510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Brilliance Volume

K060937510(k) TraditionalCleared Jun 5, 200661-day reviewClass II21 CFR 892.1750Recalls 61

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems (Cleveland), Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Jun 5, 2006 as substantially equivalent to K033326 (Philips Medical Systems, Inc., 2003) after a 61-day review. Its family tree reaches back 5 generations to 1993. 24 later devices descend from it. It has been recalled 61 times.

Decided
Jun 5, 2006
Review
61 days
Class
Class II
Product code
JAK
Recalls
61
Predicates
1

Its longest line of descent runs back 5 generations to K930090 (1993). 10 ancestors in all. 24 later devices descend from it, 8 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2017: 2 descendants cleared2008: 1 descendant cleared2013: 2 descendants cleared2020: 2 descendants cleared2009: 1 descendant cleared2012: 2 descendants cleared2014: 2 descendants cleared2010: 1 descendant cleared2025: 2 descendants cleared2026: 4 descendants cleared2011: 1 descendant cleared2015: 1 descendant cleared2021: 1 descendant cleared2024: 1 descendant cleared2022: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor4 generations back: 3 other ancestors5 generations back: 1 other ancestor 1 generation forward: 8 descendants2 generations forward: 4 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 1 descendant6 generations forward: 3 descendants7 generations forward: 2 descendants K033326 · Philips Plus Computed Tomography System · Philips Medical Systems, Inc. · 2003 · recalledK012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 1993−50+5
  1. Root · 5 generations backK930090 HelicatElscint, Inc. · 1993
  2. 4 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 3 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 2 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 1 generation backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. This deviceK060937 Brilliance VolumePhilips Medical Systems (Cleveland), Inc. · 2006
  7. Descendant · cites it directly · 11 of its ownK131773 Brilliance IctPhilips Healthcare (Cleveland) · 2013
  8. Descendant · 2 generations on · 10 of its ownK133674 Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014 · Recalled
  9. Descendant · 3 generations on · 7 of its ownK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  10. Descendant · 4 generations on · 6 of its ownK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  11. Descendant · 5 generations on · 5 of its ownK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled

Showing 5 of 24 descendants. See all descendants

Descendants cleared per year

2008: 12009: 12010: 12011: 12012: 22013: 22014: 22015: 12017: 22020: 22021: 12022: 12024: 12025: 22026: 4
2008high 4 (2026) · low 0 (2016)2026

Most influential descendants

Showing 5 of 24 · See all
  • K131773Brilliance IctPhilips Healthcare (Cleveland) · 2013 · cites it directly11descendants
  • K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014 · 2 generations on10descendants
  • K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · 3 generations on7descendants
  • K193454IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · 4 generations on6descendants
  • K203020Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · 5 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K930090HelicatElscint, Inc.1993—
−4K961464Select/SPElscint, Inc.19963
−3K982060Volumax CT Imaging SystemElscint, Inc.1998—
−2K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−1K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
0 (this device)K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.2006—
+1K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
+2K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)20146
+3K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.20172
+4K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206
+5K203020Spectral CTPhilips Medical Systems Nederland B.V.20218

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
201010
201110
201220
201320
201420
201510
201720
202020
202110
202210
202410
202520
202640
FIG. 1 — Lineage of K060937, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033326Philips Plus Computed Tomography System
Summary page 1
…e Volume" CT scanner is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: the Philips Plus in K033326. 4. Device Description The "Brilliance Volume" is a Whole Body Computed Tomography X-Ray System featuring a continuously rot…
Philips Medical Systems, Inc.2003131

Also cleared against K033326: 8 other devices →

What descends from it? (24)

See all 24

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131773Brilliance IctPhilips Healthcare (Cleveland)201311
K090462Brilliance Dual Energy OptionPhilips Medical Systems (Cleveland), Inc.20093
K203216The Multimodality Advanced Vessel AnalysisPhilips Medical Systems Nederland B.V.20201
K162838Philips iCT CT SystemPhilips Medical Systems (Cleveland), Inc.20171
K113483Iterative Reconstruction Technique Software ApplicationPhilips Healthcare (Cleveland)20121
K133603Philips CT Dynamic Myocardial Perfusion (DMP) ApplicationPhilips Medical Systems20140
K123576Imr Software ApplicationPhilips Healthcare (Cleveland)20130
K081482CSBP-001A Body Organ Perfusion SystemToshiba America Medical Systems, In.C20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 13, 2025Class IIZ-2599-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
May 1, 2021Class IIZ-1990-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Apr 13, 2021Class IIZ-1774-2021Under Investigation by firmCooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury
Apr 6, 2020Class IIZ-1921-2020Process controlThe bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
Apr 6, 2020Class IIZ-1922-2020Process controlThe bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
Feb 8, 2018Class IIZ-1409-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1408-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Nov 8, 2017Class IIZ-0332-2018Device DesignThere is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the…
Nov 8, 2017Class IIZ-0331-2018Device DesignThere is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the…
Jul 11, 2017Class IIZ-0402-2018Software designNumerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other…
Jul 11, 2017Class IIZ-0403-2018Software designNumerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other…
Dec 23, 2016Class IIZ-1058-2017Employee errorDuring a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
Nov 25, 2016Class IIZ-1004-2017Employee errorFailure to correctly document the installation of four M12 Bolts into the system rotor.
Oct 13, 2016Class IIIZ-0594-2017Labeling Change ControlThe product label does not include the correct current rating.
Oct 13, 2016Class IIIZ-0593-2017Labeling Change ControlThe product label does not include the correct current rating.
Jun 21, 2016Class IIZ-2112-2016Nonconforming Material/ComponentSupplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack.
Apr 29, 2016Class IIZ-2382-2016Software designSoftware issues found in v4.1 .3/4.1.5 in the Philips Brilliance iCT/ iCT SP products that could affect the performance of the equipment.
Apr 29, 2016Class IIZ-2381-2016Software designSoftware issues in versions v4.1 .3/4.1.5 in the Philips Brilliance iCT) iCT SP products that, could affect the performance of the equipment.
Mar 4, 2016Class IIZ-1506-2016Software designThe AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data…
Feb 5, 2016Class IIZ-1352-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Feb 5, 2016Class IIZ-1353-2016Software designThe firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
Oct 29, 2015Class IIZ-0545-2016Software designPerfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion…
Oct 29, 2015Class IIZ-0546-2016Software designPerfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion…
Sep 6, 2015Class IIZ-1713-2016Radiation Control for Health and Safety ActSoftware Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight…
Sep 6, 2015Class IIZ-1715-2016Radiation Control for Health and Safety ActSoftware Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight…
Sep 6, 2015Class IIZ-1714-2016Radiation Control for Health and Safety ActSoftware Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight…
Apr 13, 2015Class IIZ-1695-2015OtherA calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Mar 16, 2015Class IIZ-1656-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 6, 2015Class IIZ-1767-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1768-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1769-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Feb 5, 2015Class IIZ-2256-2015Software change control5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value…
Feb 5, 2015Class IIZ-1359-2015Software designPhilips discovered that a software defect exists in marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.
Feb 5, 2015Class IIZ-2255-2015Software change control5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value…
Dec 8, 2014Class IIZ-0979-2015Process controlImproper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
Jul 31, 2014Class IIZ-0236-2015Radiation Control for Health and Safety ActIt was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.
Jul 27, 2014Class IIZ-2237-2014Software designThe following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction…
Apr 1, 2014Class IIZ-1722-2014Component design/selectionA customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
Mar 27, 2014Class IIZ-1461-2014Nonconforming Material/ComponentBlock assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can…
Mar 11, 2014Class IIZ-1678-2014Software designDuring SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in…
Dec 9, 2013Class IIZ-0832-2014Software designPatient images exhibited ring artifacts.
Oct 11, 2013Class IIZ-0335-2014Device DesignIf the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with…
Jul 16, 2013Class IIZ-0831-2014Device DesignAn artifact that resembles thrombus may appear on the image.
Dec 20, 2012Class IIZ-0945-2013Component design/selectionPhilips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
Jul 2, 2012Class IIZ-2035-2012Process controlPhilips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
May 23, 2012Class IIZ-1987-2012Software designThe Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.
May 23, 2012Class IIZ-1988-2012Software designPhilips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a…
May 22, 2012Class IIZ-2000-2012Software designIn the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
May 7, 2012Not classifiedZ-1688-2012Nonconforming Material/ComponentOn 3/5/2012, Philips Healthcare was notified of one incident involving a Brilliance iCT scanner that resulted in damage to the gantry and slight injury to the user when the heat exchanger detached from the cooling unit due to some missing…
May 7, 2012Class IIZ-1791-2012Software designPhilips is issuing a software update (version 3.2.4) to the Brilliance iCT and iCT SP systems currently running software version 3.2.0, 3.2.1 or 3.2.3, addressing a number of non-conformances. A review was conducted of all nonconformances…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 4 of the 10 ancestors have a recall of any class.
  • Of its descendants, K131773 (Philips Healthcare (Cleveland), 2013) has gone furthest, with 11 descendants of its own.
  • 8 other devices also cite its predicate K033326.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The "Brilliance Volume" is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060937: Brilliance Volume.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K060937

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.