510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Spectral CT

K244008510(k) TraditionalCleared May 16, 2025141-day reviewClass II21 CFR 892.1750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Technologies, Ltd. · JAK System, X-Ray, Tomography, Computed

Cleared on May 16, 2025 as substantially equivalent to K203020 (Philips Medical Systems Nederland B.V., 2021) and 1 other predicate after a 141-day review. Its family tree reaches back 12 generations to 1993. It has been recalled 3 times.

Decided
May 16, 2025
Review
141 days
Class
Class II
Product code
JAK
Recalls
3
Predicates
2

Its longest line of descent runs back 12 generations to K930090 (1993). 20 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 12 generations backK930090 HelicatElscint, Inc. · 1993
  2. 8 generations between
  3. 3 generations backK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  4. 2 generations backK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  5. 1 generation backK240844 Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024
  6. This deviceK244008 Spectral CTPhilips Medical Systems Technologies, Ltd. · 2025
Show all 12 generations
  1. Root · 12 generations backK930090 HelicatElscint, Inc. · 1993
  2. 11 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 10 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 9 generations backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. 8 generations backK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. 7 generations backK060937 Brilliance VolumePhilips Medical Systems (Cleveland), Inc. · 2006 · Recalled
  7. 6 generations backK131773 Brilliance IctPhilips Healthcare (Cleveland) · 2013
  8. 5 generations backK133674 Philips Iqon Spectral CTPhilips Healthcare (Cleveland) · 2014 · Recalled
  9. 4 generations backK163711 IQon Spectral CTPhilips Medical Systems (Cleveland), Inc. · 2017 · Recalled
  10. 3 generations backK193454 IQon Spectral CTPhilips Medical Systems Nederland B.V. · 2020 · Recalled
  11. 2 generations backK203020 Spectral CTPhilips Medical Systems Nederland B.V. · 2021 · Recalled
  12. 1 generation backK240844 Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd. · 2024
  13. This deviceK244008 Spectral CTPhilips Medical Systems Technologies, Ltd. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K930090HelicatElscint, Inc.1993—
−11K961464Select/SPElscint, Inc.19963
−10K982060Volumax CT Imaging SystemElscint, Inc.1998—
−9K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
−8K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
−7K060937Brilliance VolumePhilips Medical Systems (Cleveland), Inc.200661
−6K131773Brilliance IctPhilips Healthcare (Cleveland)2013—
−5K133674Philips Iqon Spectral CTPhilips Healthcare (Cleveland)20146
−4K163711IQon Spectral CTPhilips Medical Systems (Cleveland), Inc.20172
−3K193454IQon Spectral CTPhilips Medical Systems Nederland B.V.20206
−2K203020Spectral CTPhilips Medical Systems Nederland B.V.20218
−1K240844Spectral CT 7500 RTPhilips Medical Systems Technologies, Ltd.2024—
0 (this device)K244008Spectral CTPhilips Medical Systems Technologies, Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200010
200110
200410
201710
FIG. 1 — Lineage of K244008, 1993–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K203020Spectral CT
Summary page 4
…nformation provided above, the proposed Spectral CT system is considered substantially equivalent to the currently marketed primary predicate device, Spectral CT system (K203020), in terms of Indications for Use. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 C…
Philips Medical Systems Nederland B.V.20218
*K240844Spectral CT 7500 RT
Summary page 5
…ice: Manufacturer: Philips Medical Systems Nederland B.V. 510(k) Clearance: K240844 Classification Name: Computed Tomography X-ray System Classi…
Philips Medical Systems Technologies, Ltd.2024—

Also cleared against K203020: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 21, 2026Class IIZ-0078-2027Under Investigation by firmSoftware issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Aug 18, 2026Class IIZ-3155-2026Radiation Control for Health and Safety ActThe Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Jan 7, 2026Class IIZ-1317-2026Process change controlAs part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 7 companies; a new generation was cleared every 2.6 years on median.
  • 12 of the 20 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K203020.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Spectral CT system is a Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment support, component parts, and accessories. The Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected X-ray spectrum and a…

Read the full statement (317 words)

The Spectral CT system is a Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment support, component parts, and accessories.

The Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected X-ray spectrum and a lower- energy detected X-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.

This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures in patients of all ages, and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology as part of treatment preparation and radiation therapy planning. The Extended field of view images and respiratory correlated scanning (4DCT) are for treatment preparation and radiation therapy planning/simulation usage only.

This device is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.

*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl. J Med 2011; 365:395-409) and subsequent literature, for further information.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K244008: Spectral CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K244008

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026