MR Elastography; MZEK-001A

K220018510(k) TraditionalCleared Mar 3, 202257-day reviewClass II21 CFR 892.1000

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Canon Medical Systems Corporation · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Mar 3, 2022 as substantially equivalent to K140666 (Philips Medical Systems Nederland B.V., 2014) and 1 other predicate after a 57-day review. Its family tree reaches back 14 generations to 1995. 11 of the 148 devices in its family, counting itself, have a Class I recall.

Decided
Mar 3, 2022
Review
57 days
Class
Class II
Product code
LNH
Recalls
None
Predicates
2

Its longest line of descent runs back 14 generations to K935595 (1995). 147 ancestors in all, the oldest K891689 (1989). No later device cites it as a predicate yet. 11 of the 148 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2020: 1 other ancestor2019: 1 other ancestor2018: 1 other ancestor2016: 1 other ancestor2012: 1 other ancestor, 1 with a Class I recall12017: 1 other ancestor2015: 1 other ancestor2011: 4 other ancestors2009: 2 other ancestors2008: 3 other ancestors2005: 15 other ancestors, 2 with a Class I recall22010: 2 other ancestors, 1 with a Class I recall12007: 3 other ancestors, 2 with a Class I recall22006: 7 other ancestors2004: 14 other ancestors, 2 with a Class I recall22002: 10 other ancestors, 1 with a Class I recall12003: 11 other ancestors, 1 with a Class I recall12001: 13 other ancestors, 1 with a Class I recall12000: 9 other ancestors1998: 4 other ancestors1997: 7 other ancestors1995: 5 other ancestors1994: 5 other ancestors1999: 9 other ancestors1996: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestorReference device: K201389 · Resoundant Acoustic Driver System · Resoundant, Inc. · 2020Reference device: K103411 · Ideal IQ Software Option · GE Medical Systems · 2011Reference device: K083421 · MR-Touch Option · Ge Medical Systems, LLC · 2009Reference device: K121434 · Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio · Siemens Medi Cal Solutions, Inc. · 2012 · Class I recallReference device: K141977 · Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo MR E11A, Magnetom Skyra with 24 RF Channel · Siemens Medical Solutions USA, Inc. · 2014Reference device: K211633 · Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR · Canon Medical Systems Corporation · 2021Reference device: K203053 · Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR · Canon Medical Systems Corporation · 2020Reference device: K212056 · Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MR · Canon Medical Systems Corporation · 2021K935595 · Acuson Model 3001 · Acuson Corp. · 1995K973767 · Harmonic Imaging with Contrast · Acuson Corp. · 1997K022567 · Sequoia Signature II Diagnostic Ultrasound System · Acuson Corp. · 2002K032114 · Sequoia 8.0 Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2003×3: 3 generations cleared close together, K041319 (2004), K042770 (2004), K050240 (2005) K052894 · Sonoline G60 S Ultrasound System · Siemens Medical Solutions USA, Inc. · 2005 · recalled×2: 2 generations cleared close together, K063138 (2006), K063803 (2007) K072786 · Acuson S2000 Ultrasound System · Siemens Medical Solutions USA, Inc. · 2007K083095 · Sonixtouch Ultrasound System · Ultrasonix Medical Corporation · 2008K083421 · MR-Touch Option · Ge Medical Systems, LLC · 2009K140666 · MR Elastography · Philips Medical Systems Nederland B.V. · 20141990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors, 1 with a Class I recall13 generations back: 8 other ancestors4 generations back: 19 other ancestors, 5 with a Class I recall55 generations back: 32 other ancestors, 4 with a Class I recall46 generations back: 25 other ancestors, 1 with a Class I recall17 generations back: 14 other ancestors8 generations back: 6 other ancestors9 generations back: 5 other ancestors10 generations back: 9 other ancestors11 generations back: 10 other ancestors K140666 · MR Elastography · Philips Medical Systems Nederland B.V. · 2014K083421 · MR-Touch Option · Ge Medical Systems, LLC · 2009K083095 · Sonixtouch Ultrasound System · Ultrasonix Medical Corporation · 2008K072786 · Acuson S2000 Ultrasound System · Siemens Medical Solutions USA, Inc. · 2007K063803 · Acuson Antares Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · recalledK063138 · Modification to Acuson Antares Ultrasound System · Siemens Medical Solutions USA, Inc. · 2006K052894 · Sonoline G60 S Ultrasound System · Siemens Medical Solutions USA, Inc. · 2005 · recalledK050240 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2005 · recalledK042770 · Modification to Acuson CV70 Cardiovascular System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK041319 · Sequoia Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2004K032114 · Sequoia 8.0 Diagnostic Ultrasound System · Siemens Medical Solutions USA, Inc. · 2003K022567 · Sequoia Signature II Diagnostic Ultrasound System · Acuson Corp. · 2002K973767 · Harmonic Imaging with Contrast · Acuson Corp. · 1997K935595 · Acuson Model 3001 · Acuson Corp. · 1995−10−50
  1. Root · 14 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
  2. 10 generations between
  3. 3 generations backK083095 Sonixtouch Ultrasound SystemUltrasonix Medical Corporation · 2008
  4. 2 generations backK083421 MR-Touch OptionGe Medical Systems, LLC · 2009
  5. 1 generation backK140666 MR ElastographyPhilips Medical Systems Nederland B.V. · 2014
  6. This deviceK220018 MR Elastography; MZEK-001ACanon Medical Systems Corporation · 2022
Show all 14 generations
  1. Root · 14 generations backK935595 Acuson Model 3001Acuson Corp. · 1995
  2. 13 generations backK973767 Harmonic Imaging with ContrastAcuson Corp. · 1997
  3. 12 generations backK022567 Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp. · 2002
  4. 11 generations backK032114 Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2003
  5. 10 generations backK041319 Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004
  6. 9 generations backK042770 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  7. 8 generations backK050240 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
  8. 7 generations backK052894 Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled
  9. 6 generations backK063138 Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2006
  10. 5 generations backK063803 Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI · 2007 · Recalled
  11. 4 generations backK072786 Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2007
  12. 3 generations backK083095 Sonixtouch Ultrasound SystemUltrasonix Medical Corporation · 2008
  13. 2 generations backK083421 MR-Touch OptionGe Medical Systems, LLC · 2009
  14. 1 generation backK140666 MR ElastographyPhilips Medical Systems Nederland B.V. · 2014
  15. This deviceK220018 MR Elastography; MZEK-001ACanon Medical Systems Corporation · 2022

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−14K935595Acuson Model 3001Acuson Corp.1995—
−13K973767Harmonic Imaging with ContrastAcuson Corp.1997—
−12K022567Sequoia Signature II Diagnostic Ultrasound SystemAcuson Corp.2002—
−11K032114Sequoia 8.0 Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2003—
−10K041319Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2004—
−9K042770Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20041
−8K050240Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20051
−7K052894Sonoline G60 S Ultrasound SystemSiemens Medical Solutions USA, Inc.20051
−6K063138Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc.2006—
−5K063803Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI20079
−4K072786Acuson S2000 Ultrasound SystemSiemens Medical Solutions USA, Inc.2007—
−3K083095Sonixtouch Ultrasound SystemUltrasonix Medical Corporation2008—
−2K083421MR-Touch OptionGe Medical Systems, LLC2009—
−1K140666MR ElastographyPhilips Medical Systems Nederland B.V.2014—
0 (this device)K220018MR Elastography; MZEK-001ACanon Medical Systems Corporation2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199210
199450
199550
199610
199770
199840
199990
200090
2001131
2002101
2003111
2004142
2005152
200670
200732
200830
200920
201021
201140
201211
201510
201610
201710
201810
201910
202010
FIG. 1 — Lineage of K220018, 1995–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

11 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140666MR Elastography
Summary page 5
…c Resonance Diagnostic Device) 13. PRODUCT CODE 90-LNH 14. PREDICATE DEVICE Predicate Device: • MR Elastography (K140666) Reference Devices: • Achieva, Intera, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Syst…
Philips Medical Systems Nederland B.V.2014—
*K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Summary page 5
…• MR Elastography (K140666) Reference Devices: • Achieva, Intera, Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems (K193215) • MR Elastography, Resoundant (K201389) Subject Device…
Philips Medical Systems Nederland B.V.2020152
†K201389Resoundant Acoustic Driver SystemResoundant, Inc.2020—
†K103411Ideal IQ Software OptionGE Medical Systems2011—
†K083421MR-Touch OptionGe Medical Systems, LLC2009—
†K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc.2012Class I
†K141977Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo MR E11A…Siemens Medical Solutions USA, Inc.2014—
†K211633Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2021—
†K203053Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2020—
†K212056Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2021—

Also cleared against K140666: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (11 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 2 generations back · ◆ in the 2012 ancestors bar of FIG. 12 Class I recalls
  • K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · 4 generations back · ◆ in the 2010 ancestors bar of FIG. 12 Class I recalls
  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 4 generations back · ◆ in the 2007 ancestors bar of FIG. 11 Class I recall
  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 4 generations back · ◆ in the 2007 ancestors bar of FIG. 11 Class I recall
  • K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · 4 generations back · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 4 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 5 generations back · ◆ in the 2005 ancestors bar of FIG. 11 Class I recall
  • K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · 5 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 5 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 6 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 27 years and passes through 7 companies; a new generation was cleared every 11 months on median.
  • 55 of the 147 ancestors have a recall of any class; 11 have a Class I recall.
  • 1 other device also cites its predicate K140666.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

MR Elastography is an optional package for Canon Medical Systems magnetic resonance imaging (MRI) systems. This option allows the user to obtain an image that reflects the stiffness, in kPa, of body tissue such as liver and muscle.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K220018: MR Elastography; MZEK-001A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220018

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026