510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Acqsim-Multislice-CT, Model 453567042271

K033357510(k) TraditionalCleared Jan 9, 200481-day reviewClass II21 CFR 892.1750Recalls 50

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Jan 9, 2004 as substantially equivalent to K001088 (Philips Medical Systems (Cleveland), Inc., 2000) and 2 other predicates after an 81-day review. Its family tree reaches back 4 generations to 1993. 10 later devices descend from it. It has been recalled 50 times.

Decided
Jan 9, 2004
Review
81 days
Class
Class II
Product code
JAK
Recalls
50
Predicates
3

Its longest line of descent runs back 4 generations to K930090 (1993). 10 ancestors in all. 10 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2000: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2016: 1 descendant cleared2008: 1 descendant cleared2017: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant cleared2018: 1 descendant cleared2026: 2 descendants cleared2022: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 4 descendants3 generations forward: 3 descendants K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 19930
  1. Root · 4 generations backK930090 HelicatElscint, Inc. · 1993
  2. 3 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 2 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 1 generation backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. This deviceK033357 Acqsim-Multislice-CT, Model 453567042271Philips Medical Systems (Cleveland), Inc. · 2004
  6. Descendant · cites it directly · 6 of its ownK171850 Philips CT Big BorePhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  7. Descendant · 2 generations on · 3 of its ownK240844 Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024
  8. Descendant · 2 generations on · 1 of its ownK181797 Philips CT Big Bore Sliding Gantry ConfigurationPhillips Medical Systems (Cleveland), Inc. · 2018
  9. Descendant · cites it directly · 1 of its ownK160611 P-ArtisP-Cure, Ltd. · 2016
  10. Descendant · 3 generations on · 0 of its ownK253649 Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd. · 2026

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2008: 12016: 12017: 12018: 12022: 12024: 12025: 22026: 2
2008high 2 (2025) · low 0 (2009)2026

Most influential descendants

Showing 5 of 10 · See all
  • K171850Philips CT Big BorePhilips Medical Systems (Cleveland), Inc. · 2017 · cites it directly6descendants
  • K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024 · 2 generations on3descendants
  • K181797Philips CT Big Bore Sliding Gantry ConfigurationPhillips Medical Systems (Cleveland), Inc. · 2018 · 2 generations on1descendant
  • K160611P-ArtisP-Cure, Ltd. · 2016 · cites it directly1descendant
  • K253649Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K930090HelicatElscint, Inc.1993—
−3K961464Select/SPElscint, Inc.19963
−2K982060Volumax CT Imaging SystemElscint, Inc.1998—
−1K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
0 (this device)K033357Acqsim-Multislice-CT, Model 453567042271Philips Medical Systems (Cleveland), Inc.2004—
+1K171850Philips CT Big BorePhilips Medical Systems (Cleveland), Inc.201714
+2K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd.2024—
+2K181797Philips CT Big Bore Sliding Gantry ConfigurationPhillips Medical Systems (Cleveland), Inc.2018—
+1K160611P-ArtisP-Cure, Ltd.2016—
+3K253649Spectral CT Verida FamilyPhilips Medical Systems Technologies, Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200010
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201610
201710
201810
202210
202410
202520
202620
FIG. 1 — Lineage of K033357, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001088Ultrasim CT System
Summary page 2
…e-CT" CT scanner is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: * AcQSim-CT, K001088 * Mx8000 IDT, K012009 · Mx8000v5.0, K010817 6. Safety and Effectiveness Considerations The safety of the device…
Philips Medical Systems (Cleveland), Inc.20002
*K012009MX8000 Idt CT System
Summary page 2
…able type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: * AcQSim-CT, K001088 * Mx8000 IDT, K012009 · Mx8000v5.0, K010817 6. Safety and Effectiveness Considerations The safety of the device is assured by adherence to G…
Marconi Medical Systems, Inc.200161
*K010817MX8000 V5.0 CT System
Summary page 2
…equivalent to the legally marketed devices currently in commercial distribution, namely: * AcQSim-CT, K001088 * Mx8000 IDT, K012009 · Mx8000v5.0, K010817 6. Safety and Effectiveness Considerations The safety of the device is assured by adherence to GMP practices and to Internati…
Philips Medical Systems (Cleveland), Inc.200112

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171850Philips CT Big BorePhilips Medical Systems (Cleveland), Inc.20176
K160611P-ArtisP-Cure, Ltd.20161
K081135Gemini CondorPhilips Medical Systems (Cleveland), Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-1983-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Mar 28, 2018Class IIZ-2273-2018OtherA problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Feb 8, 2018Class IIZ-1411-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1410-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Jan 3, 2018Class IIZ-0839-2018Under Investigation by firmSoftware issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and…
Jan 3, 2018Class IIZ-0840-2018Under Investigation by firmSoftware issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and…
Feb 23, 2017Class IIZ-1493-2017Software designPhilips identified a 4D CT pulmonary phase issue on Brilliance BigBore Oncology/Radiology CT with 4.2.0 software version through customer complaints. When the user performs a retrospective respiratory gated 4D CT scan with the Bellows…
Feb 23, 2017Class IIZ-1492-2017Software designWhen the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the…
Jul 20, 2016Class IIZ-2654-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jul 20, 2016Class IIZ-2653-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jun 21, 2016Class IIZ-2111-2016Nonconforming Material/ComponentSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack and cause injury to patients, operators, by standers or service personnel.
Jun 14, 2016Class IIZ-2110-2016Nonconforming Material/ComponentSocket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.
Jun 14, 2016Class IIZ-2109-2016Nonconforming Material/ComponentSocket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.
Mar 24, 2016Class IIZ-2367-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 24, 2016Class IIZ-2368-2016Software change controlSoftware defect causing intermittently slow response of Host.
Sep 14, 2015Class IIZ-0177-2016Software designThe firm discovered Ring/Dot artifact due to X-ray measurement error.
Mar 16, 2015Class IIZ-1648-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 6, 2015Class IIZ-1763-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1764-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Dec 8, 2014Class IIZ-0979-2015Process controlImproper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
Apr 1, 2014Class IIZ-1722-2014Component design/selectionA customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
Mar 27, 2014Class IIZ-1461-2014Nonconforming Material/ComponentBlock assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can…
Dec 19, 2013Class IIZ-1745-2014Nonconforming Material/ComponentDuring internal testing, the Brilliance CT Big Bore was found to be out of tolerance for radio frequency emissions. Specifically, at the 48MHZ frequency, the testing indicated the Brilliance CT Big Bore was 3.5dB uV/meter higher than the…
Dec 4, 2013Class IIZ-0801-2014Software designMachine Name in TumorLOC Basic Mode is obscured. When TumorLOC is in Basic Mode and Set Vertical Laser equals Yes, the Machine Name label and field are covered by the controls above it. The Machine Name cannot be viewed or changed.
Aug 30, 2013Class IIZ-0013-2014Software designThere are artifacts appearing on the Tracker images during threshold-triggered bolus scans with protocols using a Rotation Time of 0.4 seconds.
Dec 20, 2012Class IIZ-0945-2013Component design/selectionPhilips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
Oct 24, 2012Class IIZ-0880-2013Software designSoftware anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was…
Aug 31, 2012Class IIZ-2666-2014Software designPhilips Healthcare received a complaint stating the system opens e-stop while sitting idle causing all movements and scan to stop. The problem has only occurred one time when the scan was out of Ready for Scan or scanning mode.
Aug 3, 2012Class IIZ-2245-2012Device DesignThe patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
Jul 2, 2012Class IIZ-2035-2012Process controlPhilips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
Mar 26, 2012Not classifiedZ-1691-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Mar 26, 2012Not classifiedZ-1692-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Mar 26, 2012Not classifiedZ-1693-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Dec 13, 2011Not classifiedZ-0946-2012Software designThe Block and Multi Leaf Collimator (MLC) shape is not propagated to the beam that is modified by the Oppose operation. The contours can appear with the incorrect color in the Beam's Eye View Viewer, resulting in organs being incorrectly…
Oct 7, 2011Not classifiedZ-0845-2012Software designThe Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state
Sep 22, 2011Not classifiedZ-0334-2012Software designSoftware errors. A number of non-conformances were found for the Brilliance Big Bore and Brilliance 64 Computed Tomography X-ray Systems equipped with software version; 3.5 Beta1. The non-conformances stem from software errors identified…
Aug 29, 2011Not classifiedZ-0777-2012Nonconforming Material/ComponentSoftware errors. A number of non-conformances were logged in during in-house testing and external validation testing of Brilliance 64, Brilliance Big Bore, and Ingenuity CT systems, equipped with software version 3.5 Beta 2. The…
Mar 7, 2011Not classifiedZ-2207-2011OtherA program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
Mar 7, 2011Class IIZ-0791-2014Software designThe DoseRight feature suggest a mAs based on the measured patient size, a reference size and a reference mAs. When scanning large children, the suggested mAs may be higher than clinicians would expect.
Dec 10, 2010Not classifiedZ-1669-2011OtherPhilips Medical Systems has decided to recall the patient support vertical brake hub in all its Brilliance systems due to the collapse in the vertical direction. The short key connecting the vertical brake hub to the motor shaft can…
Nov 24, 2008Not classifiedZ-0286-2010Process designA leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
Nov 24, 2008Not classifiedZ-0287-2010Process designA leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
Sep 23, 2008Not classifiedZ-0597-2009Process controlThe product is being recalled because the patient support may travel downwards without being commanded to move.
Sep 22, 2008Not classifiedZ-0132-2009Process designImages not correlated: A leak between the Tube Interface and the Outlet Tube of the transducer may result in a failure to produce respiratory correlated images.
Sep 27, 2007Not classifiedZ-0121-2008Software designThe maximum intensity projection (MIP) images produced from respiratory gated images by the Tumor Localization (Tumor LOC) software application on Brilliance Big Bore CT scanners will be labeled incorrectly for certain patient…
Sep 10, 2007Not classifiedZ-0366-2008Software designIncorrect Measurement: Incorrect measurement will occur when the operator uses the distance or area measurement function, such as Graphic Tool line or ROI on zoomed and batch saved images from the Cardiac Viewer or Pulmonary Viewer and…
Jan 9, 2007Class IIZ-2678-2014Labeling designWhen the Extended Display Firled ov View reconstruction is used, the CT numbers and anatomy in teh extended region are not accurate.
Nov 6, 2006Not classifiedZ-0014-2007OtherA potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator.
Oct 2, 2006Not classifiedZ-0636-2007OtherTwo conditions have been identified with the Philips Medical System Brilliance Bore ,version 2.2.1 Tumor Localization, that may lead the user to misinterpret the displayed data.
Apr 25, 2006Class IIZ-0400-2016Software designBrilliance Big Bore- An anomaly was identified with software version 2.0 tumor localization application when printing work sheets. this causes information regarding relative locations of marked regions of interest on the patient work…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • Of its descendants, K171850 (Philips Medical Systems (Cleveland), Inc., 2017) has gone furthest, with 6 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The "AcQSim-Multislice-CT" is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033357: Acqsim-Multislice-CT, Model 453567042271.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033357

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026