Auto Syringe (R) AS50 Infusion Pump

K945942510(k) TraditionalCleared Jul 10, 1995217-day reviewClass II21 CFR 880.5725Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · FRN Pump, Infusion

Cleared on Jul 10, 1995 after a 217-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 7 of the 25 devices in its family, counting itself, have a Class I recall. 24 later devices descend from it. It has been recalled 3 times.

Decided
Jul 10, 1995
Review
217 days
Class
Class II
Product code
FRN
Recalls
3
Predicates
0

24 later devices descend from it, 1 citing it directly. 7 of the 25 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 1 descendant cleared2002: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant cleared2006: 2 descendants cleared2014: 1 descendant cleared, 1 with a Class I recall12011: 3 descendants cleared, 1 with a Class I recall12012: 3 descendants cleared2009: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12016: 2 descendants cleared2022: 2 descendants cleared, 1 with a Class I recall12023: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2013: 1 descendant cleared2015: 1 descendant cleared1995200020052010201520202025
  1. This deviceK945942 Auto Syringe (R) AS50 Infusion PumpBaxter Healthcare Corp · 1995
  2. Descendant · cites it directly · 23 of its ownK982640 Medex 3000 Series Syringe Infusion PumpMedex, Inc. · 1999 · Recalled
  3. Descendant · 2 generations on · 21 of its ownK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. Descendant · 3 generations on · 6 of its ownK053554 CompufloMilestone Scientific, Inc. · 2006
  5. Descendant · 4 generations on · 5 of its ownK101194 Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011
  6. Descendant · 5 generations on · 4 of its ownK112067 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011

Showing 5 of 24 descendants. See all descendants

Descendants cleared per year

1999: 12002: 12006: 22008: 12009: 12011: 32012: 32013: 12014: 12015: 12016: 22018: 12020: 12022: 22023: 12024: 12025: 1
1999high 3 (2011) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 24 · See all
  • K982640Medex 3000 Series Syringe Infusion PumpMedex, Inc. · 1999 · cites it directly23descendants
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations on21descendants
  • K053554CompufloMilestone Scientific, Inc. · 2006 · 3 generations on6descendants
  • K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011 · 4 generations on5descendants
  • K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K945942Auto Syringe (R) AS50 Infusion PumpBaxter Healthcare Corp1995—
+1K982640Medex 3000 Series Syringe Infusion PumpMedex, Inc.19991
+2K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
+3K053554CompufloMilestone Scientific, Inc.2006—
+4K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc.2011—
+5K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200211
200620
200810
200910
201131
201230
201310
201411
201510
201620
201810
202011
202221
202311
202411
202510
FIG. 1 — Lineage of K945942, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (24)

See all 24

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K982640Medex 3000 Series Syringe Infusion PumpMedex, Inc.199923

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 5, 2008Not classifiedZ-0152-2009Nonconforming Material/ComponentNon-conforming Electrostatic Discharge (ESD) grounding squares may cause the pump to be susceptible to shorting-out of circuitry, resulting in a loss of audio, and/or interruption of therapy.
Feb 7, 2006Not classifiedZ-0612-06OtherA non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
May 6, 2005Not classifiedZ-0159-06OtherA 'Check Flange' alarm may occur even though the syringe has been properly loaded into the AS50 Infusion Pump. The false alarm may occur during the syringe loading process or while an infusion is running. Also, the occlusion detection and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · 2 generations on · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
  • K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported…Smiths Medical MD, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 16 Class I recalls
  • K191910SpaceStation MRIB. Braun Melsugen AG · 2020 · 4 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls
  • K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · 4 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 4 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls
  • K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 5 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K982640 (Medex, Inc., 1999) has gone furthest, with 23 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—
Cite this page

510(k) Report. “K945942: Auto Syringe (R) AS50 Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K945942

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026