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc. · 2017
  9. 1 generation backK201053 PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc. · 2020
  10. This deviceK223804 PercuNav Image Fusion and Interventional Navigation SystemPhilips Ultrasound, LLC · 2023

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−8K981998Pediatric Instatrak SystemVisualization Technology, Inc.1998Class I
−7K983529Instatrak 3000Visualization Technology, Inc.1998Class I
−6K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc.2000Class I
−5K053610AbarisTraxtal Technologies, Inc.20063
−4K121498PercunavPhilips Health Care2012—
−3K132087Percunav SystemPhilips Health Care2013—
−2K170716PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc.2017—
−1K201053PercuNav Image Fusion and Interventional NavigationPhilips Ultrasound, Inc.2020—
0 (this device)K223804PercuNav Image Fusion and Interventional Navigation SystemPhilips Ultrasound, LLC2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199310
199510
199631
199720
199871
199940
200050
200170
200270
200330
200411
200520
FIG. 1 — Lineage of K223804, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K201053PercuNav Image Fusion and Interventional Navigation
Summary page 4
…l Radiology Predicate Device PercuNav Image Fusion and Interventional Navigation (K201053, cleared May 14, 2020) III. Overview The PercuNav Image Fusion and Interventional Navigation System provides image-guided di…
Philips Ultrasound, Inc.2020—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040050Instatrak System with Multiple Dataset Navigation, Model IT3500Ge Medical Systems Navigation and Visualization IN · 2004 · 6 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 6 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K982994Skull Base Instatrak SystemVisualization Technology, Inc. · 1998 · 6 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 6 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 29 years and passes through 6 companies; a new generation was cleared every 3.1 years on median.
  • 19 of the 53 ancestors have a recall of any class; 6 have a Class I recall.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The PercuNav Image Fusion and Interventional Navigation System is a stereotaxic accessory for computed tomography (CT), cone beam CT (CBCT), magnetic resonance (MR), ultrasound (US), and positron emission tomography (PET). CT, Ultrasound, PET, and MR may be fused in various combinations, such as CT with MR, MR with ultrasound, and so on. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle or…

Read the full statement (226 words)

The PercuNav Image Fusion and Interventional Navigation System is a stereotaxic accessory for computed tomography (CT), cone beam CT (CBCT), magnetic resonance (MR), ultrasound (US), and positron emission tomography (PET). CT, Ultrasound, PET, and MR may be fused in various combinations, such as CT with MR, MR with ultrasound, and so on. It may include instrumentation to display the simulated image of a tracked insertion tool such as a biopsy needle or probe on a computer monitor screen that shows images of the target organs and the current and the projected future path of the interventional instrument. The PercuNav Image Fusion and Interventional Navigation System is intended for treatment planning and to assist guidance for clinical, interventional, or diagnostic procedures in a clinical setting.

The PercuNav Image Fusion and Interventional Navigation System is also intended to supplement live imaging in clinical interventions to determine the proximity of one device relative to another.

The PercuNav Image Fusion and Interventional Navigation System is not intended to be the sole guidance for any procedure. Procedures that can be guided by the PercuNav Image Fusion and Interventional Navigation System adjunctively include, but are not limited to, the following:

• Image fusion for diagnostic clinical examinations and procedures

• Soft tissue biopsies

• Soft tissue ablation

• Bone ablation

• Bone biopsies

• Nerve blocks and pain management

• Drainage placements

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K223804: PercuNav Image Fusion and Interventional Navigation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223804

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026