Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511

K091195510(k) TraditionalCleared May 27, 200934-day reviewClass II21 CFR 892.1750Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on May 27, 2009 as substantially equivalent to K012009 (Marconi Medical Systems, Inc., 2001) after a 34-day review. Its family tree reaches back 4 generations to 1993. 1 later device descends from it. It has been recalled 8 times.

Decided
May 27, 2009
Review
34 days
Class
Class II
Product code
JAK
Recalls
8
Predicates
1

Its longest line of descent runs back 4 generations to K930090 (1993). 9 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2019: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 3 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 19930
  1. Root · 4 generations backK930090 HelicatElscint, Inc. · 1993
  2. 3 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 2 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 1 generation backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. This deviceK091195 Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511Philips Medical Systems (Cleveland), Inc. · 2009
  6. Descendant · cites it directly · 0 of its ownK191136 Access CTPhilips Healthcare (Suzhou) Co., Ltd. · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K191136Access CTPhilips Healthcare (Suzhou) Co., Ltd. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K930090HelicatElscint, Inc.1993—
−3K961464Select/SPElscint, Inc.19963
−2K982060Volumax CT Imaging SystemElscint, Inc.1998—
−1K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
0 (this device)K091195Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511Philips Medical Systems (Cleveland), Inc.2009—
+1K191136Access CTPhilips Healthcare (Suzhou) Co., Ltd.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K091195, 1993–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K012009MX8000 Idt CT System
Summary page 1
…, phase II is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: the Brilliance CT 16-slice in K012009 (a.k.a. MX8000 IDT CT Scanner). It expands the capabilities of the MX 16-slice CT system (K083498). 4. Device Description Th…
Marconi Medical Systems, Inc.200161

Also cleared against K012009: 10 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191136Access CTPhilips Healthcare (Suzhou) Co., Ltd.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2021Class IIZ-1980-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Apr 20, 2016Class IIZ-2346-2016Software designThe following issues have been found in MX 16-slice systems with software version 1.1.4.21426: 1. During the filming operation on MX16-slice console software, the clipboard used for copying and pasting images is not cleared between…
Apr 20, 2016Class IIZ-2347-2016Software designThe following issues have been found in MX 16-slice systems with software version 1.1.4.21426: 1. During the filming operation on MX16-slice console software, the clipboard used for copying and pasting images is not cleared between…
Mar 6, 2015Class IIZ-1759-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1760-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Oct 9, 2014Class IIZ-0190-2015Software Design Change"The orientation of images is displayed incorrectly. When operator chooses gantry on the right side from operation console, and the patient position is HFDL, HFDR, FFDL or FFDR, the patient position picture and text is not consistent. "L"…
Aug 24, 2012Class IIZ-2386-2012Process designOn August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
Mar 10, 2011Not classifiedZ-2059-2011OtherThe Philips MX 16-slice CT scanners are subject to recall/field correction due to a problem encountered with the units when using the 'Combine Viewing' Option. The firm determined that inaccurate dimension measurements display when using…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 3.3 years on median.
  • 3 of the 9 ancestors have a recall of any class.
  • Of its descendants, K191136 (Philips Healthcare (Suzhou) Co., Ltd., 2019) has gone furthest, with 0 descendants of its own.
  • 10 other devices also cite its predicate K012009.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Philips MX 16-slice CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating x-ray tube and detector array with multislice capability up to 16 slices simultaneously. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091195: Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K091195

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026