510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis

K231482510(k) TraditionalCleared Nov 29, 2023190-day reviewClass II21 CFR 892.1750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shanghai United Imaging Healthcare Co., Ltd. · JAK System, X-Ray, Tomography, Computed

Cleared on Nov 29, 2023 as substantially equivalent to K223028 (Shanghai United Imaging Healthcare Co., Ltd., 2023) and 1 other predicate after a 190-day review. Its family tree reaches back 8 generations to 1993. 1 later device descends from it. It has been recalled 3 times.

Decided
Nov 29, 2023
Review
190 days
Class
Class II
Product code
JAK
Recalls
3
Predicates
2

Its longest line of descent runs back 8 generations to K930090 (1993). 13 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 8 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 6 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 5 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 4 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 3 generations backK172135 uCT Computed Tomography X-ray SystemShanghai United Imaging Healthcare Co., Ltd. · 2018 · Recalled
  7. 2 generations backK203448 uCT ATLAS with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2021 · Recalled
  8. 1 generation backK223028 uCT ATLAS Astound with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd. · 2023
  9. This deviceK231482 uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with…Shanghai United Imaging Healthcare Co., Ltd. · 2023
  10. Descendant · cites it directly · 0 of its ownK243617 uCT ATLAS Astound with uWS-CT-Dual Energy Analysis; uCT ATLAS with…Shanghai United Imaging Healthcare Co., Ltd. · 2025 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K243617uCT ATLAS Astound with uWS-CT-Dual Energy Analysis; uCT ATLAS with uWS-CT-Dual Energy…Shanghai United Imaging Healthcare Co., Ltd. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K930090HelicatElscint, Inc.1993—
−7K961464Select/SPElscint, Inc.19963
−6K982060Volumax CT Imaging SystemElscint, Inc.1998—
−5K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−4K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−3K172135uCT Computed Tomography X-ray SystemShanghai United Imaging Healthcare Co., Ltd.20187
−2K203448uCT ATLAS with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd.20213
−1K223028uCT ATLAS Astound with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd.2023—
0 (this device)K231482uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual…Shanghai United Imaging Healthcare Co., Ltd.2023—
+1K243617uCT ATLAS Astound with uWS-CT-Dual Energy Analysis; uCT ATLAS with uWS-CT-Dual…Shanghai United Imaging Healthcare Co., Ltd.20253

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K231482, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K223028uCT ATLAS Astound with uWS-CT-Dual Energy Analysis
Summary page 5
…4. Identification of Predicate/Reference Device(s) Predicate Device(s) Device Name: uCT ATLAS Astound with uWS-CT-Dual Energy Analysis 510(k) Number: K223028 Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Pro…
Shanghai United Imaging Healthcare Co., Ltd.2023—
*K203448uCT ATLAS with uWS-CT-Dual Energy Analysis
Summary page 6
…care Co., Ltd. T e l: +86 (21) 67076888 Fax:+86 (21) 67076889 www.united-imaging.com Device Name: uCT ATLAS with uWS-CT-Dual Energy Analysis 510(k) Number: K203448 Regulation Name: Computed Tomography X-ray System Regulation Number: 21 CFR 892.1750 Regulatory Class: II Pro…
Shanghai United Imaging Healthcare Co., Ltd.20213
†K230162uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy AnalysisShanghai United Imaging Healthcare Co., Ltd.20234

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K243617uCT ATLAS Astound with uWS-CT-Dual Energy Analysis; uCT ATLAS with uWS-CT-Dual Energy…Shanghai United Imaging Healthcare Co., Ltd.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2026Class IIZ-3282-2026Software designPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on…
Aug 11, 2026Class IIZ-3281-2026Software designPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on…
Sep 12, 2025Class IIZ-0221-2026OtherThird-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 30 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 6 of the 13 ancestors have a recall of any class.
  • Of its descendants, K243617 (Shanghai United Imaging Healthcare Co., Ltd., 2025) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

uuCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography. uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may…

Read the full statement (460 words)

uuCT ATLAS Astound is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS Astound is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.

uCT ATLAS Astound is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.

uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography.

uCT ATLAS has the capability to image a whole organ in a single rotation. Organs include, but not limited to head, heart, liver, kidney, pancreas, joints, etc.

uCT ATLAS is intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

* Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

uWS-CT-Dual Energy Analysis is a post-processing software package that accepts UIH CT images acquired using different tube voltages and/or tube currents of the same anatomical location. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. These differences provide information on the chemical composition of the scanned body materials and enable images to be generated at multiple energies within the available spectrum. uWS-CT-Dual Energy Analysis software combines images acquired with low and high energy spectra to visualize this information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231482: uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231482

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026