510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000, HM2000) Systems

K984545510(k) TraditionalCleared Feb 25, 199965-day reviewClass II21 CFR 892.1600Recalls 17

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America, Inc. · IZI System, X-Ray, Angiographic

Cleared on Feb 25, 1999 after a 65-day review. The FDA summary names its predicate without a number. 10 of the 267 devices in its family, counting itself, have a Class I recall. 266 later devices descend from it. It has been recalled 17 times.

Decided
Feb 25, 1999
Review
65 days
Class
Class II
Product code
IZI
Recalls
17
Predicates
0

266 later devices descend from it, 8 citing it directly. 10 of the 267 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 1 descendant cleared2004: 2 descendants cleared2002: 2 descendants cleared2006: 4 descendants cleared2005: 1 descendant cleared2007: 5 descendants cleared2013: 7 descendants cleared, 1 with a Class I recall12003: 2 descendants cleared2010: 7 descendants cleared2009: 4 descendants cleared2008: 4 descendants cleared2014: 14 descendants cleared2011: 4 descendants cleared2017: 10 descendants cleared, 2 with a Class I recall22015: 6 descendants cleared2012: 8 descendants cleared2019: 10 descendants cleared, 1 with a Class I recall12024: 22 descendants cleared, 2 with a Class I recall22023: 21 descendants cleared, 1 with a Class I recall12021: 16 descendants cleared2018: 13 descendants cleared2016: 10 descendants cleared, 1 with a Class I recall12022: 18 descendants cleared2020: 14 descendants cleared, 1 with a Class I recall12025: 32 descendants cleared, 1 with a Class I recall12026: 29 descendants cleared1995200020052010201520202025
  1. This deviceK984545 Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000…Philips Medical Systems North America, Inc. · 1999
  2. Descendant · cites it directly · 136 of its ownK042961 Spy Intra-Operative Imaging System: SP2000Novadaq Technologies, Inc. · 2005 · Recalled
  3. Descendant · 2 generations on · 134 of its ownK060867 Spy Intraoperative Imaging SystemNovadaq Technologies, Inc. · 2006 · Recalled
  4. Descendant · 2 generations on · 133 of its ownK061871 Leica Model FL800Leica Microsystems, Inc. · 2006
  5. Descendant · 2 generations on · 131 of its ownK071037 Spy Intra-Operative Imaging System, Model SP2000Novadaq Technologies, Inc. · 2007 · Recalled
  6. Descendant · 2 generations on · 131 of its ownK063345 Spy Imaging SystemNovadaq Technologies, Inc. · 2007 · Recalled

Showing 5 of 266 descendants. See all descendants

Descendants cleared per year

1999: 12002: 22003: 22004: 22005: 12006: 42007: 52008: 42009: 42010: 72011: 42012: 82013: 72014: 142015: 62016: 102017: 102018: 132019: 102020: 142021: 162022: 182023: 212024: 222025: 322026: 29
1999high 32 (2025) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 266 · See all
  • K042961Spy Intra-Operative Imaging System: SP2000Novadaq Technologies, Inc. · 2005 · cites it directly136descendants
  • K060867Spy Intraoperative Imaging SystemNovadaq Technologies, Inc. · 2006 · 2 generations on134descendants
  • K061871Leica Model FL800Leica Microsystems, Inc. · 2006 · 2 generations on133descendants
  • K071037Spy Intra-Operative Imaging System, Model SP2000Novadaq Technologies, Inc. · 2007 · 2 generations on131descendants
  • K063345Spy Imaging SystemNovadaq Technologies, Inc. · 2007 · 2 generations on131descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K984545Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000, HM2000)…Philips Medical Systems North America, Inc.1999—
+1K042961Spy Intra-Operative Imaging System: SP2000Novadaq Technologies, Inc.20051
+2K060867Spy Intraoperative Imaging SystemNovadaq Technologies, Inc.20061
+2K061871Leica Model FL800Leica Microsystems, Inc.2006—
+2K071037Spy Intra-Operative Imaging System, Model SP2000Novadaq Technologies, Inc.20071
+2K063345Spy Imaging SystemNovadaq Technologies, Inc.20071

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200220
200320
200420
200510
200640
200750
200840
200940
201070
201140
201280
201371
2014140
201560
2016101
2017102
2018130
2019101
2020141
2021160
2022180
2023211
2024222
2025321
2026290
FIG. 1 — Lineage of K984545, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

10 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Philips Integris V5000
Named without a submission number
—
*—Philips Integris H5000
Named without a submission number
—

What descends from it? (266)

See all 266

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042961Spy Intra-Operative Imaging System: SP2000Novadaq Technologies, Inc.2005136
K020055Philips Integris Allura 9 with FD OptionPhilips Medical Systems2002124
K031836Philips Fresco Release 1Philips Medical Systems North America Co.200351
K062647E-View Fluoroscopy System, Model CS-50Omega Medical Imaging, Inc.200610
K070834CS-Series Fluoroscopy System, CS-10 Through CS-65 (11 Configurations)Omega Medical Imaging, Inc.20079
K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura… Class IPhilips Medical Systems Nederland B.V.20130
K041780New BridMedical Imaging Solutions, Inc.20040
K991922Digital Imaging System (BSR)Siemens Medical Solutions USA, Inc.19990

Has anything in its family been recalled? (10 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 19, 2019Class IIZ-1140-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Feb 19, 2019Class IIZ-1141-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Sep 24, 2014Class IIZ-1280-2016Device DesignLoss of key image functionality due to a bent pedal of the Footswitch.
Oct 7, 2010Not classifiedZ-1655-2011Nonconforming Material/ComponentA wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
Feb 4, 2010Not classifiedZ-2449-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Feb 4, 2010Not classifiedZ-2448-2010Software designA lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator…
Dec 10, 2008Not classifiedZ-0344-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Dec 10, 2008Not classifiedZ-0345-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Dec 10, 2008Not classifiedZ-0346-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Dec 10, 2008Not classifiedZ-0348-2009Radiation Control for Health and Safety ActTube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
Oct 23, 2008Not classifiedZ-2475-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Oct 23, 2008Not classifiedZ-2471-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Oct 23, 2008Not classifiedZ-2473-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Oct 23, 2008Not classifiedZ-2472-2010OtherA high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Dec 14, 2006Not classifiedZ-0367-2007OtherMovement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
Oct 12, 2005Not classifiedZ-0206-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.
Oct 12, 2005Not classifiedZ-0207-06OtherPotential for high voltage exposure involving the injector interface connector of patient support tables.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130842Allura Xper FD or Table Series; Allura Xper FD10 Series; Allura Xper FD20 Series; Allura Xper FD Biplane…Philips Medical Systems Nederland B.V. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 111 Class I recalls
  • K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · 9 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 9 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 9 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 10 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · 10 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · 10 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K230309Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve…Centerline Biomedical, Inc. · 2023 · 10 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · 10 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 11 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • Of its descendants, K042961 (Novadaq Technologies, Inc., 2005) has gone furthest, with 136 descendants of its own.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—

What is it indicated for?

The Philips Integris Series Release 2 systems are intended for use in acquiring diagnostic quality images during cardiac, angiographic, vascular, neurovascular, and interventional applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K984545: Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000, HM2000) Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K984545

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.