510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

NeuViz 128 Multi-slice CT Scanner System

K151383510(k) TraditionalCleared Nov 4, 2015166-day reviewClass II21 CFR 892.1750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Neusoft Medical Systems Co., Ltd. · JAK System, X-Ray, Tomography, Computed

Cleared on Nov 4, 2015 as substantially equivalent to K033326 (Philips Medical Systems, Inc., 2003) after a 166-day review. Its family tree reaches back 5 generations to 1993. 2 later devices descend from it. It has been recalled 3 times.

Decided
Nov 4, 2015
Review
166 days
Class
Class II
Product code
JAK
Recalls
3
Predicates
1

Its longest line of descent runs back 5 generations to K930090 (1993). 10 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor4 generations back: 3 other ancestors5 generations back: 1 other ancestor 1 generation forward: 2 descendants K033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50
  1. Root · 5 generations backK930090 HelicatElscint, Inc. · 1993
  2. 4 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 3 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 2 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 1 generation backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. This deviceK151383 NeuViz 128 Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd. · 2015
  7. Descendant · cites it directly · 0 of its ownK230220 NeuViz 128 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd. · 2023
  8. Descendant · cites it directly · 0 of its ownK171201 NeuViz Prime Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd. · 2017

Most influential descendants

Showing 2 of 2 · See all
  • K230220NeuViz 128 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd. · 2023 · cites it directly0descendants
  • K171201NeuViz Prime Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd. · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K930090HelicatElscint, Inc.1993—
−4K961464Select/SPElscint, Inc.19963
−3K982060Volumax CT Imaging SystemElscint, Inc.1998—
−2K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−1K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
0 (this device)K151383NeuViz 128 Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.2015—
+1K230220NeuViz 128 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.2023—
+1K171201NeuViz Prime Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202310
FIG. 1 — Lineage of K151383, 1993–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033326Philips Plus Computed Tomography System
Summary page 6
…diac Function Analysis z Cardiac Viewer z Fat Analysis z CTDSA z Tumor Assessment Predicated Device: Philips Ingenuity Core128 CT System(K033326) Reference Device NeuViz 64e CT system (K121792). Statement of Substantial Equivalence: The NeuViz 128 Multi-slice CT Sc…
Philips Medical Systems, Inc.2003131
†K121792Neuviz 64 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20125

Also cleared against K033326: 8 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230220NeuViz 128 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20230
K171201NeuViz Prime Multi-slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 9, 2019Class IIZ-0773-2019Software designFor surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and…
Oct 22, 2018Class IIZ-1280-2019Software designSoftware defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot…
May 19, 2017Class IIZ-3045-2017Software designSoftware defect

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • Of its descendants, K230220 (Neusoft Medical Systems Co., Ltd., 2023) has gone furthest, with 0 descendants of its own.
  • 8 other devices also cite its predicate K033326.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The NeuViz 128 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151383: NeuViz 128 Multi-slice CT Scanner System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151383

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.