510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

MX8000 V5.0 CT System

K010817510(k) TraditionalCleared Jun 15, 200188-day reviewClass II21 CFR 892.1750Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Jun 15, 2001 after an 88-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 264 later devices descend from it. It has been recalled 12 times.

Decided
Jun 15, 2001
Review
88 days
Class
Class II
Product code
JAK
Recalls
12
Predicates
0

264 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2001: 1 descendant cleared2004: 7 descendants cleared2003: 3 descendants cleared2008: 7 descendants cleared2009: 4 descendants cleared2005: 3 descendants cleared2007: 3 descendants cleared2016: 5 descendants cleared2006: 3 descendants cleared2015: 3 descendants cleared2018: 14 descendants cleared2017: 12 descendants cleared2011: 4 descendants cleared2019: 9 descendants cleared2014: 9 descendants cleared2013: 6 descendants cleared2010: 2 descendants cleared2020: 22 descendants cleared2012: 6 descendants cleared2023: 20 descendants cleared2025: 35 descendants cleared2022: 22 descendants cleared2024: 19 descendants cleared2026: 26 descendants cleared2021: 19 descendants cleared200020052010201520202025
  1. This deviceK010817 MX8000 V5.0 CT SystemPhilips Medical Systems (Cleveland), Inc. · 2001
  2. Descendant · cites it directly · 263 of its ownK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  3. Descendant · cites it directly · 214 of its ownK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  4. Descendant · 2 generations on · 147 of its ownK042293 Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  5. Descendant · 3 generations on · 92 of its ownK130159 Qlab Quantification Software Heart ModelPhilips Health Care · 2013
  6. Descendant · 4 generations on · 91 of its ownK132165 Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · Recalled

Showing 5 of 264 descendants. See all descendants

Descendants cleared per year

2001: 12003: 32004: 72005: 32006: 32007: 32008: 72009: 42010: 22011: 42012: 62013: 62014: 92015: 32016: 52017: 122018: 142019: 92020: 222021: 192022: 222023: 202024: 192025: 352026: 26
2001high 35 (2025) · low 0 (2002)2026

Most influential descendants

Showing 5 of 264 · See all
  • K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · cites it directly263descendants
  • K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · cites it directly214descendants
  • K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · 2 generations on147descendants
  • K130159Qlab Quantification Software Heart ModelPhilips Health Care · 2013 · 3 generations on92descendants
  • K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · 4 generations on91descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010817MX8000 V5.0 CT SystemPhilips Medical Systems (Cleveland), Inc.2001—
+1K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
+1K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
+2K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.20041
+3K130159Qlab Quantification Software Heart ModelPhilips Health Care2013—
+4K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc.20131

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200330
200470
200530
200630
200730
200870
200940
201020
201140
201260
201360
201490
201530
201650
2017120
2018140
201990
2020220
2021190
2022220
2023200
2024190
2025350
2026260
FIG. 1 — Lineage of K010817, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (264)

See all 264

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.2001263
K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003214
K033357Acqsim-Multislice-CT, Model 453567042271Philips Medical Systems (Cleveland), Inc.200410

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-1979-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Mar 16, 2015Class IIZ-1647-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Apr 1, 2014Class IIZ-1722-2014Component design/selectionA customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
Dec 20, 2012Class IIZ-0945-2013Component design/selectionPhilips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
Jun 3, 2011Not classifiedZ-2914-2011Device DesignPhilips Medical Systems decided to conduct a recall/field correction of the patient support vertical brake hub component in all of their CT MX8000 Dual v. Exp Computed Tomography X-Ray System scanners due to a component defect which could…
Sep 23, 2008Not classifiedZ-0597-2009Process controlThe product is being recalled because the patient support may travel downwards without being commanded to move.
Nov 6, 2006Not classifiedZ-0014-2007OtherA potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator.
Dec 27, 2005Class IIZ-2709-2014Device DesignThe system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.
Mar 11, 2005Not classifiedZ-1414-05OtherWhen two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.
Mar 11, 2005Not classifiedZ-1410-05OtherWhen two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.
Mar 11, 2005Not classifiedZ-1412-05OtherWhen two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.
Apr 2, 2003Not classifiedZ-0836-03OtherPatient images may be incorrectly stored in the archive directory of a different patient.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K012009 (Marconi Medical Systems, Inc., 2001) has gone furthest, with 263 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—
Cite this page

510(k) Report. “K010817: MX8000 V5.0 CT System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K010817

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026