510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Philips Iqon Spectral CT

K133674510(k) TraditionalCleared Nov 21, 2014357-day reviewClass II21 CFR 892.1750Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Healthcare (Cleveland) · JAK System, X-Ray, Tomography, Computed

Cleared on Nov 21, 2014 as substantially equivalent to K131773 (Philips Healthcare (Cleveland), 2013) after a 357-day review. Its family tree reaches back 7 generations to 1993. 10 later devices descend from it. It has been recalled 6 times.

Decided
Nov 21, 2014
Review
357 days
Class
Class II
Product code
JAK
Recalls
6
Predicates
1

Its longest line of descent runs back 7 generations to K930090 (1993). 12 ancestors in all. 10 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2015: 1 descendant cleared2017: 1 descendant cleared2026: 3 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant cleared2022: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalledK060937 · Brilliance Volume · Philips Medical Systems (Cleveland), Inc. · 2006 · recalledK131773 · Brilliance Ict · Philips Healthcare (Cleveland) · 20131990200020102020
Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor6 generations back: 3 other ancestors7 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 3 descendants5 generations forward: 2 descendants K131773 · Brilliance Ict · Philips Healthcare (Cleveland) · 2013K060937 · Brilliance Volume · Philips Medical Systems (Cleveland), Inc. · 2006 · recalledK033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50+5
  1. Root · 7 generations backK930090 HelicatElscint, Inc. · 1993
  2. 6 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 5 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 4 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 3 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 2 generations backK060937 Brilliance VolumePhilips Medical Systems (Cleveland), Inc. · 2006 · Recalled
  7. 1 generation backK131773 Brilliance IctPhilips Healthcare (Cleveland) · 2013
  8. This deviceK133674 Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014
  9. Descendant · cites it directly · 7 of its ownK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  10. Descendant · 2 generations on · 6 of its ownK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  11. Descendant · 3 generations on · 5 of its ownK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  12. Descendant · 4 generations on · 3 of its ownK240844 Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024
  13. Descendant · cites it directly · 1 of its ownK150665 Philips Spectral CT ApplicationsPhilips Medical Systems Nethrlands BV · 2015 · Recalled

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2015: 12017: 12020: 12021: 12022: 12024: 12025: 12026: 3
2015high 3 (2026) · low 0 (2016)2026

Most influential descendants

Showing 5 of 10 · See all
  • K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · cites it directly7descendants
  • K193454IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · 2 generations on6descendants
  • K203020Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · 3 generations on5descendants
  • K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024 · 4 generations on3descendants
  • K150665Philips Spectral CT ApplicationsPhilips Medical Systems Nethrlands BV · 2015 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K930090HelicatElscint, Inc.1993—
−6K961464Select/SPElscint, Inc.19963
−5K982060Volumax CT Imaging SystemElscint, Inc.1998—
−4K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−3K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−2K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200661
−1K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
0 (this device)K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)2014—
+1K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.20172
+2K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206
+3K203020Spectral CTPhilips Medical Systems Nederland B.V.20218
+4K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd.2024—
+1K150665Philips Spectral CT ApplicationsPhilips Medical Systems Nethrlands BV20151

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201710
202010
202110
202210
202410
202510
202630
FIG. 1 — Lineage of K133674, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131773Brilliance Ict
Summary page 4
…750 Class: II Product Code: JAK Panel: Radiology Predicate devices Philips Brilliance iCT (K131773) Premarket Notification [510(k)] Submission Philips Medical Systems (Cleveland), Inc. Sectio…
Philips Healthcare (Cleveland)2013—

What descends from it? (10)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.20177
K150665Philips Spectral CT ApplicationsPhilips Medical Systems Nethrlands BV20151

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 8, 2018Class IIZ-1414-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Jan 11, 2018Class IIZ-0841-2018Under Investigation by firm23 Software related issues for the IQon Spectral CT with 4.7.2 software version. These issues include: " Software errors that may result in CT spectral images misrepresentation or incorrect treatment; " Software errors that may result in…
Nov 8, 2017Class IIZ-0333-2018Device DesignThere is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the…
Dec 16, 2016Class IIZ-1006-2017Software designMultiple issues have caused the device to result in CT rescans or incorrect scan location or misrepresentation of image results.
Nov 14, 2016Class IIZ-0705-2017Software designA software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images
Oct 13, 2016Class IIIZ-0598-2017Labeling Change ControlThe product label does not include the correct current rating.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 5 of the 12 ancestors have a recall of any class.
  • Of its descendants, K163711 (Philips Medical Systems (Cleveland), Inc., 2017) has gone furthest, with 7 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Philips IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum…

Read the full statement (167 words)

The Philips IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.

This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133674: Philips Iqon Spectral CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K133674

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026