510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Spectral CT on Rails

K212875510(k) TraditionalCleared May 13, 2022246-day reviewClass II21 CFR 892.1750Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · JAK System, X-Ray, Tomography, Computed

Cleared on May 13, 2022 as substantially equivalent to K203020 (Philips Medical Systems Nederland B.V., 2021) and 1 other predicate after a 246-day review. Its family tree reaches back 11 generations to 1993. It has been recalled 6 times.

Decided
May 13, 2022
Review
246 days
Class
Class II
Product code
JAK
Recalls
6
Predicates
2

Its longest line of descent runs back 11 generations to K930090 (1993). 20 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 11 generations backK930090 HelicatElscint, Inc. · 1993
  2. 7 generations between
  3. 3 generations backK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  4. 2 generations backK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  5. 1 generation backK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  6. This deviceK212875 Spectral CT on RailsPhilips Medical Systems Nederland B.V. · 2022
Show all 11 generations
  1. Root · 11 generations backK930090 HelicatElscint, Inc. · 1993
  2. 10 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 9 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 8 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 7 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 6 generations backK060937 Brilliance VolumePhilips Medical Systems (Cleveland), Inc. · 2006 · Recalled
  7. 5 generations backK131773 Brilliance IctPhilips Healthcare (Cleveland) · 2013
  8. 4 generations backK133674 Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014 · Recalled
  9. 3 generations backK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  10. 2 generations backK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  11. 1 generation backK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  12. This deviceK212875 Spectral CT on RailsPhilips Medical Systems Nederland B.V. · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K930090HelicatElscint, Inc.1993—
−10K961464Select/SPElscint, Inc.19963
−9K982060Volumax CT Imaging SystemElscint, Inc.1998—
−8K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−7K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−6K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200661
−5K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
−4K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)20146
−3K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.20172
−2K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206
−1K203020Spectral CTPhilips Medical Systems Nederland B.V.20218
0 (this device)K212875Spectral CT on RailsPhilips Medical Systems Nederland B.V.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200010
200110
200410
201710
201810
FIG. 1 — Lineage of K212875, 1993–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K203020Spectral CT
Summary page 4
…Spectral CT Predicate Manufacturer: Philips Medical Systems Nederland B.V. Device: 510(k) Clearance: K203020 Classification Name: Computed Tomography x-Ray System Classification Regulation:…
Philips Medical Systems Nederland B.V.20218
*K181797Philips CT Big Bore Sliding Gantry Configuration
Summary page 4
…iding Gantry Configuration Predicate Manufacturer: Philips Medical Systems Nederland B.V. Device: 510(k) Clearance: K181797 Classification Name: Computed Tomography x-Ray System Classification Regulation:…
Phillips Medical Systems (Cleveland), Inc.2018—

Also cleared against K203020: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 7, 2026Class IIZ-1825-2026Under Investigation by firmPhilips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview…
Jan 7, 2026Class IIZ-1318-2026Process change controlAs part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to…
Feb 28, 2025Class IIZ-1324-2025Under Investigation by firmMultiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
Dec 13, 2023Class IIZ-1054-2024Under Investigation by firmA software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment.
Oct 9, 2023Class IIZ-0306-2024Under Investigation by firmWhen preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.
Oct 9, 2023Class IIZ-0319-2024Under Investigation by firmThis recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 6 companies; a new generation was cleared every 2.4 years on median.
  • 12 of the 20 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K203020.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Spectral CT on Rails is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The Spectral CT on Rails system acquires one CT dataset — composed of data from a higher-energy detected x-ray spectrum…

Read the full statement (270 words)

The Spectral CT on Rails is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

The Spectral CT on Rails system acquires one CT dataset — composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.

This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures in patients of all ages, and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology.

The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.

The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl. ] Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K212875: Spectral CT on Rails.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K212875

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026